  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Glendora",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66173",
      "recalling_firm": "Ormco/Sybronendo",
      "address_1": "1332 S Lone Hill Ave",
      "address_2": "N/A",
      "postal_code": "91740-5339",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the countries of Canada, Australia, New Zealand, EU and ROW consignees.",
      "recall_number": "Z-2194-2013",
      "product_description": "Brand Name: Vector TAS 8mm Screw.    The Device Listing Number is E670535.    Vector T AS 8mm Screw is intended to provide a fixed anchorage point for attachment of orthodontic appliances to facilitate the orthodontic movement of teeth. It is used temporarily and removed after orthodontic treatment has been completed. It is surgically placed in the bone of the upper or lower jaw arches to provide support for orthodontic devices.",
      "product_quantity": "4,786",
      "reason_for_recall": "Ormco Corporation is voluntarily recalling one lot of 8 MM Vector TAS Screws because some screws of this lot were machined undersized at the top of the threads may potentially cause a weakening of the screw. Use of the undersized product may result in breakage of the screw during insertion or removal when torqued, possibly requiring surgical removal of any remaining pieces.",
      "recall_initiation_date": "20130712",
      "center_classification_date": "20130912",
      "termination_date": "20140911",
      "report_date": "20130918",
      "code_info": "Lot Number 112911: 8 MM Vector TAS Screw, Part Number 601-0021  Lot Number 112911: Vector TAS Starter Kit (10 Patient), Part Number 601-0010",
      "more_code_info": ""
    }
  ]
}