{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80094",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "N/A",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution including US states of CA, MD, NV, OH, TX, and WI, Afghanistan, Canada, Germany, Iraq, Italy, Korea, Kuwait, and Lebanon.",
      "recall_number": "Z-2176-2018",
      "product_description": "Brilliance 16 P with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728246. Computed Tomography X-ray system",
      "product_quantity": "20",
      "reason_for_recall": "The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore.  This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.",
      "recall_initiation_date": "20180330",
      "center_classification_date": "20180612",
      "termination_date": "20200414",
      "report_date": "20180620",
      "code_info": "System Serial Numbers: 5380, 5379, 5520, 5524, 5538, 5539, 5549, 5779, 5792, 5778, 5808, 5887, 5888, 5973, 6020, 6026, 6110, 6097, 6352, 6424"
    }
  ]
}