  "results": [
    {
      "country": "United States",
      "city": "Lake Forest",
      "address_1": "275 N Field Dr",
      "reason_for_recall": "Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.",
      "address_2": "",
      "product_quantity": "6,268 infusion pump systems",
      "code_info": "List No. 0F782; Serial Numbers: 100 ;  101 ;  102 ;  105 ;  105 ;  106 ;  106 ;  107 ;  108 ;  108 ;  109 ;  111 ;  112 ;  114 ;  116 ;  117 ;  119 ;  120 ;  121 ;  122 ;  123 ;  124 ;  125 ;  125 ;  126 ;  126 ;  127 ;  127 ;  128 ;  128 ;  129 ;  129 ;  130 ;  130 ;  131 ;  131 ;  132 ;  132 ;  133 ;  133 ;  134 ;  134 ;  135 ;  135 ;  136 ;  136 ;  137 ;  137 ;  138 ;  138 ;  139 ;  140 ;  141 ;  142 ;  142 ;  143 ;  143 ;  144 ;  144 ;  145 ;  146 ;  147 ;  147 ;  148 ;  148 ;  149 ;  150 ;  150 ;  151 ;  152 ;  153 ;  154 ;  154 ;  155 ;  155 ;  156 ;  157 ;  158 ;  158 ;  159 ;  160 ;  161 ;  162 ;  162 ;  163 ;  164 ;  164 ;  165 ;  165 ;  166 ;  166 ;  167 ;  167 ;  168 ;  169 ;  170 ;  171 ;  172 ;  172 ;  173 ;  173 ;  174 ;  174 ;  175 ;  175 ;  176 ;  176 ;  177 ;  177 ;  178 ;  178 ;  179 ;  180 ;  181 ;  182 ;  183 ;  183 ;  184 ;  184 ;  185 ;  186 ;  186 ;  187 ;  188 ;  188 ;  189 ;  189 ;  190 ;  191 ;  192 ;  193 ;  194 ;  195 ;  196 ;  197 ;  198 ;  199 ;  200 ;  200 ;  201 ;  201 ;  202 ;  202 ;  203 ;  203 ;  204 ;  204 ;  205 ;  205 ;  206 ;  206 ;  207 ;  207 ;  208 ;  208 ;  209 ;  209 ;  210 ;  210 ;  211 ;  212 ;  212 ;  213 ;  213 ;  214 ;  214 ;  215 ;  216 ;  216 ;  217 ;  217 ;  218 ;  218 ;  219 ;  220 ;  221 ;  222 ;  223 ;  224 ;  224 ;  225 ;  225 ;  226 ;  226 ;  227 ;  227 ;  228 ;  229 ;  229 ;  230 ;  231 ;  232 ;  233 ;  234 ;  235 ;  236 ;  237 ;  238 ;  239 ;  240 ;  241 ;  241 ;  242 ;  242 ;  243 ;  243 ;  244 ;  245 ;  245 ;  246 ;  246 ;  247 ;  247 ;  248 ;  248 ;  249 ;  250 ;  250 ;  251 ;  251 ;  252 ;  252 ;  253 ;  253 ;  254 ;  255 ;  256 ;  256 ;  257 ;  258 ;  259 ;  259 ;  260 ;  260 ;  261 ;  262 ;  263 ;  263 ;  264 ;  265 ;  265 ;  266 ;  266 ;  267 ;  267 ;  268 ;  269 ;  269 ;  270 ;  270 ;  271 ;  271 ;  272 ;  273 ;  273 ;  274 ;  274 ;  275 ;  275 ;  276 ;  277 ;  277 ;  278 ;  279 ;  279 ;  280 ;  280 ;  281 ;  281 ;  282 ;  283 ;  283 ;  284 ;  284 ;  285 ;  286 ;  287 ;  288 ;  288 ;  289 ;  290 ;  290 ;  291 ;  291 ;  292 ;  292 ;  293 ;  293 ;  294 ;  294 ;  295 ;  295 ;  296 ;  296 ;  297 ;  298 ;  299 ;  299 ;  300 ;  301 ;  301 ;  302 ;  303 ;  303 ;  304 ;  305 ;  305 ;  306 ;  307 ;  307 ;  308 ;  309 ;  310 ;  310 ;  311 ;  311 ;  312 ;  312 ;  313 ;  314 ;  315 ;  315 ;  316 ;  317 ;  318 ;  319 ;  319 ;  320 ;  321 ;  323 ;  324 ;  325 ;  326 ;  327 ;  327 ;  328 ;  329 ;  330 ;  330 ;  331 ;  332 ;  332 ;  333 ;  334 ;  335 ;  335 ;  336 ;  336 ;  337 ;  338 ;  339 ;  339 ;  340 ;  340 ;  341 ;  341 ;  342 ;  343 ;  343 ;  344 ;  345 ;  346 ;  346 ;  347 ;  347 ;  348 ;  349 ;  349 ;  350 ;  350 ;  351 ;  352 ;  353 ;  353 ;  354 ;  354 ;  355 ;  356 ;  357 ;  357 ;  358 ;  359 ;  359 ;  360 ;  360 ;  361 ;  361 ;  362 ;  362 ;  363 ;  363 ;  364 ;  364 ;  365 ;  366 ;  366 ;  367 ;  367 ;  368 ;  369 ;  370 ;  370 ;  371 ;  372 ;  372 ;  373 ;  373 ;  374 ;  374 ;  375 ;  376 ;  376 ;  377 ;  377 ;  378 ;  378 ;  379 ;  379 ;  380 ;  381 ;  382 ;  382 ;  383 ;  383 ;  384 ;  384 ;  385 ;  385 ;  386 ;  386 ;  387 ;  387 ;  388 ;  388 ;  389 ;  389 ;  390 ;  390 ;  391 ;  391 ;  392 ;  393 ;  393 ;  394 ;  394 ;  395 ;  395 ;  396 ;  396 ;  397 ;  398 ;  398 ;  399 ;  399 ;  400 ;  400 ;  401 ;  401 ;  402 ;  402 ;  403 ;  403 ;  404 ;  404 ;  405 ;  406 ;  406 ;  407 ;  407 ;  408 ;  408 ;  409 ;  409 ;  410 ;  410 ;  411 ;  411 ;  412 ;  412 ;  413 ;  413 ;  414 ;  414 ;  415 ;  415 ;  416 ;  416 ;  417 ;  417 ;  418 ;  418 ;  419 ;  420 ;  421 ;  422 ;  423 ;  424 ;  425 ;  426 ;  427 ;  428 ;  429 ;  430 ;  431 ;  432 ;  433 ;  434 ;  435 ;  436 ;  437 ;  438 ;  439 ;  440 ;  441 ;  442 ;  443 ;  444 ;  445 ;  446 ;  447 ;  448 ;  449 ;  450 ;  451 ;  452 ;  453 ;  454 ;  455 ;  456 ;  457 ;  458 ;  459 ;  460 ;  461 ;  462 ;  463 ;  464 ;  465 ;  466 ;  467 ;  468 ;  469 ;  470 ;  471 ;  472 ;  473 ;  474 ;  475 ;  476 ;  477 ;  478 ;  