  "results": [
    {
      "status": "Ongoing",
      "city": "Lugano",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85469",
      "recalling_firm": "Inpeco S.A.",
      "address_1": "Via San Gottardo 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of IL, NY.",
      "recall_number": "Z-2200-2020",
      "product_description": "Aptio Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation.",
      "product_quantity": "61 systems",
      "reason_for_recall": "In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.",
      "recall_initiation_date": "20200407",
      "center_classification_date": "20200601",
      "report_date": "20200610",
      "code_info": "Serial Numbers:  AP2.0005  AP2.0008  AP2.0010  AP2.0014  AP2.0017  AP2.0018  AP2.0022  AP2.0023  AP2.0026  AP2.0027  AP2.0033  AP2.0036  AP2.0039  AP2.0043  AP2.0045  AP2.0056  AP2.0060  AP2.0065  AP2.0069  AP2.0070  AP2.0071  AP2.0076  AP2.0078  AP2.0081  AP2.0087  AP2.0089  AP2.0102  AP2.0104  AP2.0105  AP2.0110  AP2.0111  AP2.0113  AP2.0120  AP2.0127  AP2.0137  AP2.0139  AP2.0146  AP2.0147  AP2.0149  AP2.0151  AP2.0155  AP2.0165  AP2.0170  AP2.0172  AP2.0184  AP2.0189  AP2.0201  AP2.0202  AP2.0203  AP2.0209  AP2.0211  AP2.0213  AP2.0217  AP2.0223  APT.0009  APT.0014  APT.0058  APT.0065  APT.0093  APT.0091  APT.0125"
    }
  ]
}