{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Miami Lakes",
      "state": "FL",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88197",
      "recalling_firm": "Heartware, Inc.",
      "address_1": "14400 Nw 60th Ave",
      "address_2": "N/A",
      "postal_code": "33014-2807",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution.  US nationwide and Armenia, AUSTRIA, Bahrain, Belgium, CROATIA, CZECH REPUBLIC, Denmark, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, Hungary, Israel, ITALY, Jordan, Kazakhstan, Kuwait, Latvia, Lebanon, LITHUANIA, LUXEMBOURG, Macedonia, NETHERLANDS, Norway, POLAND, Qatar, Romania, SAUDI ARABIA, Serbia, Slovakia, SOUTH AFRICA, Spain, Sweden, SWITZERLAND, TAIWAN, Turkey, Ukraine, UNITED ARAB EMIRATES, and United Kingdom.",
      "recall_number": "Z-2194-2021",
      "product_description": "HeartWare HVAD Implant Kit, REF 1104JP",
      "product_quantity": "63 units",
      "reason_for_recall": "Medtronic is stopping the distribution and sale of the HVAD System due to a growing body of observational clinical comparisons that shows a higher frequency of neurological adverse events and mortality associated with the system when compared to other commercially available left ventricular assist devices, as well as complaints that the internal pump may delay or fail to restart.",
      "recall_initiation_date": "20210603",
      "center_classification_date": "20210806",
      "report_date": "20210818",
      "code_info": "GTIN:  00888707005838  All serial numbers"
    }
  ]
}