  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Erlangen",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76808",
      "recalling_firm": "Hans Pausch Rontgengeratebau Gmbh",
      "address_1": "Graf Zeppelin Strasse 1",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "FDA Mandated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution",
      "recall_number": "Z-2152-2017",
      "product_description": "Pausch Uroview FD fluoroscopic X-ray system  UDI number (01)04260491620040  Primarily for urological applications (functional X-ray diagnostics, endourology and minimal invasive urology / surgery). The system, which includes a radiologic / urologic treatment table, maybe used for urological treatment, planning and diagnostic procedures, including but not limited to.",
      "product_quantity": "10 systems",
      "reason_for_recall": "During patient interventions with the urology table the radiation is activated with a footswitch connected to the table. This footswitch is actively involved in the safety circuit of the urological table. Only if footswitch + exposure + first primary barrier are closed the signal will be transmitted to the generator for activating   X-ray. During intended use only with this footswitch X-ray should be activated.",
      "recall_initiation_date": "20170314",
      "center_classification_date": "20170530",
      "termination_date": "20190614",
      "report_date": "20170607",
      "code_info": "model numbers: 01169033 and 01169034 with the serial numbers 0103-0112",
      "more_code_info": ""
    }
  ]
}