{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Roswell",
      "address_1": "1400 Holcomb Bridge Rd",
      "reason_for_recall": "A defect of the \"fuse holder\" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke.",
      "address_2": "",
      "product_quantity": "129 units",
      "code_info": "Catalogue #01000-01 and 01000-01R, Units with 4-Digit Serial Numbers: Lot #4297, 4324, 4332, 4519, 4559, 4572; Units with 10 Character Serial Numbers: SA08017027, SA08017028, SA08017029, SA08017030, SA08017031, SA08017032, SA08017033, SA08017034, SA08017035, SA08017036, SA08017037, SA08017038, SA08017039, SA08017040, SA08017041, SA08017042, SA08017043, SA08017044, SA08017045, SA08017046, SA08017047, SA08017048, SA08017049, SA08343100, SA11137004, SA11137005, SA11137006, SA11137007, SA11137008, SA08094124, SA08094125, SA08094126, SA08094127, SA08094128, SA08094129, SA08094130, SA08094131, SA08094132, SA08094133, SA08094134, SA08094135, SA08094136, SA08094137, SA08094138, SA08094139, SA08094140, SA08094141, SA08094142, SA08094143, SA08094144, SA08094145, SA08094146, SA08094147, SA08094148, SA08094149, SA08094150, SA08094151, SA08094152, SA08094153, SA09020004, SA09020005, SA09020007, SA09020008, SA09020010, SA09020013, SA11298011, SA11298012, SA11298013, SA11298014, SA11298015, SA11298016, SA11298017, SA11298018, SA11298019, SA08136001, SA08136002, SA08136003, SA08136004, SA08136005, SA08136006,  SA08136007, SA08136008, SA08136009, SA08136010,   SA09043001, SA09043002, SA09043003, SA09043004, SA09043007, SA09043008, SA09043009, SA09043010, SA09043010, SA09043011, SA09043012, SA09043013, SA09043014, SA09043015, SA09043016, SA09043017, SA09043018, SA09043019, SA09043020, SA08343061, SA08343082, SA08343083, SA08343084, SA08343085, SA08343086, SA08343087, SA08343088, SA08343089, SA08343090, SA08343091,   SA08343092, SA08343093, SA08343094, SA08343095, SA08343096, SA10337006, SA10337007, SA10337008, SA10337009 and SA10337010.",
      "center_classification_date": "20120809",
      "distribution_pattern": "Nationwide Distribution-including the states of AZ, CA, CO, DE, FL, GA, IA, ID, IN, IL, KS, KY, LA, MA, MD, ME, MO, MS, NE, NM, OH, OR, PA, RI, SC, TN, TX, VA, WI and WV.",
      "state": "GA",
      "product_description": "Kimberly Clark Patient Warming System - Model 1000 Control Unit.     Kimberly Clark Patient Warming System - Model 1000 Control Unit and Energy Transfer Pads is a thermal regulating system, indicated for monitoring and controlling patient temperature.",
      "report_date": "20120815",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Kimberly-Clark Corporation",
      "recall_number": "Z-2181-2012",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "62596",
      "termination_date": "20120905",
      "more_code_info": "",
      "recall_initiation_date": "20120427",
      "postal_code": "30076",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}