  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "South Jordan",
      "state": "UT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88191",
      "recalling_firm": "Merit Medical Systems, Inc.",
      "address_1": "1600 W Merit Pkwy",
      "address_2": "N/A",
      "postal_code": "84095-2416",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - U.S. Nationwide distribution in the states of AL, AZ, CA, CO, DE, FL, GA, ID, IL IN, KS, LA, MI, MN, MO, NC, NE, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, WA, WI, WV, and WY. The countries of Australia, Austria, China, Denmark, Finland, France, Germany, Israel, Japan, Kenya, Kyrgyzstan, Netherlands, New Zealand, Norway, Qatar, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates and United Kingdom.",
      "recall_number": "Z-2166-2021",
      "product_description": "basixALPHA Inflation Device 30 atm/bar 20 mL - Product Usage: Used to inflate and deflate an angioplasty balloon or other interventional device, and to measure the pressure within the balloon.",
      "product_quantity": "7,211 devices",
      "reason_for_recall": "Due to a design issue, inflation syringe handle separating from the inflation device when handle is withdrawn.",
      "recall_initiation_date": "20210610",
      "center_classification_date": "20210729",
      "termination_date": "20240515",
      "report_date": "20210804",
      "code_info": "Catalog Number: IN3130  UDI:  00884450440179;  00884450510155;  Lot Numbers:  H1862928;  H1896261;  H1923761;  H1929106;  H1939255;  H1966485;  H1978295;  H2068264;  H2073193;  H2073247;  H2077173;    Catalog Number: IN3152  UDI: 00884450511220  Lot Numbers:  H1896262;  H1909581;  H1923762;  H1929107;  H1978299;  H2007415;  H2031599;  H2068265;  H2073248;  H2077175;  H2084368;    Catalog Number: IN3302  UDI: 00884450511305  Lot Numbers:  H1896263;  H1966489;    Catalog Number: IN3802  UDI: 00884450511510  Lot Numbers:  H18962650;  H1939259;  H1966490;  H1973390;  H1978305;  H1993118;  H2001426;  H2007417;  H2012534;  H2031601;  H2073198;  H2073249;  H2077176;"
    }
  ]
}