  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "American Fork",
      "address_1": "752 E 1180 S Ste 200",
      "reason_for_recall": "Potential for images to be flipped while streaming, which could incorrectly display image orientation markers.",
      "address_2": "",
      "product_quantity": "2695",
      "code_info": "All NovaPAC software versions 7.4, 7.5, 7.6, and 8.0",
      "center_classification_date": "20160711",
      "distribution_pattern": "Worldwide distribution including US nationwide. No Canadian distribution.",
      "state": "UT",
      "product_description": "NovaPACS \u0013 versions 7.4, 7.5, 7.6, and 8.0, Diagnostic Viewer. Distributed by Novarad Corporation. Picture archiving and communication system (PACS)",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Novarad Corporation",
      "recall_number": "Z-2151-2016",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "74556",
      "termination_date": "20161201",
      "more_code_info": "",
      "recall_initiation_date": "20130509",
      "postal_code": "84003-3561",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}