{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Round Lake",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88219",
      "recalling_firm": "BAXTER HEALTHCARE CORPORATION",
      "address_1": "25212 W Il Route 120",
      "address_2": "",
      "postal_code": "60073-9799",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the country of Japan.",
      "recall_number": "Z-2173-2021",
      "product_description": "Sharesource Connectivity Platform for Use with the Amia Automated PD System",
      "product_quantity": "1 unit",
      "reason_for_recall": "Baxter Healthcare Corporation has identified that the AMIA Sharesource User Guide that is used with the AMIA Automated PD System cyclers incorrectly describes the function of the UF (Ultrafiltration) Limit setting as the ending criteria of the cycler\u0019s Extra Last Drain Mode.  In some locations in the guide, it states that the UF volume achieved must exceed the programmed UF Limit before the cycler will end the Extra Last Drain, allowing the patient to end therapy or proceed to their last fill.",
      "recall_initiation_date": "20210630",
      "center_classification_date": "20210802",
      "termination_date": "20240410",
      "report_date": "20210811",
      "code_info": "Product Code 5C9400, UDI: 45413765565692, 55413765565699",
      "more_code_info": ""
    }
  ]
}