  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Minneapolis",
      "state": "MN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68838",
      "recalling_firm": "Medtronic Neuromodulation",
      "address_1": "7000 Central Ave Ne",
      "address_2": "N/A",
      "postal_code": "55432-3568",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) including the states of AL,AR,AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, Ml, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI and WV., and the countries of  Austria, Belgium, Cyprus, Denmark, France, Germany, Iraq, Italy, Lebanon, Netherlands, Panama, Poland, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, United Kingdom and Uruguay.",
      "recall_number": "Z-2172-2014",
      "product_description": "Medtronic Ascenda Intrathecal Catheter models 8780 and 8781, and Ascenda Intrathecal Catheter Pump Segment Revision Kit, model 8784.      The Ascenda Intrathecal Catheter is part of an infusion system that stores and delivers parenteral drugs to the intrathecal space.  The implanted infusion system components consist of a Medtronic pump and an Ascenda Intrathecal Catheter.",
      "product_quantity": "1200 (551 US, 649 OUS)",
      "reason_for_recall": "Medtronic is removing specific lots of the Ascenda Intrathecal Catheters and Revision Kits, which are used with the implantable SynchroMed drug infusion pump.  This recall is being conducted due to a single component of the catheter, the retainer ring, not meeting specification criteria.  There is a possibility of unintentional disconnection of the catheter from the pump, or difficulty in disconne",
      "recall_initiation_date": "20140711",
      "center_classification_date": "20140808",
      "termination_date": "20150717",
      "report_date": "20140820",
      "code_info": "Serial Numbers are in the following format:  \"N\" + 9 digits (e.g. N12345001.  Product beginning with the following letter/number combinations are affected:  N464184, N464200, N464206, N464254, N464257, N464258, N464262, N464263, N464264, N464271, N464693, N464703, N464708, N464725, N464727, N464730, N464734, N464736, N464738, N464742, N464746, N464750, N464752, N464841, N464924, N464930, N464995, N464995, N465020, N465038, N465063, N465065, N465073, N465082, N465088, N465091, N465094, N465111, N465125, N465129, N465131, N465136, N465138, N465144, N465145, N465146, N465148, N465150, N466065, N466067, N466067, N466068, N466069, N466071, N466072, N466075, N466134, N466167, N466177, N466196, N466201, N466240, N466244, N466259, N466289, N466290, N466291, N466306, N466368, N466390, N466392, N466394, N466398, N466399, N466631, N466639, N466645, N466678, N466678, N466754, N466758, N466761, N466886, N466897, N466901, N466902, N466911, N466914, N467103, N467109, N467113, N467114, N467116, N467126, N467138, N467171, N467180, N467183, N467192, N467193, N467194, N467195, N467197, N467198, N467203, N467204, N467206, N467209, N467351, N467351, N467352, N467355, N467356, N467357, N467395, N467399, N467401, N467402, N467407, N467444, N467451, N467453, N467454, N467563, N467568, N467570, N467601, N467606, N467608, N467609, N467611, N467612, N467614, N467618, N467621, N467626, N468288, N468343, N468401, N468404, N468406, N468406, N468448, N468494, N468500, N468507, N468510, N468511, N468514, N468516, N468519, N468522, N468574, N468580, N468621, N468622, N468624, N468626, N468641, N469182, N469262, N469332, N469360, N469360, N469388, N469420, N469422, N469428, N469431, N469433, N469434, N469436, N469438, N469440, N469446, N469447, N469449, N469452, N469460, N469465, N469466, N469469, N469474, N469478, N469507, N469511, N469530, N469533, N469549, N469579, N469583, N469586, N469588, N470163, N470197, N470234, N470303, N470307, N470401, N470408, N470414, N470420, N470424, N470453, N470459, N470919, N470923, N470929, N470931, N470931, N470934, N470938, N470940, N470941, N470951, N470953, N470954, N471442, N471443, N471446, N471447, N471448, N471449, N471487, N471488, N471490, N471490, N471491, N471764, N471791, N471840, N471853, N471856, N471860, N471863, N471864, N471865, N471866, N471870, N471871, N471875, N471880, N472310, N472326, N472333, N472336, N472342, N472345, N472347, N472857, N472861, N472866, N472868, N472873, N472875, N472876, N472878, N472882, N472885, N472886, N472890, N472894, N472900, N472905, N473140, N473144, N473148, N473154, N473156, N473176, N473178, N473180, N473202, N473207, N473211, N473221, N473222, N473223, N473228, N473230, N473311, N473311, N473323, N473377, N473401, N473426, N473438, N473492, N473493, N473989, N473996, N473999, N474001, N474005, N474006, N474007, N474010, N474018, N474019, N474023, N474027, N474037, N474038, N474047, N474047, N474421, N474423, N474427, N474431, N474433."
    }
  ]
}