      "status": "Terminated",
      "city": "Clearwater",
      "state": "FL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71411",
      "recalling_firm": "Bausch & Lomb Surgical, Inc.",
      "address_1": "21 N Park Place Blvd",
      "address_2": "N/A",
      "postal_code": "33759-3917",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MD, MN, MO, MS, NC, NH, NJ, NM, NV, NY,  OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WV, Hawaii and Puerto Rico and countries of: Canada, Brazil, Argentina, Austria, Belgium, Switzerland, Germany, Spain, France, Great Britain, Georgia, Greece, Kyrgyzstan, Norway, Russia, Ukraine, United Arab Emirates, China, Australia, India Malaysia, Korea, and Taiwan.",
      "recall_number": "Z-2154-2015",
      "product_description": "Bausch & Lomb SoFlex SE Foldable intraocular Lens    Product Usage:  The SofPort/SoFlex IOLs are intended for primary implantation in the capsular bag of the eye for the visual correction of aphakia secondary to the removal of a cataractous lens in adult patients.",
      "product_quantity": "72,773 units; 52,787 units (US) 19,986 units (Foreign)",
      "reason_for_recall": "Haptics breaking during lens loading and insertion.",
      "recall_initiation_date": "20150526",
      "center_classification_date": "20150720",
      "termination_date": "20160825",
      "report_date": "20150729",
      "code_info": "Produce Code:  LI61AO and LI61SE.   LOT #'s 4455501, 4455723, 4456117, 4456118, 4456123, 4456207, 4456213, 4456214, 4456306, 4456316, 4456318, 4456319, 4456403, 4456406, 4456412, 4456413, 4456414, 4456415, 4456416, 4456703, 4456704, 4456707, 4456710,   4456711, 4456712, 4456713, 4456714, 4456715, 4456716, 4456717, 4456718,   4456719, 4456720, 4456721, 4456722, 4456723, 4456724, 4456725, 4456726,   4456727, 4456801, 4456803, 4456804, 4456805, 4456807, 4456808, 4456809,   4456810, 4456811, 4456812, 4456813, 4456814, 4456815, 4456816, 4456817,   4456818, 4456819, 4456820, 4456821, 4456822, 4456823, 4456824, 4456825,   4456826, 4456827, 4456828, 4456829, 4456830, 4456831, 4456832, 4456901,  4456902, 4456903, 4456904, 4456905, 4456906, 4456907, 4456908, 4456909,   4456910, 4456911, 4456912, 4457001, 4457002, 4457003, 4457004, 4457005,   4457006, 4457007, 4457008, 4457009, 4457010, 4457011, 4457012, 4457013,   4457014, 4457015, 4457016, 4457017, 4457018, 4457019, 4457020, 4457021,   4457022, 4457101, 4457102, 4457103, 4457104, 4457105, 4457106, 4457107,   4457108, 4457109, 4457110, 4457111, 4457401, 4457402, 4457403, 4457404, 4457406, 4457407, 4457408, 4457409, 4457410, 4457411, 4457412, 4457413,  4457414, 4457415, 4457416, 4457417, 4457418, 4457419, 4457420, 4457421,  4457422, 4457423, 4457424, 4457426, 4457427, 4457428, 4457501, 4457502,  4457503, 4457504, 4457505, 4457506, 4457507, 4457508, 4457509, 4457510,  4457511, 4457512, 4457513, 4457514, 4457515, 4457516, 4457517, 4457518,  4457519, 4457520, 4457521, 4457522, 4457523, 4457601, 4457602, 4457603,  4457604, 4457606, 4457607, 4457608, 4457609, 4457610, 4457611, 4457612,  4457613, 4457614, 4457615, 4457616, 4457617, 4457618, 4457619, 4457620,   4457701  4457702  4457703  4457704  4457705  4457706  4457707  4457708  4457709  4457710  4457711  4457712  4457713  4457714  4457715  4457716  4457717  4457718  4457719  4457720  4457721  4457722  4457723  4457724  4457725  4457726  4457727  4457801  4457802  4457803  4457804  4457805  4457806  4457807  4457808  4457809  4457810  4457811  4457812  4457813  4457814  4457815  4457816  4457817  4457818  4457819  4457820  4457821  4457822  4458101  4458102  4458103  4458104  4458105  