{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Mahwah",
      "state": "NJ",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92287",
      "recalling_firm": "Datascope Corp.",
      "address_1": "1300 Macarthur Blvd",
      "address_2": "N/A",
      "postal_code": "07430-2052",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution: US (Nationwide) and OUS (foreign) to countries of: ALBANIA ALGERIA, ANGOLA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELARUS, BELGIUM, BOSNIA and HERZEGOVINA, BRAZIL, BRUNEI, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, C¿te d'Ivoire, CROATIA, CZECH REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, ESTONIA, FINLAND, FRANCE, GERMANY, GHANA, GUATEMALA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRAQ, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, KOSOVO, KUWAIT, LATVIA, LEBANON, LIBYA, LITHUANIA, LUXEMBOURG, MACEDONIA, MALAYSIA, MEXICO, MOROCCO, MYANMAR, NETHERLANDS, NEW ZEALAND, NIGERIA, NORWAY, OMAN, PAKISTAN, PALESTINE, PERU, PHILIPPINES, POLAND, PORTUGAL, QATAR, RUSSIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SRI LANKA, SWEDEN, SWITZERLAND, SYRIA, TAIWAN, TANZANIA, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, TURKMENISTAN, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, UZBEKISTAN, VIETNAM, YEMEN, ZAMBIA",
      "recall_number": "Z-2150-2023",
      "product_description": "Cardiosave Rescue Intra-Aortic Balloon Pump (IABP), Model numbers   0998-00-0800-75  0998-00-0800-83  0998-00-0800-85",
      "product_quantity": "9174 (4586 US; 4588 OUS)",
      "reason_for_recall": "IABP may lose the ability to charge batteries in one or both bay slots.    Therapy may be interrupted if batteries fail to charge and the device is disconnected from AC  power. Low battery alarms may alert the User to the issue prior to interruption of therapy.",
      "recall_initiation_date": "20230605",
      "center_classification_date": "20230719",
      "report_date": "20230726",
      "code_info": "All Unit Serial Numbers  Model                    UDI 0998-00-0800-75 10607567112312 0998-00-0800-83 10607567108407 0998-00-0800-85 10607567113449",
      "more_code_info": ""
    }
  ]
}