479 ;  480 ;  481 ;  482 ;  483 ;  484 ;  485 ;  486 ;  487 ;  488 ;  489 ;  490 ;  491 ;  492 ;  493 ;  494 ;  495 ;  496 ;  497 ;  498 ;  499 ;  500 ;  500 ;  501 ;  501 ;  502 ;  502 ;  503 ;  503 ;  504 ;  504 ;  505 ;  505 ;  506 ;  506 ;  507 ;  507 ;  508 ;  508 ;  509 ;  509 ;  510 ;  510 ;  511 ;  511 ;  512 ;  512 ;  513 ;  513 ;  514 ;  514 ;  515 ;  515 ;  516 ;  516 ;  517 ;  517 ;  518 ;  518 ;  519 ;  519 ;  520 ;  520 ;  521 ;  521 ;  522 ;  522 ;  523 ;  523 ;  524 ;  524 ;  525 ;  525 ;  526 ;  526 ;  527 ;  527 ;  528 ;  528 ;  529 ;  529 ;  530 ;  530 ;  531 ;  531 ;  532 ;  533 ;  533 ;  534 ;  534 ;  535 ;  535 ;  536 ;  536 ;  537 ;  537 ;  538 ;  538 ;  539 ;  540 ;  540 ;  541 ;  541 ;  542 ;  542 ;  543 ;  543 ;  544 ;  544 ;  545 ;  546 ;  546 ;  547 ;  548 ;  548 ;  549 ;  549 ;  550 ;  550 ;  551 ;  551 ;  552 ;  552 ;  553 ;  554 ;  555 ;  555 ;  556 ;  556 ;  557 ;  558 ;  558 ;  559 ;  559 ;  560 ;  560 ;  561 ;  561 ;  562 ;  562 ;  563 ;  563 ;  564 ;  564 ;  565 ;  565 ;  566 ;  566 ;  567 ;  567 ;  568 ;  568 ;  569 ;  570 ;  570 ;  571 ;  571 ;  572 ;  573 ;  573 ;  574 ;  574 ;  575 ;  576 ;  576 ;  577 ;  577 ;  578 ;  578 ;  579 ;  579 ;  580 ;  580 ;  581 ;  581 ;  582 ;  582 ;  583 ;  583 ;  584 ;  584 ;  585 ;  585 ;  586 ;  587 ;  588 ;  588 ;  589 ;  589 ;  590 ;  590 ;  591 ;  591 ;  592 ;  592 ;  593 ;  593 ;  594 ;  594 ;  595 ;  595 ;  596 ;  597 ;  597 ;  598 ;  598 ;  599 ;  599 ;  600 ;  600 ;  601 ;  601 ;  602 ;  602 ;  603 ;  603 ;  604 ;  604 ;  605 ;  605 ;  606 ;  607 ;  607 ;  608 ;  608 ;  609 ;  610 ;  610 ;  611 ;  612 ;  612 ;  613 ;  613 ;  614 ;  614 ;  615 ;  615 ;  616 ;  616 ;  617 ;  617 ;  618 ;  619 ;  619 ;  620 ;  620 ;  621 ;  621 ;  622 ;  622 ;  623 ;  623 ;  624 ;  624 ;  625 ;  625 ;  626 ;  626 ;  627 ;  627 ;  628 ;  628 ;  629 ;  629 ;  630 ;  630 ;  631 ;  631 ;  632 ;  632 ;  634 ;  634 ;  635 ;  636 ;  637 ;  637 ;  638 ;  638 ;  639 ;  640 ;  641 ;  641 ;  642 ;  642 ;  643 ;  643 ;  644 ;  645 ;  645 ;  646 ;  646 ;  647 ;  648 ;  648 ;  649 ;  649 ;  650 ;  650 ;  651 ;  651 ;  652 ;  653 ;  654 ;  654 ;  655 ;  656 ;  656 ;  657 ;  657 ;  658 ;  659 ;  659 ;  660 ;  660 ;  662 ;  663 ;  664 ;  666 ;  667 ;  668 ;  669 ;  670 ;  671 ;  672 ;  673 ;  674 ;  675 ;  676 ;  677 ;  678 ;  679 ;  681 ;  682 ;  683 ;  684 ;  685 ;  686 ;  687 ;  688 ;  689 ;  690 ;  691 ;  692 ;  693 ;  694 ;  696 ;  697 ;  698 ;  699 ;  700 ;  701 ;  702 ;  703 ;  704 ;  705 ;  706 ;  708 ;  709 ;  710 ;  711 ;  712 ;  713 ;  715 ;  716 ;  717 ;  718 ;  719 ;  720 ;  721 ;  723 ;  724 ;  725 ;  728 ;  729 ;  730 ;  731 ;  733 ;  734 ;  735 ;  737 ;  738 ;  741 ;  742 ;  744 ;  745 ;  747 ;  749 ;  751 ;  752 ;  753 ;  754 ;  755 ;  756 ;  757 ;  758 ;  759 ;  760 ;  761 ;  762 ;  763 ;  764 ;  765 ;  766 ;  767 ;  768 ;  769 ;  770 ;  771 ;  772 ;  773 ;  774 ;  775 ;  776 ;  777 ;  778 ;  779 ;  780 ;  781 ;  782 ;  783 ;  784 ;  785 ;  786 ;  787 ;  788 ;  789 ;  791 ;  793 ;  794 ;  795 ;  797 ;  798 ;  799 ;  800 ;  803 ;  804 ;  805 ;  806 ;  807 ;  808 ;  809 ;  810 ;  811 ;  812 ;  813 ;  814 ;  815 ;  816 ;  817 ;  818 ;  819 ;  820 ;  822 ;  823 ;  824 ;  826 ;  827 ;  828 ;  829 ;  830 ;  831 ;  832 ;  833 ;  834 ;  835 ;  836 ;  837 ;  838 ;  839 ;  840 ;  841 ;  842 ;  843 ;  844 ;  845 ;  846 ;  847 ;  848 ;  849 ;  850 ;  851 ;  852 ;  853 ;  854 ;  855 ;  856 ;  857 ;  858 ;  859 ;  860 ;  861 ;  862 ;  863 ;  864 ;  865 ;  866 ;  868 ;  869 ;  870 ;  871 ;  872 ;  873 ;  874 ;  875 ;  876 ;  877 ;  879 ;  880 ;  881 ;  882 ;  883 ;  884 ;  885 ;  886 ;  887 ;  888 ;  889 ;  890 ;  891 ;  892 ;  893 ;  894 ;  895 ;  896 ;  897 ;  898 ;  899 ;  900 ;  901 ;  902 ;  903 ;  904 ;  905 ;  906 ;  907 ;  908 ;  909 ;  910 ;  911 ;  912 ;  913 ;  914 ;  916 ;  917 ;  918 ;  919 ;  920 ;  921 ;  922 ;  923 ;  924 ;  925 ;  926 ;  927 ;  928 ;  929 ;  930 ;  931 ;  932 ;  933 ;  934 ;  935 ;  936 ;  937 ;  939 ;  940 ;  942 ;  943 ;  947 ;  948 ;  950 ;  951 ;  953 ;  955 ;  958 ;  959 ;  960 ;  961 ;  962 ;  963 ;  964 ;  965 ;  966 ;  967 ;  968 ;  970 ;  971 ;  972 ;  973 ;  974 ;  976 ;  977 ;  978 ;  979 ;  980 ;  981 ;  982 ;  983 ;  984 ;  986 ;  987 ;  988 ;  990 ;  991 ;  992 ;  993 ;  994 ;  995 ;  996 ;  997 ;  998 ;  999 ;  1000 ;  1001 ;  1002 ;  1003 ;  1004 ;  1005 ;  1007 ;  1008 ;  1009 ;  1010 ;  1011 ;  1012 ;  1013 ;  1014 ;  1015 ;  1016 ;  1017 ;  1018 ;  1019 ;  1020 ;  1021 ;  1022 ;  1023 ;  1024 ;  1025 ;  1026 ;  1027 ;  1028 ;  1029 ;  1030 ;  1032 ;  1033 ;  1034 ;  1035 ;  1036 ;  1037 ;  1038 ;  1039 ;  1040 ;  1041 ;  1042 ;  1043 ;  1044 ;  1045 ;  1046 ;  1047 ;  1048 ;  1049 ;  1050 ;  1051 ;  1052 ;  1053 ;  1054 ;  1055 ;  1056 ;  1057 ;  1059 ;  1060 ;  1061 ;  1062 ;  1063 ;  1064 ;  1065 ;  1066 ;  1067 ;  1068 ;  1069 ;  1070 ;  1071 ;  1072 ;  1073 ;  1074 ;  1075 ;  1077 ;  1078 ;  1079 ;  1080 ;  1081 ;  1082 ;  1083 ;  1084 ;  1085 ;  1086 ;  1087 ;  1088 ;  1089 ;  1090 ;  1091 ;  1092 ;  1093 ;  1094 ;  1095 ;  1096 ;  1097 ;  1098 ;  1099 ;  1100 ;  1101 ;  1102 ;  1103 ;  1104 ;  1105 ;  1106 ;  1107 ;  1108 ;  1109 ;  1110 ;  1111 ;  1112 ;  1113 ;  1114 ;  1115 ;  1116 ;  1117 ;  1118 ;  1119 ;  1120 ;  1121 ;  1122 ;  1123 ;  1124 ;  1125 ;  1126 ;  1127 ;  1128 ;  1129 ;  1130 ;  1131 ;  1132 ;  1133 ;  1135 ;  1136 ;  1138 ;  1139 ;  1140 ;  1141 ;  1142 ;  1143 ;  1145 ;  1147 ;  1148 ;  1149 ;  1150 ;  1151 ;  1152 ;  1153 ;  1155 ;  1157 ;  1158 ;  1159 ;  1160 ;  1161 ;  1162 ;  1163 ;  1164 ;  1165 ;  1166 ;  1167 ;  1168 ;  1169 ;  1170 ;  1171 ;  1173 ;  1174 ;  1175 ;  1176 ;  1177 ;  1178 ;  1179 ;  1180 ;  1181 ;  1182 ;  1183 ;  1185 ;  1186 ;  1187 ;  1188 ;  1189 ;  1190 ;  1191 ;  1192 ;  1193 ;  1194 ;  1198 ;  1199 ;  1201 ;  1202 ;  1203 ;  1205 ;  1206 ;  1207 ;  1209 ;  1210 ;  1212 ;  1213 ;  1214 ;  1215 ;  1216 ;  1217 ;  1218 ;  1220 ;  1221 ;  1222 ;  1223 ;  1224 ;  1225 ;  1227 ;  1229 ;  1230 ;  1231 ;  1232 ;  1233 ;  1235 ;  1236 ;  1237 ;  1238 ;  1239 ;  1240 ;  1241 ;  1243 ;  1245 ;  1246 ;  1247 ;  1248 ;  1249 ;  1250 ;  1252 ;  1253 ;  1254 ;  1255 ;  1256 ;  1257 ;  1258 ;  1261 ;  1262 ;  1263 ;  1265 ;  1266 ;  1267 ;  1268 ;  1269 ;  1270 ;  1271 ;  1272 ;  1273 ;  1274 ;  1275 ;  1276 ;  1277 ;  1278 ;  6268 ;  10000 ;  10000 ;  10000 ;  10000 ;  10000 ;  10001 ;  10001 ;  10001 ;  10001 ;  10002 ;  10002 ;  10002 ;  10002 ;  10003 ;  10003 ;  10003 ;  10003 ;  10003 ;  10004 ;  10004 ;  10004 ;  10004 ;  10004 ;  10004 ;  10005 ;  10005 ;  10005 ;  10005 ;  10005 ;  10005 ;  10005 ;  10006 ;  10006 ;  10006 ;  10006 ;  10006 ;  10007 ;  10007 ;  10007 ;  10007 ;  10007 ;  10008 ;  10008 ;  10008 ;  10008 ;  10008 ;  10008 ;  10009 ;  10009 ;  10009 ;  10009 ;  10009 ;  10010 ;  10010 ;  10010 ;  10010 ;  10010 ;  10011 ;  10011 ;  10011 ;  10011 ;  10011 ;  10011 ;  10011 ;  10012 ;  10012 ;  10012 ;  10012 ;  10012 ;  10013 ;  10013 ;  10013 ;  10013 ;  10013 ;  10013 ;  10014 ;  10014 ;  10014 ;  10014 ;  10014 ;  10014 ;  10015 ;  10015 ;  10015 ;  10015 ;  10015 ;  10015 ;  10016 ;  10016 ;  10016 ;  10016 ;  10016 ;  10016 ;  10017 ;  10017 ;  10017 ;  10017 ;  10017 ;  10017 ;  10018 ;  10018 ;  10018 ;  10018 ;  10018 ;  10019 ;  10019 ;  10019 ;  10019 ;  10019 ;  10019 ;  10020 ;  10020 ;  10020 ;  10020 ;  10020 ;  10021 ;  10021 ;  10021 ;  10021 ;  10021 ;  10021 ;  10022 ;  10022 ;  10022 ;  10022 ;  10022 ;  10023 ;  10023 ;  10023 ;  10023 ;  10023 ;  10023 ;  10024 ;  10024 ;  10024 ;  10024 ;  10024 ;  10025 ;  10025 ;  10025 ;  10025 ;  10025 ;  10025 ;  10026 ;  10026 ;  10026 ;  10026 ;  10026 ;  10026 ;  10027 ;  10027 ;  10027 ;  10028 ;  10028 ;  10028 ;  10028 ;  10028 ;  10029 ;  10029 ;  10029 ;  10029 ;  10030 ;  10030 ;  10030 ;  10030 ;  10030 ;  10030 ;  10031 ;  10031 ;  10031 ;  10031 ;  10031 ;  10031 ;  10032 ;  10032 ;  10032 ;  10032 ;  10033 ;  10033 ;  10033 ;  10033 ;  10034 ;  10034 ;  10034 ;  10034 ;  10034 ;  10035 ;  10035 ;  10035 ;  10035 ;  10036 ;  10036 ;  10036 ;  10036 ;  10036 ;  10036 ;  10037 ;  10037 ;  10037 ;  10037 ;  10037 ;  10037 ;  10038 ;  10038 ;  10038 ;  10039 ;  10039 ;  10039 ;  10039 ;  10039 ;  10040 ;  10040 ;  10040 ;  10040 ;  10040 ;  10041 ;  10041 ;  10041 ;  10041 ;  10042 ;  10042 ;  10043 ;  10043 ;  10043 ;  10043 ;  10044 ;  10044 ;  10044 ;  10044 ;  10045 ;  10045 ;  10045 ;  10045 ;  