  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Twinsburg",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65980",
      "recalling_firm": "Assuramed",
      "address_1": "1810 Summit Commerce Par",
      "address_2": "K",
      "postal_code": "44087",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution including AR, AZ, CA, CO, FL, IL, KY, LA, MI, NJ, NY, OH,OK, OR, SC, TX, UT, and WA.",
      "recall_number": "Z-2196-2013",
      "product_description": "DiaScreen Liquid Urine Controls: Positive and Negative Assayed Liquid Quality Controls for use with Dipsticks in Routine Urinalysis; Product Number D76004 (4x12 mL bottles) consists of: 2x12mL Positive controls [and] 2x12mL Negative controls; Manufactured for ARKRAY USA, Inc.; Minneapolis, MN 55439    Diascreen liquid urine control is a synthetic control designed to react with commercial urine dipsticks to register listed responses on color reagent pads. It is intended for use in a clinical laboratory as a urine control for qualitative procedures used in chemical determinations. The control is intended to be used exactly as if it were a patient sample to assist in the assessment of the listed analytical procedures.",
      "product_quantity": "10 units",
      "reason_for_recall": "The products may have been stored at temperatures outside their required storage conditions.",
      "recall_initiation_date": "20130716",
      "center_classification_date": "20130913",
      "termination_date": "20171227",
      "report_date": "20130925",
      "code_info": "Product Number: D76004;   Lots: All Lots;   Independence Medical Catalog Number: CJD76004",
      "more_code_info": ""
    }
  ]
}