  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 GBC Drive, Mailstop 514",
      "reason_for_recall": "Siemens has received complaints of customers experiencing slopes >1.05 when calibrating Flex(R) reagent cartridge lot 12095AD.  This results in a slope failure and prevents the VANC assay from being run on the Dimension Vista(R) System.",
      "address_2": "PO BOX 6101",
      "product_quantity": "1,894 cartons",
      "code_info": "lot 12095AD, exp 2013-04-04",
      "center_classification_date": "20120809",
      "distribution_pattern": "Worldwide Distribution -- US, including Washington, DC and the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WV, and the countries of Canada, New Zealand, Germany, and Puerto Rico.",
      "state": "DE",
      "product_description": "Dimension Vista(R) Vancomycin Flex(R) Reagent Cartridge (VANC, Catalog # K4086).    For the quantitative measurement of vancomycin in human serum and plasma on the Dimension Vista system.",
      "report_date": "20120815",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2175-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62609",
      "termination_date": "20140114",
      "more_code_info": "",
      "recall_initiation_date": "20120717",
      "postal_code": "19714-6101",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}