      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71689",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-US: All states in the US including DC and PR. OUS: Australia, Austria, Bangladesh, Belgium, Brazil, China, Denmark, France, Germany, Hong Kong, India, Indonesia, Ireland, Italy, Japan, Malaysia, Mexico, Netherlands,New Zealand, Oman, Philippines, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Kingdom, Viet Nam.    UPDATED:   Canada, Bulgaria, Egypt, Hungary, Jordan, Luxembourg, Poland, Reunion, Romania, Russia, Sauth Arabia, Turkey.",
      "recall_number": "Z-2200-2015",
      "product_description": "Pinnacle3  Software Version 9.0, 9.2, 9.4 and 9.6, Model Numbers  453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451.    UPDATED:     Pinnacle3  Software Version 8.0h, 8.0k, 8.0m, 8.0n.",
      "product_quantity": "1642",
      "reason_for_recall": "Philips, Pinnacle Radiation Treatment Planning System version 8.0h, 8.0k, 8.0m, 8.0n, 9 0, 9 2 9 4, 9 6 is being recalled because the  dose may be inconsistent with the density of a density-overridden ROl.",
      "recall_initiation_date": "20140731",
      "center_classification_date": "20150722",
      "termination_date": "20190508",
      "report_date": "20150729",
      "code_info": "453560446041, 459800091001, 459800220161, 459800232931, 459800235871, 459800338451    UPDATED:    453560446051 [A]   459800096361 [Al  453560446061 IA]   459800096541 [A]  453560446091 [A]   459800096741 [A]  453560446101 [A]   459800096921 [A]  453560446111 [A]   459800097111 [A]  453560446131 [A]   459800097291 [A]  453560446141 [A]   459800097471 [A]  453560446161 [A]   459800097651 [B]  453560446181 [A]   459800097831 [B]  453560446201 [A]   459800098171 [B]  459800089221 [A]   459800098351 [A]  459800091451 [A]   459800109261 [A]  459800138101 [BJ  459800189431 [A)  459800220131 [B]  459800239171 [A]  459800233271 [8]  459800233281 [A]  459800233291 [B]  459800233301 [A]  459800233311 [8]  459800233321 [B]  459800233331 [B]  459800233341 [A]  459800233351 [A]  459800233361 [A]  459800234101 [A]  459800234111 [A]  459800234121 [A]  459800318391 [Al  459800405751 [A]  459800417961 [A]  459800249651 [8]  459800249721 [A]  459800249851 [A]  459800249931 [A]  459800250001 [Al  459800250081 [A]  459800250151 [A]  459800250221 [A]  459800250291 [A]  459800250361 [A]  459800250431 [A]  459800250501 [A]  459800250571 [A]  459800338961 [B]  459800340181 [A]  459800340201 [A]  459800341171 [A]  459800341191 [A]  459800341481 [A]  459800341501 [A]  459800345001 [A]  459800345021 [A]  459800345041 [A]  459800345061 [A]  459800345081 [A]  459800345101 [A]  459800454251 [A]  459800455341 [A]  459800493591 [A]  459800494011 [A]  459800554831 [A]  459800554881 [A]  453560435761  453560434881  453560454791  453560461331  453560461341"
    }
  ]
}