{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Fremont",
      "address_1": "47900 Bayside Pkwy",
      "reason_for_recall": "Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended  stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.",
      "address_2": "",
      "product_quantity": "45,302 (8,188 in the U.S.) - total, all model numbers",
      "code_info": "MODEL Number: M0035421540    Lot Number/Expiration Date:  18808912 31-Jan-19  18741434 31-Dec-18  18714072 30-Nov-18  18677229 30-Nov-18  18625483 31-Oct-18  18625541 31-Oct-18  18642784 31-Oct-18  18633921 31-Oct-18  18598299 30-Sep-18  18590470 30-Sep-18  18542363 31-Aug-18  18420011 31-May-18  18419897 31-May-18  18833256 28-Feb-19  18808618 31-Jan-19  18787244 31-Jan-19  18677171 30-Nov-18  18524564 31-Jul-18  18524622 31-Jul-18  18832645 31-Jan-19  18768683 31-Dec-18  18756119 31-Dec-18  18725503 30-Nov-18  18666721 31-Oct-18  18642842 31-Oct-18  18642668 31-Oct-18  18598183 30-Sep-18  18590528 30-Sep-18  18578953 30-Sep-18  18406806 31-May-18  18809794 31-Jan-19  18787859 31-Jan-19  18755053 31-Dec-18  18741144 31-Dec-18  18701144 30-Nov-18  18590586 30-Sep-18  18579069 30-Sep-18  18549270 31-Aug-18  18420182 31-May-18  18420239 31-May-18  18809206 31-Jan-19  18817972 28-Feb-19  18701318 30-Nov-18  18677113 30-Nov-18  18488340 31-Jul-18  18598125 30-Sep-18  18542305 31-Aug-18  18787449 31-Jan-19  18714246 30-Nov-18  18625599 31-Oct-18  18634037 31-Oct-18  18598241 30-Sep-18  18559192 31-Aug-18  18832034 31-Jan-19  18817309 28-Feb-19  18787654 31-Jan-19  18768041 31-Dec-18  18701260 30-Nov-18  18809500 31-Jan-19  18701202 30-Nov-18  18677287 30-Nov-18  18642726 31-Oct-18  18590412 30-Sep-18  18559250 31-Aug-18  19080127 31-May-19  19010580 30-Apr-19  18931017 31-Mar-19  18885811 28-Feb-19  18846227 28-Feb-19  19026877 30-Apr-19  19025829 30-Apr-19  19043083 31-May-19  18974705 30-Apr-19  18957691 31-Mar-19  18911592 31-Mar-19  18845848 28-Feb-19  19049933 31-May-19  18998580 30-Apr-19  18853712 28-Feb-19  18887225 28-Feb-19  19080305 31-May-19  19026353 30-Apr-19  18998638 30-Apr-19  18975142 30-Apr-19  18956531 31-Mar-19  19079949 31-May-19  18964774 31-Mar-19  19042039 30-Apr-19  19010180 30-Apr-19  18963382 31-Mar-19  18912801 31-Mar-19  18911995 31-Mar-19  18846606 28-Feb-19  18963846 31-Mar-19  18930672 31-Mar-19  18899164 28-Feb-19  18912398 31-Mar-19  18974268 30-Apr-19  18987031 30-Apr-19  19050975 31-May-19  19048891 31-May-19  19042561 31-May-19  19041517 30-Apr-19  18964310 31-Mar-19  18975579 30-Apr-19  18957111 31-Mar-19  18940339 31-Mar-19  18938917 31-Mar-19  18929982 31-Mar-19  18845469 28-Feb-19.",
      "center_classification_date": "20160711",
      "distribution_pattern": "Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.",
      "state": "CA",
      "product_description": "Target Detachable Coils:  TARGET 360 NANO 1.5 MM X 4 CM  MODEL Number:M0035421540    Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Neurovascular",
      "recall_number": "Z-2164-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74425",
      "termination_date": "20170303",
      "more_code_info": "",
      "recall_initiation_date": "20160606",
      "postal_code": "94538-6515",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}