  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Westlake",
      "address_1": "810 Sharon Dr",
      "reason_for_recall": "A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5.   If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission.   If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t",
      "address_2": "",
      "product_quantity": "1427 ABL 90 Instruments",
      "code_info": "Part Number: 393-090, all production runs, 393-090R0026N0010 to 393-090R0179N010",
      "center_classification_date": "20120808",
      "distribution_pattern": "Worldwide distribution: USA (nationwide) including the states of AK, AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Canada, Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cuba, Cyprus, Czech republic, Denmark, Ecuador, Egypt, Eritrea, Estonia, Finland, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Malta, Mauritius, Mexico, Moldova, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Peru, Philippines, Poland, Puerto Rico, Qatar, Republic of Korea, Romania, Russia, Saudi Arabia, Switzerland, Singapore, Slovenia, South Africa, Spain, Sweden Syria, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Yugoslavia, and Zimbabwe.",
      "state": "OH",
      "product_description": "ABL90 FLEX software version 2.7 MR3.  The ABL90 FLEX analyzer is intended for in vitro testing of samples of heparinized whole blood.",
      "report_date": "20120815",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Radiometer America Inc",
      "recall_number": "Z-2164-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62503",
      "termination_date": "20130820",
      "more_code_info": "",
      "recall_initiation_date": "20120705",
      "postal_code": "44145-1521",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}