{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rastatt",
      "state": "N/A",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80076",
      "recalling_firm": "Maquet Cardiopulmonary Ag",
      "address_1": "Kehler Strasse 31",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution including US state of VA and Puerto Rico,  Australia, Belgium, Brazil, Canada, Chile, Colombia, Cuba, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, Iran, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Martinique, Mauritius, Mexico, Netherlands, New Zealand, Norway, Pakistan, Poland, Portugal, Reunion, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-2160-2018",
      "product_description": "Aortic Arterial Cannula 22 Fr (O.D.); 23 cm (L); with 3/8-inch connector with LL,  Part Number 701002246, Article Number A22-7107.  For use in cardiopulmonary bypass",
      "product_quantity": "470 US",
      "reason_for_recall": "This lot of arterial cannula may have been supplied with 3/8-width de-airing plugs that could break during the process of removal from the arterial cannula.",
      "recall_initiation_date": "20180514",
      "center_classification_date": "20180611",
      "termination_date": "20200515",
      "report_date": "20180620",
      "code_info": "UDI CODE: 04037691123813, Lot Number: 92181729 and 92181287"
    }
  ]
}