  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Parsippany",
      "state": "NJ",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "70961",
      "recalling_firm": "Biomet Spine, LLC",
      "address_1": "399 Jefferson Rd",
      "address_2": "N/A",
      "postal_code": "07054-3707",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution US Nationwide (AL, CA, CO, FL, GA, LA, MA, MD, MI, MO, NY, OH, PA, TX, WA, WI) and The Netherlands.",
      "recall_number": "Z-2152-2015",
      "product_description": "Polaris 4.75 TI Spinal System, 4.0 x 20 MM, Closed MAS, Closed Head Multiaxial Screw, Non-Sterile, Rx Only, REF 14-583020, Biomet Spine and Bone Healing Technologies.  Spinal fixation component.",
      "product_quantity": "4,765 units",
      "reason_for_recall": "Multiaxial screws may not meet internal requirements related to fatigue strength.  Compression testing result and surface treatment are out of specification.",
      "recall_initiation_date": "20150312",
      "center_classification_date": "20150720",
      "termination_date": "20151221",
      "report_date": "20150729",
      "code_info": "Part Numbers:  14-582020, 14-582022, 14-582025, 14-582030, 14-582035, 14-582040, 14-582120, 14-582122, 14-582125, 14-582127, 14-582130, 14-582135, 14-582140, 14-582220, 14-582222, 14-582225, 14-582230, 14-582235, 14-582240, 14-582245, 14-582250, 14-592320, 14-582325, 14-582330, 14-582335, 14-582340, 14-582345, 14-582350, 14-582530, 14-582535, 14-582540, 14-582545, 14-582550, 14-582565, 14-582570, 14-582575, 14-583020, 14-583025, 14-583030, 14-583035, 14-583040, 14-583120, 14-583125, 14-583130, 14-583135, 14-583140, 14-583220, 14-583225, 14-583230, 14-583235, 14-583240, 14-583245, 14-583250, 14-583320, 14-583325, 14-583330, 14-583335, 14-583340, 14-583345 and 14-583350.  All lots."
    }
  ]
}