{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Fremont",
      "address_1": "47900 Bayside Pkwy",
      "reason_for_recall": "Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended  stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.",
      "address_2": "",
      "product_quantity": "45,302 (8,188 in the U.S.) - total, all model numbers",
      "code_info": "MODEL Number: M0035431030;    Lot Number/Expiration Date:  18801491 31-Jan-19  18757273 31-Dec-18  18709428 30-Nov-18  18602133 30-Sep-18  18523870 31-Jul-18  18802806 31-Jan-19  18664601 31-Oct-18  18569137 31-Aug-18  18517839 31-Jul-18  18486090 31-Jul-18  18842022 31-Jan-19  18651923 31-Oct-18  18651865 31-Oct-18  18607288 30-Sep-18  18518071 31-Jul-18  18774611 31-Dec-18  18762227 31-Dec-18  18637042 31-Oct-18  18564861 31-Aug-18  18843756 28-Feb-19  18842889 31-Jan-19  18781019 31-Jan-19  18651981 31-Oct-18  18546861 31-Aug-18  18447942 30-Jun-18  18830688 31-Jan-19  18815198 31-Jan-19  18814328 31-Jan-19  18736978 30-Nov-18  18592210 30-Sep-18  18517955 31-Jul-18  18414488 31-May-18  18829659 31-Jan-19  18583312 31-Aug-18  18984903 30-Apr-19  19050397 31-May-19  18883354 28-Feb-19  18911821 31-Mar-19  19041285 30-Apr-19  18944244 31-Mar-19  18922785 31-Mar-19  18902333 28-Feb-19  18911418 31-Mar-19  18857263 28-Feb-19  18973207 30-Apr-19  19041807 30-Apr-19  19065916 31-May-19  19063344 31-May-19  18901691 28-Feb-19  19074243 31-May-19  19049355 31-May-19  19042329 31-May-19  19005113 30-Apr-19  18944767 31-Mar-19  18973553 30-Apr-19  18943198 31-Mar-19  18943721 31-Mar-19  19042851 31-May-19  18855184 28-Feb-19  19048313 31-May-19  18972861 30-Apr-19  18953849 31-Mar-19  18912224 31-Mar-19  19005171 30-Apr-19  18902975 28-Feb-19  18912627 31-Mar-19.",
      "center_classification_date": "20160711",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.",
      "state": "CA",
      "product_description": "Target Detachable Coils:  TARGET HELICAL NANO 1 MM X 3 CM  MODEL Number:M0035431030        Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Neurovascular",
      "recall_number": "Z-2167-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74425",
      "termination_date": "20170303",
      "more_code_info": "",
      "recall_initiation_date": "20160606",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}