4458106  4458107  4458108  4458109  4458110  4458111  4458112  4458113  4458114  4458115  4458116  4458201  4458202  4458203  4458204  4458205  4458206  4458207  4458208  4458209  4458210  4458211  4458212  4458213  4458214  4458215  4458216  4458217  4458218  4458219  4458220  4458221  4458222  4458223  4458224  4458226  4458227  4458228  4458229  4458230  4458231  4458232  4458233  4458234  4458236  4458237  4458238  4458239  4458241  4458242  4458243  4458244  4458245  4458246  4458247  4458301  4458302  4458303  4458304  4458305  4458306  4458307  4458308  4458312  4458313  4458314  4458316  4458317  4458821  4458909  4459202  4459203  4459205  4459212  4459501  4459513  4459520  4459620  4460228  4460231  4460234  4460235  4460413  4460601  4460602  4460603  4460604  4460605  4460606  4460607  4460608  4460609  4460610  4461001  4461002  4461003  4461004  4461005  4461006  4461007  4461008  4461009  4461010  4461011  4461012  4461013  4461112  4461113  4461202  4461208  4461212  4461214  4461218  4461220  4461221  4461222  4461223  4461224  4461227  4461231  4461233  4461234  4461235  4461238  4461239  4461240  4461304  4461305  4461308  4461309  4461310  4461311  4461312  4461313  4461314  4461315  4461316  4461317  4461318  4461319  4461601  4461602  4461603  4461604  4461605  4461606  4461607  4461608  4461609  4461610  4461611  4461612  4461613  4461614  4461615  4461616  4461617  4461618  4461619  4461620  4461621  4461622  4461701  4461702  4461703  4461704  4461705  4461706  4461707  4461708  4461709  4461710  4461711  4461712  4461713  4461714  4461715  4461716  4461717  4461718  4461719  4461720  4461721  4461722  4461723  4461724  4461801  4461802  4461803  4461804  4461901  4461902  4461903  4461904  4461905  4461906  4461907  4461908  4461909  4461910  4461911  4461912  4461913  4461914  4461915  4462001  4462002  4462003  4462004  4462005  4462006  4462007  4462008  4462009  4462010  4462011  4462012  4462013  4462014  4462015  4462016  4462017  4462018  4462019  4462020  4462021  4462301  4462302  4462303  4462304  4462305  4462306  4462307  4462308  4462309  4462310  4462311  4462312  4462313  4462314  4462315  4462316  4462317  4462318  4462319  4462320  4462321  4462322  4462401  4462402  4462403  4462404  4462405  4462406  4462407  4462408  4462409  4462410  4462411  4462412  4462413  4462414  4462415  4462416  4462417  4462418  4462419  4462420  4462421  4462422  4462423  4462501  4462502  4462503  4462504  4462505  4462506  4462507  4462508  4462509  4462510  4462511  4462512  4462513  4462514  4462515  4462516  4462517  4462518  4462701  4462702  4462703  4462704  4462705  4462706  4462708  4462709  4462710  4462711  4462713  4462714  4462715  4462716  4462717  4462718  4462719  4462720  4462721  4462722  4462723  4462724  4462725  4462726  4462727  4463001  4463002  4463003  4463004  4463005  4463006  4463007  4463008  4463009  4463010  4463011  4463012  4463013  4463014  4463015  4463016  4463017  4463018  4463019  4463020  4463021  4463022  4463023  4463024  4463025  4463026  4463101  4463102  4463103  4463104  4463105  4463106  4463107  4463108  4463109  4463110  4463111  4463112  4463113  4463114  4463115  4463116  4463117  4463118  4463201  4463202  4463203  4463204  4463205  4463206  4463207  4463208  4463209  4463210  4463211  4463212  4463213  4463214  4463215  4463216  4463217  4463218  4463219  4463220  4463221  4463222  4463223  4463224  4463225  4463226  4463227  4463228  4463229  4463230  4463231  4463232  4463233  4463234  4463235  4463236  4463237  4463238  4463239  4463240  4463241  4463242  4463243  4463244  