  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rancho Santa Margarita",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85436",
      "recalling_firm": "Applied Medical Resources Corp",
      "address_1": "22872 Avenida Empresa",
      "address_2": "N/A",
      "postal_code": "92688-2650",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US:  AL, AR, AZ, CA, CO, CT, FL, GA, HI,  IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA,  PR, RI,  SC, SD, TN, TX, UT, VA, WA, WI, WV, WY  OUS:  Argentina, Australia, Austria, Belgium, Canada, Chile, Costa Rica, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Latvia, Liechtenstein, Luxembourg, Netherlands, New Zealand, Peru, Portugal, Serbia, Singapore, South Africa, Spain, Thailand, United Kingdom",
      "recall_number": "Z-2158-2020",
      "product_description": "Epix latis GRASPER, 5 mm x 35 cm, REF C4130, Qty:  10, Sterile R, CE, Rx ONLY, UDI:  (01)00607915110147",
      "product_quantity": "32,760 units",
      "reason_for_recall": "There may be a slightly protruding rivet on the distal end of the laparoscopic grasper which has the potential for tissue to catch on it during use and lead to tissue damage.",
      "recall_initiation_date": "20200317",
      "center_classification_date": "20200528",
      "termination_date": "20220217",
      "report_date": "20200603",
      "code_info": "Batch:  1364578/Expiration Date:  July 28, 2022;   Batch:  1366453/Expiration Date:  August 13, 2022;    Batch: 1371249/Expiration Date: October 3, 2022;   Batch: 1373050/Expiration Date:  October 23, 2022;   Batch: 1373248/Expiration Date:  November 4, 2022;   Batch:  1373249/Expiration Date:  November 11, 2022;  Batch: 1373255/Expiration Date:  November 18, 2022;   Batch: 1373260/Expiration Date:  November 19, 2022;   Batch; 1373261/Expiration Date:  November 24, 2022;   Batch: 1373262/Expiration Date:  December 2, 2022;   Batch: 1373263/Expiration Date:  December 5, 2022;   Batch: 1377363/Expiration Date:  December 11, 2022;   Batch: 1377364/Expiration Date:  December 17, 2022"
    }
  ]
}