10046 ;  10046 ;  10046 ;  10047 ;  10047 ;  10047 ;  10047 ;  10048 ;  10048 ;  10048 ;  10049 ;  10049 ;  10049 ;  10050 ;  10050 ;  10050 ;  10050 ;  10051 ;  10051 ;  10051 ;  10051 ;  10052 ;  10052 ;  10052 ;  10052 ;  10053 ;  10053 ;  10053 ;  10053 ;  10054 ;  10054 ;  10054 ;  10054 ;  10055 ;  10055 ;  10055 ;  10055 ;  10056 ;  10056 ;  10056 ;  10056 ;  10057 ;  10057 ;  10057 ;  10058 ;  10058 ;  10059 ;  10059 ;  10059 ;  10059 ;  10060 ;  10060 ;  10061 ;  10061 ;  10062 ;  10062 ;  10062 ;  10062 ;  10063 ;  10063 ;  10063 ;  10064 ;  10064 ;  10064 ;  10065 ;  10065 ;  10065 ;  10066 ;  10066 ;  10067 ;  10067 ;  10068 ;  10068 ;  10068 ;  10069 ;  10069 ;  10069 ;  10070 ;  10070 ;  10070 ;  10071 ;  10071 ;  10071 ;  10072 ;  10072 ;  10072 ;  10073 ;  10073 ;  10073 ;  10074 ;  10074 ;  10074 ;  10075 ;  10075 ;  10075 ;  10075 ;  10076 ;  10076 ;  10076 ;  10077 ;  10077 ;  10077 ;  10078 ;  10078 ;  10079 ;  10079 ;  10079 ;  10080 ;  10080 ;  10080 ;  10080 ;  10081 ;  10081 ;  10081 ;  10082 ;  10082 ;  10082 ;  10082 ;  10083 ;  10083 ;  10083 ;  10084 ;  10084 ;  10084 ;  10085 ;  10085 ;  10085 ;  10085 ;  10086 ;  10086 ;  10086 ;  10087 ;  10087 ;  10087 ;  10088 ;  10088 ;  10088 ;  10089 ;  10089 ;  10089 ;  10090 ;  10090 ;  10090 ;  10090 ;  10091 ;  10091 ;  10091 ;  10091 ;  10092 ;  10092 ;  10092 ;  10093 ;  10093 ;  10094 ;  10094 ;  10094 ;  10094 ;  10095 ;  10095 ;  10095 ;  10095 ;  10096 ;  10096 ;  10096 ;  10097 ;  10097 ;  10097 ;  10098 ;  10098 ;  10098 ;  10098 ;  10098 ;  10099 ;  10099 ;  10099 ;  10099 ;  10100 ;  10100 ;  10100 ;  10101 ;  10101 ;  10101 ;  10101 ;  10102 ;  10102 ;  10102 ;  10103 ;  10103 ;  10103 ;  10103 ;  10103 ;  10104 ;  10104 ;  10104 ;  10104 ;  10105 ;  10105 ;  10105 ;  10105 ;  10106 ;  10106 ;  10106 ;  10106 ;  10107 ;  10107 ;  10107 ;  10107 ;  10108 ;  10108 ;  10108 ;  10108 ;  10109 ;  10109 ;  10109 ;  10109 ;  10110 ;  10110 ;  10110 ;  10110 ;  10110 ;  10111 ;  10111 ;  10111 ;  10111 ;  10112 ;  10112 ;  10112 ;  10113 ;  10113 ;  10113 ;  10113 ;  10114 ;  10114 ;  10114 ;  10115 ;  10115 ;  10115 ;  10115 ;  10116 ;  10116 ;  10116 ;  10117 ;  10117 ;  10118 ;  10118 ;  10118 ;  10119 ;  10119 ;  10119 ;  10119 ;  10120 ;  10120 ;  10121 ;  10121 ;  10121 ;  10121 ;  10122 ;  10122 ;  10122 ;  10123 ;  10123 ;  10123 ;  10123 ;  10124 ;  10124 ;  10124 ;  10125 ;  10126 ;  10126 ;  10126 ;  10127 ;  10127 ;  10128 ;  10128 ;  10128 ;  10129 ;  10129 ;  10129 ;  10130 ;  10130 ;  10131 ;  10131 ;  10131 ;  10132 ;  10132 ;  10132 ;  10132 ;  10133 ;  10133 ;  10133 ;  10133 ;  10134 ;  10134 ;  10134 ;  10134 ;  10135 ;  10135 ;  10135 ;  10136 ;  10137 ;  10137 ;  10137 ;  10138 ;  10138 ;  10138 ;  10138 ;  10139 ;  10139 ;  10139 ;  10140 ;  10140 ;  10141 ;  10141 ;  10141 ;  10142 ;  10142 ;  10143 ;  10143 ;  10143 ;  10144 ;  10145 ;  10145 ;  10145 ;  10146 ;  10146 ;  10146 ;  10146 ;  10147 ;  10147 ;  10148 ;  10148 ;  10149 ;  10149 ;  10150 ;  10150 ;  10150 ;  10151 ;  10151 ;  10151 ;  10152 ;  10152 ;  10152 ;  10153 ;  10153 ;  10154 ;  10154 ;  10154 ;  10155 ;  10155 ;  10156 ;  10156 ;  10156 ;  10157 ;  10157 ;  10158 ;  10158 ;  10158 ;  10158 ;  10159 ;  10159 ;  10159 ;  10160 ;  10160 ;  10160 ;  10160 ;  10161 ;  10161 ;  10161 ;  10162 ;  10162 ;  10162 ;  10163 ;  10163 ;  10163 ;  10164 ;  10164 ;  10164 ;  10165 ;  10165 ;  10165 ;  10166 ;  10166 ;  10166 ;  10167 ;  10167 ;  10167 ;  10168 ;  10168 ;  10169 ;  10169 ;  10169 ;  10169 ;  10170 ;  10170 ;  10171 ;  10171 ;  10171 ;  10172 ;  10172 ;  10172 ;  10172 ;  10173 ;  10173 ;  10173 ;  10174 ;  10175 ;  10176 ;  10176 ;  10177 ;  10177 ;  10177 ;  10178 ;  10178 ;  10178 ;  10178 ;  10179 ;  10179 ;  10179 ;  10179 ;  10180 ;  10180 ;  10180 ;  10181 ;  10181 ;  10181 ;  10182 ;  10182 ;  10183 ;  10183 ;  10184 ;  10184 ;  10184 ;  10185 ;  10185 ;  10185 ;  10186 ;  10186 ;  10187 ;  10187 ;  10187 ;  10188 ;  10188 ;  10189 ;  10189 ;  10190 ;  10190 ;  10190 ;  10191 ;  10192 ;  10192 ;  10193 ;  10193 ;  10194 ;  10194 ;  10195 ;  10195 ;  10195 ;  10195 ;  10196 ;  10196 ;  10196 ;  10197 ;  10197 ;  10197 ;  10198 ;  10198 ;  10198 ;  10199 ;  10199 ;  10200 ;  10200 ;  10200 ;  10201 ;  10201 ;  10201 ;  10202 ;  10202 ;  10203 ;  10203 ;  10203 ;  10204 ;  10204 ;  10205 ;  10205 ;  10206 ;  10206 ;  10206 ;  10207 ;  10207 ;  10207 ;  10208 ;  10208 ;  10209 ;  10209 ;  10209 ;  10210 ;  10210 ;  10211 ;  10211 ;  10211 ;  10212 ;  10212 ;  10212 ;  10213 ;  10213 ;  10213 ;  10214 ;  10214 ;  10215 ;  10215 ;  10216 ;  10216 ;  10216 ;  10217 ;  10217 ;  10218 ;  10218 ;  10219 ;  10220 ;  10220 ;  10221 ;  10221 ;  10222 ;  10222 ;  10222 ;  10223 ;  10223 ;  10223 ;  10224 ;  10224 ;  10224 ;  10224 ;  10225 ;  10225 ;  10225 ;  10226 ;  10226 ;  10226 ;  10226 ;  10227 ;  10227 ;  10227 ;  10228 ;  10228 ;  10229 ;  10229 ;  10230 ;  10230 ;  10230 ;  10231 ;  10231 ;  10231 ;  10232 ;  10232 ;  10232 ;  10233 ;  10233 ;  10233 ;  10234 ;  10234 ;  10234 ;  10235 ;  10235 ;  10235 ;  10236 ;  10236 ;  10236 ;  10236 ;  10237 ;  10237 ;  10237 ;  10237 ;  10238 ;  10238 ;  10238 ;  10239 ;  10239 ;  10239 ;  10240 ;  10240 ;  10240 ;  10241 ;  10241 ;  10241 ;  10242 ;  10242 ;  10242 ;  10243 ;  10243 ;  10244 ;  10244 ;  10244 ;  10244 ;  10245 ;  10245 ;  10245 ;  10246 ;  10246 ;  10247 ;  10247 ;  10247 ;  10248 ;  10248 ;  10248 ;  10249 ;  10249 ;  10249 ;  10249 ;  10250 ;  10250 ;  10250 ;  10250 ;  10251 ;  10251 ;  10251 ;  10252 ;  10252 ;  10252 ;  10253 ;  10253 ;  10254 ;  10254 ;  10254 ;  10255 ;  10255 ;  10255 ;  10256 ;  10256 ;  10256 ;  10257 ;  10257 ;  10257 ;  10258 ;  10258 ;  10258 ;  10259 ;  10259 ;  10259 ;  10259 ;  10260 ;  10260 ;  10260 ;  10260 ;  10261 ;  10261 ;  10261 ;  10261 ;  10262 ;  10262 ;  10262 ;  10263 ;  10263 ;  10264 ;  10264 ;  10265 ;  10265 ;  10265 ;  10266 ;  10266 ;  10266 ;  10267 ;  10267 ;  10267 ;  10267 ;  10268 ;  10268 ;  10268 ;  10268 ;  10269 ;  10269 ;  10269 ;  10269 ;  10270 ;  10270 ;  10270 ;  10271 ;  10271 ;  10271 ;  10272 ;  10272 ;  10272 ;  10273 ;  10273 ;  10274 ;  10274 ;  10274 ;  10274 ;  10275 ;  10275 ;  10275 ;  10275 ;  10276 ;  10276 ;  10276 ;  10276 ;  10277 ;  10277 ;  10277 ;  10278 ;  10278 ;  10278 ;  10278 ;  10279 ;  10279 ;  10279 ;  10280 ;  10280 ;  10280 ;  10280 ;  10281 ;  10281 ;  10281 ;  10282 ;  10282 ;  10282 ;  10283 ;  10283 ;  10283 ;  10283 ;  10284 ;  10284 ;  10284 ;  10284 ;  10285 ;  10285 ;  10285 ;  10285 ;  10286 ;  10286 ;  10286 ;  10287 ;  10287 ;  10287 ;  10287 ;  10288 ;  10288 ;  10288 ;  10289 ;  10289 ;  10289 ;  10290 ;  10290 ;  10290 ;  10290 ;  10291 ;  10291 ;  10291 ;  10292 ;  10292 ;  10293 ;  10293 ;  10293 ;  10294 ;  10294 ;  10294 ;  10295 ;  10295 ;  10295 ;  10295 ;  10296 ;  10296 ;  10296 ;  10297 ;  10297 ;  10297 ;  10298 ;  10298 ;  10298 ;  10299 ;  10299 ;  10299 ;  10300 ;  10300 ;  10300 ;  10301 ;  10301 ;  10302 ;  10302 ;  10302 ;  10302 ;  10303 ;  10303 ;  10303 ;  10304 ;  10304 ;  10305 ;  10305 ;  10305 ;  10306 ;  10306 ;  10306 ;  10307 ;  10307 ;  10307 ;  10307 ;  10308 ;  10308 ;  10308 ;  10309 ;  10309 ;  10309 ;  10310 ;  10310 ;  10311 ;  10311 ;  10311 ;  10312 ;  10312 ;  10312 ;  10312 ;  10313 ;  10313 ;  10314 ;  10314 ;  10314 ;  10315 ;  10315 ;  10315 ;  10315 ;  10316 ;  10316 ;  10316 ;  10317 ;  10317 ;  10317 ;  10317 ;  10318 ;  10318 ;  10318 ;  10319 ;  10319 ;  10320 ;  10320 ;  10321 ;  10321 ;  10322 ;  10322 ;  10323 ;  10324 ;  10324 ;  10324 ;  10325 ;  10325 ;  10326 ;  10327 ;  10327 ;  10328 ;  10328 ;  10328 ;  10329 ;  10329 ;  10330 ;  10330 ;  10331 ;  10332 ;  10332 ;  10333 ;  10334 ;  10334 ;  10335 ;  10336 ;  10336 ;  10337 ;  10337 ;  10337 ;  10339 ;  10340 ;  10340 ;  10341 ;  10341 ;  10342 ;  10342 ;  10342 ;  10343 ;  10343 ;  10344 ;  10344 ;  10345 ;  10346 ;  10346 ;  10347 ;  10348 ;  10348 ;  10349 ;  10349 ;  10350 ;  10350 ;  10351 ;  10352 ;  10353 ;  10353 ;  10354 ;  10354 ;  10354 ;  10355 ;  10355 ;  10355 ;  10356 ;  10356 ;  10357 ;  10357 ;  10357 ;  10358 ;  10358 ;  10359 ;  10359 ;  10360 ;  10360 ;  10361 ;  10361 ;  10361 ;  10363 ;  10363 ;  10364 ;  10365 ;  10365 ;  10366 ;  10367 ;  10367 ;  10368 ;  10369 ;  10369 ;  10370 ;  10370 ;  10371 ;  10371 ;  10372 ;  10372 ;  10373 ;  10373 ;  10374 ;  10375 ;  10375 ;  10376 ;  10376 ;  10377 ;  10377 ;  10378 ;  10379 ;  10379 ;  10380 ;  10381 ;  10382 ;  10383 ;  10384 ;  10385 ;  10386 ;  10387 ;  10388 ;  10389 ;  10390 ;  10390 ;  10391 ;  10391 ;  10394 ;  10395 ;  10396 ;  10397 ;  10398 ;  10399 ;  10400 ;  10401 ;  10401 ;  10402 ;  10402 ;  10403 ;  10404 ;  10405 ;  10406 ;  10407 ;  10407 ;  10408 ;  10409 ;  10410 ;  10410 ;  10411 ;  10411 ;  10412 ;  10413 ;  10414 ;  10415 ;  10415 ;  10416 ;  10417 ;  10417 ;  10418 ;  10419 ;  10419 ;  10420 ;  10421 ;  10422 ;  10422 ;  10423 ;  10424 ;  10424 ;  10425 ;  10426 ;  10426 ;  10427 ;  10427 ;  10428 ;  10429 ;  10430 ;  10430 ;  10431 ;  10432 ;  10433 ;  10434 ;  10434 ;  10435 ;  10436 ;  10437 ;  10438 ;  10438 ;  10439 ;  10440 ;  10440 ;  10441 ;  10441 ;  10442 ;  10443 ;  10443 ;  10444 ;  10444 ;  10445 ;  10446 ;  10447 ;  10448 ;  10449 ;  10449 ;  10450 ;  10451 ;  10452 ;  10453 ;  10454 ;  10455 ;  10455 ;  10456 ;  10457 ;  10458 ;  10459 ;  10460 ;  10461 ;  10461 ;  10462 ;  10462 ;  10463 ;  10464 ;  10465 ;  10466 ;  10467 ;  10467 ;  10468 ;  10469 ;  10469 ;  10470 ;  10470 ;  10471 ;  10471 ;  10472 ;  10472 ;  10473 ;  10473 ;  10474 ;  10474 ;  10475 ;  10476 ;  10477 ;  10478 ;  10479 ;  10479 ;  10480 ;  10481 ;  10482 ;  10483 ;  10484 ;  10484 ;  10485 ;  10485 ;  10486 ;  10486 ;  10487 ;  10487 ;  10488 ;  10488 ;  10489 ;  10490 ;  10490 ;  10491 ;  10491 ;  10492 ;  10492 ;  10493 ;  10493 ;  10494 ;  10495 ;  10495 ;  10496 ;  10497 ;  10498 ;  10498 ;  10499 ;  10500 ;  10500 ;  10501 ;  10501 ;  10502 ;  10503 ;  10504 ;  10504 ;  10505 ;  10506 ;  10507 ;  10508 ;  10509 ;  10509 ;  10510 ;  10510 ;  10511 ;  10511 ;  10512 ;  10512 ;  10513 ;  10514 ;  10514 ;  10515 ;  10515 ;  10516 ;  10516 ;  10517 ;  10517 ;  10518 ;  10518 ;  10519 ;  10519 ;  10520 ;  10521 ;  10522 ;  10524 ;  10525 ;  10526 ;  10526 ;  10527 ;  10527 ;  10528 ;  10528 ;  10529 ;  10530 ;  10530 ;  10533 ;  10533 ;  10534 ;  10534 ;  10535 ;  10536 ;  10536 ;  10537 ;  10538 ;  10539 ;  10539 ;  10540 ;  10540 ;  10541 ;  10541 ;  10542 ;  10543 ;  10544 ;  10545 ;  10546 ;  10547 ;  10547 ;  10548 ;  10548 ;  10549 ;  10549 ;  10550 ;  10551 ;  10552 ;  10553 ;  10553 ;  10554 ;  10555 ;  10556 ;  10557 ;  10558 ;  10558 ;  10559 ;  10560 ;  10561 ;  10561 ;  10562 ;  10563 ;  10564 ;  10565 ;  10566 ;  10567 ;  10568 ;  10569 ;  10570 ;  10571 ;  10571 ;  10572 ;  10573 ;  10573 ;  10574 ;  10574 ;  10575 ;  10575 ;  10576 ;  10576 ;  10577 ;  10578 ;  10579 ;  10580 ;  10580 ;  10581 ;  10581 ;  10582 ;  10582 ;  10583 ;  10584 ;  10585 ;  10585 ;  10586 ;  10587 ;  10587 ;  10588 ;  10589 ;  10590 ;  10591 ;  10591 ;  10592 ;  10593 ;  10594 ;  10595 ;  10595 ;  10596 ;  10597 ;  10597 ;  10598 ;  10598 ;  10599 ;  10599 ;  10600 ;  10600 ;  10601 ;  10602 ;  10602 ;  10603 ;  10603 ;  10604 ;  10605 ;  10605 ;  10606 ;  10607 ;  10608 ;  10608 ;  10609 ;  10609 ;  10610 ;  10610 ;  10611 ;  10611 ;  10612 ;  10612 ;  10613 ;  10614 ;  10614 ;  10615 ;  10616 ;  10616 ;  10617 ;  10618 ;  10619 ;  10619 ;  10620 ;  10621 ;  10622 ;  10623 ;  10623 ;  10624 ;  10624 ;  10625 ;  10626 ;  10626 ;  10627 ;  10628 ;  10628 ;  10629 ;  10629 ;  10630 ;  10631 ;  10631 ;  10632 ;  10633 ;  10633 ;  10634 ;  10634 ;  10635 ;  10635 ;  10636 ;  10637 ;  10637 ;  10638 ;  10638 ;  10639 ;  10640 ;  10640 ;  10641 ;  10641 ;  10642 ;  10642 ;  10643 ;  10643 ;  10644 ;  10645 ;  10646 ;  10647 ;  10648 ;  10649 ;  10650 ;  10651 ;  10652 ;  10653 ;  10654 ;  10655 ;  10656 ;  10657 ;  10658 ;  10659 ;  10660 ;  10661 ;  10662 ;  10663 ;  10664 ;  10665 ;  10666 ;  10667 ;  10668 ;  10669 ;  10670 ;  10671 ;  10672 ;  10674 ;  10675 ;  10676 ;  10677 ;  10678 ;  10679 ;  10680 ;  10681 ;  10682 ;  10683 ;  10684 ;  10685 ;  10686 ;  10687 ;  10688 ;  10689 ;  10690 ;  10691 ;  10692 ;  10693 ;  10694 ;  10695 ;  10696 ;  10697 ;  10698 ;  10699 ;  10700 ;  10701 ;  10702 ;  10703 ;  10704 ;  10706 ;  10707 ;  10708 ;  10709 ;  10710 ;  10711 ;  10712 ;  10713 ;  10714 ;  10715 ;  10716 ;  10717 ;  10718 ;  10719 ;  10721 ;  10722 ;  10723 ;  10724 ;  10725 ;  10726 ;  10727 ;  10728 ;  10729 ;  10730 ;  10731 ;  10732 ;  10733 ;  10734 ;  10735 ;  10736 ;  10737 ;  10738 ;  10739 ;  10740 ;  10741 ;  10742 ;  10743 ;  10744 ;  10745 ;  10746 ;  10747 ;  10748 ;  10749 ;  10750 ;  10751 ;  10752 ;  10753 ;  10754 ;  10755 ;  10756 ;  10757 ;  10758 ;  10759 ;  10760 ;  10761 ;  10762 ;  10763 ;  10764 ;  10765 ;  10766 ;  10767 ;  10768 ;  10769 ;  10770 ;  10771 ;  10772 ;  10773 ;  10774 ;  10776 ;  10778 ;  10779 ;  10780 ;  10781 ;  10782 ;  10783 ;  10784 ;  10785 ;  10786 ;  10787 ;  10788 ;  10789 ;  10790 ;  10791 ;  10792 ;  10793 ;  10794 ;  10795 ;  10796 ;  10797 ;  10798 ;  10799 ;  10800 ;  10801 ;  10802 ;  10803 ;  10804 ;  10805 ;  10806 ;  10807 ;  10808 ;  10809 ;  10810 ;  10811 ;  10812 ;  10813 ;  10814 ;  10815 ;  10816 ;  10817 ;  10818 ;  10819 ;  10820 ;  10821 ;  10822 ;  10823 ;  10824 ;  10825 ;  10826 ;  10827 ;  10828 ;  10829 ;  10830 ;  10831 ;  10832 ;  10833 ;  10835 ;  10836 ;  10837 ;  10838 ;  10839 ;  10840 ;  10841 ;  10842 ;  10843 ;  10844 ;  10845 ;  10846 ;  10847 ;  10848 ;  10849 ;  10850 ;  10851 ;  10852 ;  10853 ;  10854 ;  10855 ;  10856 ;  10857 ;  10858 ;  10859 ;  10860 ;  10861 ;  10862 ;  10863 ;  10864 ;  10865 ;  10866 ;  10867 ;  10868 ;  10869 ;  10870 ;  10871 ;  10872 ;  10873 ;  10874 ;  10875 ;  10876 ;  10877 ;  10878 ;  10879 ;  10880 ;  10881 ;  10882 ;  10883 ;  10885 ;  10886 ;  10887 ;  10888 ;  10890 ;  10892 ;  10893 ;  10894 ;  10895 ;  10896 ;  10897 ;  10898 ;  10899 ;  10900 ;  10901 ;  10903 ;  10904 ;  10905 ;  10907 ;  10909 ;  10910 ;  10911 ;  10912 ;  10914 ;  10915 ;  10916 ;  10917 ;  10918 ;  10920 ;  10922 ;  10923 ;  10924 ;  10926 ;  10929 ;  10930 ;  10931 ;  10932 ;  10933 ;  10935 ;  10936 ;  10937 ;  10938 ;  10939 ;  10940 ;  10941 ;  10943 ;  10944 ;  10945 ;  10946 ;  10947 ;  10948 ;  10949 ;  10950 ;  10951 ;  10953 ;  10954 ;  10956 ;  10957 ;  10958 ;  10959 ;  10960 ;  10961 ;  10962 ;  10963 ;  10964 ;  10965 ;  10966 ;  10967 ;  10968 ;  10969 ;  10970 ;  10971 ;  10972 ;  10973 ;  10974 ;  10975 ;  10976 ;  10978 ;  10979 ;  10980 ;  10981 ;  10982 ;  10983 ;  10985 ;  10986 ;  10987 ;  10988 ;  10989 ;  10990 ;  10991 ;  10993 ;  10994 ;  10995 ;  10996 ;  10997 ;  10998 ;  10999 ;  11000 ;  11001 ;  11002 ;  11003 ;  11004 ;  11005 ;  11006 ;  11007 ;  11008 ;  11009 ;  11010 ;  11011 ;  11012 ;  11013 ;  11014 ;  11015 ;  11016 ;  11017 ;  11018 ;  11019 ;  11020 ;  11021 ;  11022 ;  11023 ;  11024 ;  11025 ;  11026 ;  11027 ;  11028 ;  11029 ;  11030 ;  11031 ;  11032 ;  11033 ;  11034 ;  11035 ;  11036 ;  11037 ;  11038 ;  11039 ;  11040 ;  11041 ;  11042 ;  11043 ;  11044 ;  11045 ;  11046 ;  11047 ;  11048 ;  11049 ;  11050 ;  11051 ;  11052 ;  11053 ;  11054 ;  11055 ;  11056 ;  11057 ;  11058 ;  11059 ;  11060 ;  11061 ;  11062 ;  11063 ;  11064 ;  11065 ;  11066 ;  11067 ;  11068 ;  11069 ;  11071 ;  11072 ;  11073 ;  11074 ;  11075 ;  11076 ;  11077 ;  11078 ;  11079 ;  11080 ;  11082 ;  11083 ;  11084 ;  11085 ;  11086 ;  11087 ;  11088 ;  11089 ;  11090 ;  18024 ;  40757 ;  40861 ;  102745 ;  102805 ;  85670037 ;  95590001 ;  95590002 ;  95590003 ;  95590004 ;  95590005 ;  95590006 ;  95590007 ;  95590008 ;  95590009 ;  95590010 ;  95590011 ;  95590012 ;  95590013 ;  95590014 ;  95590015 ;  95590016 ;  95590017 ;  95590018 ;  95590019 ;  95590020 ;  95590021 ;  95590022 ;  95590023 ;  95590024 ;  95590025 ;  95590026 ;  95590027 ;  95590028 ;  95590029 ;  95590030 ;  95590031 ;  95590032 ;  95590033 ;  95590034 ;  95590035 ;  95590036 ;  95590037 ;  95590038 ;  95590039 ;  95590040 ;  95590041 ;  95590042 ;  95590043 ;  95590044 ;  95590045 ;  95590046 ;  95590047 ;  95590048 ;  95590049 ;  95590050 ;  95590051 ;  95590052 ;  95590053 ;  95590054 ;  95590055 ;  95590056 ;  95590057 ;  95590058 ;  95590059 ;  95590060 ;  95590061 ;  95590062 ;  95590063 ;  95590064 ;  95590065 ;  95590066 ;  95590067 ;  95590068 ;  95590069 ;  95590070 ;  95590071 ;  95590072 ;  95590073 ;  95590074 ;  95590075 ;  95590076 ;  95590077 ;  95590078 ;  95590079 ;  95590080 ;  95590081 ;  95590082 ;  95590083 ;  95590084 ;  95590085 ;  95590086 ;  95590087 ;  95590088 ;  95590089 ;  95590090 ;  95590091 ;  95590092 ;  95590093 ;  