{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Irving",
      "address_1": "1915 Hurd Dr",
      "reason_for_recall": "There is a potential for 15% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results.",
      "address_2": "",
      "product_quantity": "4,727 units",
      "code_info": "Lot Numbers / Expiration Date    83459UN20 /09JUL2021,  03168UN20 /25SEP2021,  09787UN20 /24OCT2021,  37977UN20 /01JAN2022,  63000UN20 /01JAN2022,  77745UN20 /08JAN2022,  77743UN21 /02MAR2022,  09237UN21 /09APR2022",
      "center_classification_date": "20210802",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, MONTENEGRO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL, PR CHINA, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UNITED KINGDOM, VIETNAM.",
      "state": "TX",
      "product_description": "REF: 08P1820,  Abbott Alinity c Activated Alanine Aminotransferase reagent Kit, Alinity c Activated Alanine Aminotransferase Reagent Kit,, IVD, CE, UDI: (01) 00380740132569",
      "report_date": "20210811",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Abbott Laboratories",
      "recall_number": "Z-2184-2021",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "88123",
      "more_code_info": "",
      "recall_initiation_date": "20210610",
      "postal_code": "75038",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}