  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Marcy L'Etoile",
      "state": "N/A",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80111",
      "recalling_firm": "BioMerieux SA",
      "address_1": "Chemin De L'Orme",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "International distribution.",
      "recall_number": "Z-2166-2018",
      "product_description": "The VITEK¿ 2 Anaerobic and Corynebacteria identification card (ANC) is intended for use with the VITEK¿ 2 Systems for the automated identification of most clinically significant anaerobic organisms and Corynebatcerium species. The VITEK¿ 2 ANC identification card is a single-use disposable.     A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated with test reagents, that are usually contained in individual compartments and used to differentiate and identify selected microorganisms. The device aids in the diagnosis of disease.",
      "product_quantity": "135 (78 boxes of various product) in total",
      "reason_for_recall": "A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.",
      "recall_initiation_date": "20180323",
      "center_classification_date": "20180612",
      "termination_date": "20200623",
      "report_date": "20180620",
      "code_info": "VITEK¿ 2 ANC (ref: 21347): Lot Number 2440318203"
    }
  ]
}