      "country": "United States",
      "city": "Fremont",
      "address_1": "47900 Bayside Pkwy",
      "reason_for_recall": "Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended  stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.",
      "address_2": "",
      "product_quantity": "45,302 (8,188 in the U.S.) - total, all model numbers",
      "code_info": "MODEL Number:M0035443560:    Lot Number/Expiration Date:  18736425 31-Dec-18  18521457 31-Jul-18  18838703 31-Jan-19  18703233 30-Nov-18  18443288 30-Jun-18  18735209 30-Nov-18  18659056 31-Oct-18  18781251 31-Jan-19  18801665 31-Jan-19  18747459 31-Dec-18  18802980 31-Jan-19  18659087 31-Oct-18  18649087 31-Oct-18  18536525 31-Aug-18  19046694 31-May-19  19047725 31-May-19  18875709 28-Feb-19  18987089 30-Apr-19  19006438 30-Apr-19  18959391 31-Mar-19  18936943 31-Mar-19  19022830 30-Apr-19  18984771 30-Apr-19  18922696 31-Mar-19  18898321 28-Feb-19  19061363 31-May-19  18853805 28-Feb-19.",
      "center_classification_date": "20160711",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.",
      "state": "CA",
      "product_description": "Target Detachable Coils:  TARGET 360 NANO 3.5 MM X 6CM  MODEL Number:M0035443560      Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Neurovascular",
      "recall_number": "Z-2177-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74425",
      "termination_date": "20170303",
      "more_code_info": "",
      "recall_initiation_date": "20160606",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}