      "country": "United States",
      "city": "Newark",
      "address_1": "500 GBC Drive, Mailstop 514",
      "reason_for_recall": "Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagent Management System (RMS) power input module can cause the plug to overheat and potentially create a fire hazard.  RMS without a power cord retention clip may be affected.",
      "address_2": "PO BOX 6101",
      "product_quantity": "2,554",
      "code_info": "Serial numbers 99060001 - 99062554",
      "center_classification_date": "20120809",
      "distribution_pattern": "Distributed in the US including Washington, DC,  Puerto Rico and the states of  AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WI, WV, and WY and the Virgin and Pacific Islands.",
      "state": "DE",
      "product_description": "Dimension(R) Reagent Management System (RMS) - Power Cord.    The Siemens Healthcare Diagnostics Dimension(R) clinical chemistry analyzers are floor model, fully automated, microprocessor-controlled, integrated instrument systems that use prepackaged Flex(R) reagent test cartridges to measure a variety of analytes in human body fluids.",
      "report_date": "20120815",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2182-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62611",
      "termination_date": "20140403",
      "more_code_info": "",
      "recall_initiation_date": "20120717",
      "postal_code": "19714-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}