{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88282",
      "recalling_firm": "Straumann USA LLC",
      "address_1": "60 Minuteman Rd",
      "address_2": "",
      "postal_code": "01810-1008",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2198-2021",
      "product_description": "Straumann SCS Screwdriver, driver, for ratchet, short, L 21mm, stainless steel\t  Article Number: 046.401. dental surgical instrument.",
      "product_quantity": "130 pieces",
      "reason_for_recall": "Tip of the screwdriver does not have the correct configuration / profile therefore the screwdriver does not properly fit the opposing screw head and cannot pick up the screws, may result in injury",
      "recall_initiation_date": "20210527",
      "center_classification_date": "20210803",
      "termination_date": "20230321",
      "report_date": "20210811",
      "code_info": "Packaged lot number: EHG85 Lot Number engraved on part: CXV34",
      "more_code_info": ""
    }
  ]
}