  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lugano",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85444",
      "recalling_firm": "Inpeco S.A.",
      "address_1": "Via San Gottardo 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide including in the states of IL, NY.",
      "recall_number": "Z-2142-2020",
      "product_description": "Aptio Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00) - Product Usage: is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.",
      "product_quantity": "70 systems",
      "reason_for_recall": "The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.",
      "recall_initiation_date": "20200407",
      "center_classification_date": "20200528",
      "termination_date": "20240206",
      "report_date": "20200603",
      "code_info": "Serial Numbers: AP2.0003  AP2.0007  AP2.0008  AP2.0009  AP2.0010  AP2.0017  AP2.0022  AP2.0025  AP2.0036  AP2.0049  AP2.0052  AP2.0057  AP2.0058  AP2.0059  AP2.0064  AP2.0065  AP2.0077  AP2.0082  AP2.0083  AP2.0089  AP2.0090  AP2.0092  AP2.0098  AP2.0100  AP2.0105  AP2.0109  AP2.0110  AP2.0129  AP2.0136  AP2.0149  AP2.0156  AP2.0173  AP2.0179  AP2.0180  AP2.0181  AP2.0193  AP2.0194  AP2.0201  AP2.0202  AP2.0211  AP2.0230  AP2.0233  AP2.0235  AP2.0236  AP2.0238  AP2.0251  AP2.0254  AP2.0258  AP2.0264  AP2.0269  AP2.0270  AP2.0275  AP2.0289  AP2.0295  AP2.0317  AP2.0325  AP2.0332  AP2.0339  AP2.0375  AP2.0381  AP2.0394  AP2.0411  AP2.0427  APT.0009  APT.0014  APT.0058  APT.0065  APT.0091  APT.0136  APT.0247"
    }
  ]
}