{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marcy L'Etoile",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80111",
      "recalling_firm": "BioMerieux SA",
      "address_1": "Chemin De L'Orme",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution.",
      "recall_number": "Z-2168-2018",
      "product_description": "The VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic gram negative bacilli to antimicrobial agents.     Fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.",
      "product_quantity": "135 (78 boxes of various product) in total",
      "reason_for_recall": "A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.",
      "recall_initiation_date": "20180323",
      "center_classification_date": "20180612",
      "termination_date": "20200623",
      "report_date": "20180620",
      "code_info": "VITEK¿ 2 AST-N233 (ref: 413117): Lot Number 6330343103"
    }
  ]
}