  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Heerbrugg",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64601",
      "recalling_firm": "Leica Microsystems (Schweiz) Ag",
      "address_1": "201 Max Schmidheiny-Strasse",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide distribution: USA (Nationwide) including states of: AK, CA, CO, NY, and OR; and countries of:  Belarus, China, Hong Kong, India, Indonesia, Italy, Norway, South Africa, Spain, Sweden, Turkey, and Vietnam.",
      "recall_number": "Z-2177-2013",
      "product_description": "Surgical microscope    The device is used to improve the visibility of objects through magnification and illumination and can be applied for observation and documentation and for human veterinary medical treatment.",
      "product_quantity": "28 units",
      "reason_for_recall": "The brake in the Z axis and/or brakes in the AB and C axes failed during installation.",
      "recall_initiation_date": "20121218",
      "center_classification_date": "20130910",
      "termination_date": "20170424",
      "report_date": "20130918",
      "code_info": "Serial Numbers: OH5-45912,   OH5-46012,   OH5-46112,   OH5-46212,   OH5-46312,   OH5-46412,   OH5-46512,   OH5-46612,   OH5-46712,   OH5-46912,   OH5-47012,   OH5-47112,   OH5-47212,   OH5-47312,   OH5-47412,   OH5-47512,   OH5-47612,   OH5-47812,   OH5-47912,   OH5-48012,   OH5-48112,   OH5-48212,   OH5-48312,   OH5-48312",
      "more_code_info": ""
    }
  ]
}