  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lugano",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85469",
      "recalling_firm": "Inpeco S.A.",
      "address_1": "Via San Gottardo 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution including in the states of IL, NY.",
      "recall_number": "Z-2199-2020",
      "product_description": "Accelerator a3600 Automation System using the Aliquoter Module (Inpeco P/N FLX-212) - Product Usage: The systems consolidate multiple Analytical instruments into a unified workstation.",
      "product_quantity": "97 systems",
      "reason_for_recall": "In certain firmware versions, in case a Clot Detection error (error code E0E0 or 13E0) is generated during the sample aspiration the current error recovery procedure dispenses 2/3 of sample volume back into the Primary Tube. Evidence from the field showed that in case of Clot Detection error, this management may lead to the dilution of the Primary Tube with the distilled water of the hydraulic circuit of the Aliquoter Module.",
      "recall_initiation_date": "20200407",
      "center_classification_date": "20200601",
      "report_date": "20200610",
      "code_info": "Serial Numbers:  ACP.002  ACP.004  ACP.009  ACP.011  ACP.012  ACP.015  ACP.018  ACP.019  ACP.026  ACP.027  ACP.028  ACP.033  ACP.034  ACP.042  ACP.045  ACP.047  ACP.048  ACP.050  ACP.052  ACP.059  ACP.061  ACP.063  ACP.064  ACP.066  ACP.069  ACP.070  ACP.071  ACP.073  ACP.076  ACP.085  ACP.088  ACP.092  ACP.093  ACP.101  ACP.102  ACP.104  ACP.113  ACP.119  ACP.120  ACP.123  ACP.131  ACP.141  ACP.142  ACP.146  ACP.163  ACP.164  ACP.171  ACP.172  ACP.182  ACP.185  ACP.186  ACP.194  ACP.205  ACP.209  ACP.214  ACP.217  ACP.218  ACP.226  ACP.229  ACP.230  ACP.232  ACP.233  ACP.235  ACP.236  ACP.237  ACP.242  ACP.243  ACP.244  ACP.248  ACP.250  ACP.252  ACP.253  ACP.257  ACP.258  ACP.259  ACP.263  ACP.265  ACP.267  ACP.268  ACP.275  ACP.276  ACP.280  ACP.286  ACP.289  ACP.290  ACP.291  ACP.294  ACP.295  ACP.297  ACP.298  ACP.299  ACP.303  ACP.304  ACP.314  ACP.323  ACP.332  ACP.336"
    }
  ]
}