  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oxford",
      "state": "CT",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "66024",
      "recalling_firm": "Villa Radiology Systems LLC",
      "address_1": "91 Willenbrock Rd Ste B1",
      "address_2": "N/A",
      "postal_code": "06478-1036",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution including the states of AZ, MI, IN, UT, CT, AL, NY, PA, IL, MD, CA, MA,  IA, OH, MS and MO.",
      "recall_number": "Z-2160-2013",
      "product_description": "Juno DRF System. Mobile X-ray system    These are general, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures",
      "product_quantity": "22 units within US",
      "reason_for_recall": "It has been discovered that the Air Kerma rate information is inconsistent in Continuous Fluoro modes 30x30 (12\"x12\") and 15x15 (6\"x6). This could result in a deviation of the displayed values of Air Kerma Rate and Cumulated Air Kerma from the actual values greater then +/-35% and then in a failure to comply with the requirements of 21 CFR 1020.32(k)(6",
      "recall_initiation_date": "20130802",
      "center_classification_date": "20130913",
      "termination_date": "20160725",
      "report_date": "20130925",
      "code_info": "PN 709020",
      "more_code_info": ""
    }
  ]
}