4463245  4463246  4463247  4463248  4463249  4463701  4463702  4463703  4463704  4463705  4463706  4463707  4463708  4463709  4463710  4463711  4463712  4463713  4463714  4463715  4463716  4463717  4463718  4463719  4463720  4463721  4463722  4463801  4463802  4463803  4463804  4463805  4463806  4463808  4463809  4463810  4463811  4463812  4463813  4463814  4463815  4463816  4463817  4463819  4463822  4463823  4463824  4463825  4463826  4463827  4463828  4463829  4463901  4463902  4463904  4463905  4463906  4463909  4463910  4463911  4463912  4463913  4463914  4463916  4463917  4463919  4463920  4463921  4463922  4463923  4464001  4464002  4464003  4464004  4464005  4464006  4464007  4464008  4464010  4464011  4464012  4464013  4464014  4464015  4464016  4464017  4464018  4464019  4464020  4464021  4464022  4464023  4464024  4464025  4464026  4464101  4464102  4464103  4464104  4464105  4464106  4464107  4464108  4464109  4464110  4464111  4464112  4464113  4464114  4464115  4464116  4464117  4464118  4464119  4464120  4464121  4464122  4464123  4464124  4464125  4464126  4464127  4464128  4464501  4464502  4464503  4464504  4464505  4464506  4464507  4464508  4464509  4464510  4464511  4464512  4464601  4464602  4464603  4464604  4464605  4464606  4464607  4464608  4464609  4464610  4464611  4464612  4464613  4464614  4464615  4464616  4464617  4464618  4464619  4464620  4464621  4464622  4464623  4464624  4464625  4464626  4464627  4464628  4464629  4464630  4464631  4464632  4464633  4464634  4464636  4464637  4464638  4464639  4464640  4464641  4464642  4464643  4464644  4464645  4464646  4464647  4464648  4464649  4464701  4464702  4464703  4464704  4464705  4464706  4464707  4464708  4464801  4464802  4464803  4464804  4464805  4464806  4464807  4464808  4464809  4464810  4464811  4464812  4464813  4465101  4465102  4465103  4465104  4465105  4465106  4465107  4465108  4465109  4465110  4465111  4465112  4465113  4465114  4465115  4465116  4465117  4465118  4465119  4465120  4465121  4465122  4465123  4465124  4465125  4465126  4465127  4465128  4465129  4465130  4465201  4465202  4465203  4465204  4465205  4465206  4465207  4465208  4465209  4465210  4465211  4465212  4465213  4465214  4465215  4465216  4465301  4465302  4465303  4465304  4465305  4465306  4465307  4465308  4465309  4465310  4465311  4465312  4465313  4465314  4465315  4465316  4465401  4465402  4465403  4465404  4465405  4465406  4465407  4465408  4465409  4465410  4465411  4465412  4465414  4465415  4465416  4465417  4465418  4465419  4465420  4465421  4465422  4465423  4465424  4465425  4465426  4465427  4465428  4465429  4465430  4465501  4465502  4465503  4465504  4465505  4465506  4465507  4465508  4465509  4465510  4465511  4465512  4465513  4465514  4465515  4465516  4465518  4465520  4465523  4465524  4465525  4465801  4465802  4465803  4465805  4465806  4465808  4465809  4465810  4465811  4465812  4465813  4465814  4465815  4465816  4465817  4465818  4465820  4465821  4465822  4465823  4465824  4465825  4465826  4465827  4465901  4465902  4465903  4465904  4465905  4465906  4465907  4465908  4465909  4465910  4465911  4465913  4465914  4465915  4465916  4466001  4466002  4466004  4466005  4466007  4466008  4466010  4466011  4466012  4466013  4466014  4466016  4466017  4466101  4466102  4466103  4466104  4466105  4466106  4466107  4466109  4466110  4466112  4466113  4466116  4466121  4466122  4466134  4466136  4466137  4466138  4466140  4466143  4466144  4466145  4466148  4466149  4466150  4466151  4466153  4466157  4466159  4466161  4466167  4466168"
    }
  ]
}