  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Fremont",
      "address_1": "47900 Bayside Pkwy",
      "reason_for_recall": "Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended  stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.",
      "address_2": "",
      "product_quantity": "45,302 (8,188 in the U.S.) - total, all model numbers",
      "code_info": "MODEL Number:M0035443040 :    Lot Number/Expiration Date:  18442838 30-Jun-18  18442388 30-Jun-18  18490600 31-Jul-18  18831276 31-Jan-19  18760116 31-Dec-18  18760119 31-Dec-18  18584035 30-Sep-18  18470827 31-Jul-18  18442358 30-Jun-18  18830247 31-Jan-19  18470767 30-Jun-18  18470857 31-Jul-18  18800019 31-Jan-19  18760117 31-Dec-18  18568317 31-Aug-18  18462748 30-Jun-18  18443108 30-Jun-18  18442178 30-Jun-18  18568224 31-Aug-18  18490538 31-Jul-18  18462838 30-Jun-18  18568193 31-Aug-18  18584004 30-Sep-18  18490662 31-Jul-18  18442688 30-Jun-18  19066241 31-May-19  19035290 31-May-19  18902944 28-Feb-19  18877825 28-Feb-19  18973721 30-Apr-19  18846664 31-Jan-19  19081717 31-May-19  19026562 30-Apr-19  18973029 30-Apr-19  18876957 28-Feb-19  18905865 28-Feb-19  19027086 30-Apr-19  19080897 31-May-19  19063669 31-May-19  18845906 31-Jan-19  18845527 31-Jan-19  19026038 30-Apr-19  18902302 28-Feb-19  18846285 31-Jan-19  19081102 31-May-19  18901660 28-Feb-19.",
      "center_classification_date": "20160711",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.",
      "state": "CA",
      "product_description": "Target Detachable Coils:  TARGET 360 NANO 3MM X 4CM  MODEL Number:M0035443040      Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Neurovascular",
      "recall_number": "Z-2175-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74425",
      "termination_date": "20170303",
      "more_code_info": "",
      "recall_initiation_date": "20160606",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}