      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68727",
      "recalling_firm": "Carestream Health, Inc.",
      "address_1": "1049 W Ridge Rd",
      "address_2": "N/A",
      "postal_code": "14615-2731",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of  (United Kingdom, United Arab Emirates, Turkey, Taiwan, Switzerland, Sweden, Spain, South Africa, Portugal, Poland, Oman, New Zealand, Netherlands, Kuwait, Italy, Israel, Hong Kong, Germany, France, Finland, Denmark, Dubai, Czech Republic, China, Korea, Belgium, Australia, and Amsterdam.",
      "recall_number": "Z-2177-2014",
      "product_description": "Kodak DirectView DR 7500 Diagnostic X-Ray System     Product Usage:  The DR 7500 system is a permanently installed diagnostic x-ray system composed of 4 main components: an operator console, over head tube crane, x-ray tube assembly including a collimator and stationary generator. In addition the device can use an x-ray table and /or wallstand/bucky to complete x-ray exposures. The DR 7500 System is a permanently installed diagnostic x-ray system for generation of x-rays for examination of various anatomical regions.",
      "product_quantity": "589 systems",
      "reason_for_recall": "An on-site evaluation of the equipment parts that attach the Beta Assembly (x-ray tube and collimator) to the Over Head Tube Crane (OTC) Telescope was completed by Carestream Health (CSH) Service Engineering.   Inspection of these parts revealed that the tab on the lock washer was not bent into the lock nut recess and because the Beta Assembly detached from the OTC telescope, the lock nut was not sufficiently tightened.",
      "recall_initiation_date": "20140709",
      "center_classification_date": "20140809",
      "termination_date": "20170512",
      "report_date": "20140820",
      "code_info": "The device can be tracked according to service code.  The service code for the DR 7500 system is 8087."
    }
  ]
}