  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Fremont",
      "address_1": "47900 Bayside Pkwy",
      "reason_for_recall": "Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended  stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.",
      "address_2": "",
      "product_quantity": "45,302 (8,188 in the U.S.) - total, all model numbers",
      "code_info": "MODEL Number:M0035452560;    Lot Number/Expiration Date:  18426504 31-May-18  18780350 31-Jan-19  18426444 31-May-18  18532556 31-Aug-18  18794332 31-Jan-19  18830452 31-Jan-19  19036717 31-May-19  19005693 30-Apr-19  19060276 31-May-19  18939358 31-Mar-19  19056621 31-May-19  19037970 31-May-19  18971989 30-Apr-19  18906402 28-Feb-19  18953265 31-Mar-19  18897969 28-Feb-19",
      "center_classification_date": "20160711",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.",
      "state": "CA",
      "product_description": "Target Detachable Coils:  TARGET HELICAL NANO 3MM X 4CM  MODEL Number:M0035452560    Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Neurovascular",
      "recall_number": "Z-2184-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74425",
      "termination_date": "20170303",
      "more_code_info": "",
      "recall_initiation_date": "20160606",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}