  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Fremont",
      "address_1": "47900 Bayside Pkwy",
      "reason_for_recall": "Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended  stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.",
      "address_2": "",
      "product_quantity": "45,302 (8,188 in the U.S.) - total, all model numbers",
      "code_info": "MODEL Number: M0035431510;    Lot Number/Expiration Date:  18581295 30-Sep-18  18800256 31-Jan-19  18775373 31-Dec-18  18744204 30-Nov-18  18812788 31-Jan-19  18607578 30-Sep-18  18602655 30-Sep-18  19003141 30-Apr-19  18935435 31-Mar-19  18923481 31-Mar-19  18953907 31-Mar-19  18883412 28-Feb-19  18852906 28-Feb-19  19056331 31-May-19.",
      "center_classification_date": "20160711",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.",
      "state": "CA",
      "product_description": "Target Detachable Coils:  TARGET HELICAL NANO 1.5 MM X 1 CM  MODEL Number:M0035431510    Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Neurovascular",
      "recall_number": "Z-2168-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74425",
      "termination_date": "20170303",
      "more_code_info": "",
      "recall_initiation_date": "20160606",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}