{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Boerne",
      "state": "TX",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88323",
      "recalling_firm": "Stanbio Laboratory, LP",
      "address_1": "1261 N Main St",
      "address_2": "",
      "postal_code": "78006-3014",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MO, NC, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, TN, TX, VA, VT, WA, WI, and WV. The countries of  Australia, Canada, Chile, Federal Republic of Germany, Hong Kong, Singapore, Taiwan, Thailand, United Kingdom, and Viet Nam.",
      "recall_number": "Z-2199-2021",
      "product_description": "Beta-Hydroxybutyrate LiquiColor.  in vitro diagnostic reagent.",
      "product_quantity": "7,108 products",
      "reason_for_recall": "Deterioration in the stability of the reagents which has resulted in lower than expected or out of range Quality Control values.",
      "recall_initiation_date": "20210712",
      "center_classification_date": "20210805",
      "termination_date": "20220329",
      "report_date": "20210811",
      "code_info": "Catalog Number/Product Description: 2440-058/¿-Hydroxybutyrate LiquiColor Device Identifier: 00657498000519 Lot Numbers: 164312; 161081; 161039; 161031; 158972;",
      "more_code_info": ""
    }
  ]
}