95590094 ;  95590095 ;  95590096 ;  95590097 ;  95590098 ;  95590099 ;  95590100 ;  95590101 ;  95590102 ;  95590103 ;  95590104 ;  95590105 ;  95590106 ;  95590107 ;  95590108 ;  95590109 ;  95590110 ;  95590111 ;  95590112 ;  95590113 ;  95590114 ;  95590115 ;  95590116 ;  95590117 ;  95590118 ;  95590119 ;  95590120 ;  95590121 ;  95590122 ;  95590123 ;  95590124 ;  95590125 ;  95590126 ;  95590127 ;  95590128 ;  95590129 ;  95590130 ;  95590131 ;  95590132 ;  95590133 ;  95590134 ;  95590135 ;  95590136 ;  95590137 ;  95590138 ;  95590139 ;  95590140 ;  95590141 ;  95590142 ;  95590143 ;  95590144 ;  95590145 ;  95590146 ;  95590147 ;  95590148 ;  95590149 ;  95590150 ;  95590151 ;  95590152 ;  95590153 ;  95590154 ;  95590155 ;  95590156 ;  95590157 ;  95590158 ;  95590159 ;  95590160 ;  95590161 ;  95590162 ;  95590163 ;  95590164 ;  95590165 ;  95590166 ;  95590167 ;  95590168 ;  95590169 ;  95590170 ;  95590171 ;  95590172 ;  95590173 ;  95590174 ;  95590175 ;  95590176 ;  95590177 ;  95590178 ;  95590179 ;  95590180 ;  95590181 ;  95590182 ;  95590183 ;  95590184 ;  95590185 ;  95590186 ;  95590187 ;  95590188 ;  95590189 ;  95590190 ;  95590191 ;  95590192 ;  95590193 ;  95590194 ;  95590195 ;  95590196 ;  95590197 ;  95590198 ;  95590199 ;  95590200 ;  95590201 ;  95590202 ;  95590203 ;  95590204 ;  95590205 ;  95590206 ;  95590207 ;  95590208 ;  95590209 ;  95590210 ;  95590211 ;  95590212 ;  95590213 ;  95590214 ;  95590215 ;  95590216 ;  95590217 ;  95590218 ;  95590219 ;  95590220 ;  95590221 ;  95590222 ;  95590223 ;  95590224 ;  95590225 ;  95590226 ;  95590227 ;  95590228 ;  95590229 ;  95590230 ;  95590231 ;  95590232 ;  95590233 ;  95590234 ;  95590235 ;  95590236 ;  95590237 ;  95590238 ;  95590239 ;  95590240 ;  95590241 ;  95590242 ;  95590243 ;  95590244 ;  95590245 ;  95590246 ;  95590247 ;  95590248 ;  95590249 ;  95590250 ;  95590251 ;  95590252 ;  95590253 ;  95590254 ;  95590255 ;  95590256 ;  95590257 ;  95590258 ;  95590259 ;  95590260 ;  95590261 ;  95590262 ;  95590263 ;  95590264 ;  95590265 ;  95590266 ;  95590267 ;  95590268 ;  95590269 ;  95590270 ;  95590271 ;  95590272 ;  95590273 ;  95590274 ;  95590275 ;  95590276 ;  95590277 ;  95590278 ;  95590279 ;  95590280 ;  95590281 ;  95590282 ;  95590283 ;  95590284 ;  95590285 ;  95590286 ;  95590287 ;  95590288 ;  95590289 ;  95590290 ;  95590291 ;  95590292 ;  95590293 ;  95590294 ;  95590295 ;  95590296 ;  95590297 ;  95590298 ;  95590299 ;  95590300 ;  95590301 ;  95590302 ;  95590303 ;  95590304 ;  95590305 ;  95590306 ;  95590307 ;  95590308 ;  95590309 ;  95590310 ;  95590311 ;  95590312 ;  95590313 ;  95590314 ;  95590315 ;  95590316 ;  95590317 ;  95590318 ;  95590319 ;  95590320 ;  95590321 ;  95590322 ;  95590323 ;  95590324 ;  95590325 ;  95590326 ;  95590327 ;  95590328 ;  95590329 ;  95590330 ;  95590331 ;  95590332 ;  95590333 ;  95590334 ;  95590335 ;  95590336 ;  95590337 ;  95590338 ;  95590339 ;  95590340 ;  95590341 ;  95590342 ;  95590343 ;  95590344 ;  95590345 ;  95590346 ;  95590347 ;  95590348 ;  95590349 ;  95590350 ;  95590351 ;  95590352 ;  95590353 ;  95590354 ;  95590355 ;  95590356 ;  95590357 ;  95590358 ;  95590359 ;  95590360 ;  95590361 ;  95590362 ;  95590363 ;  95590364 ;  95590365 ;  95590366 ;  95590367 ;  95590368 ;  95590369 ;  95590370 ;  95590371 ;  95590372 ;  95590373 ;  95590374 ;  95590375 ;  95590376 ;  95590377 ;  95590378 ;  95590379 ;  95590380 ;  95590381 ;  95590382 ;  95590383 ;  95590384 ;  95590385 ;  95590386 ;  95590387 ;  95590388 ;  9",
      "center_classification_date": "20160708",
      "distribution_pattern": "Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:",
      "state": "IL",
      "product_description": "The LifeCare Pump Model 4200 PCA Infuser makes possible the practical implementation of patient-controlled analgesia. The health care professional sets the prescribed dose, rates, and limits for the drug to be infused. The infuser can be set to operate in one of three modes: PCA Mode, Continuous Mode, and PCA + Continuous Mode. The LifeCare Pump Model 4200 PCA Infuser is designed with a number of mechanical, electrical, and software safeguards that prevent patient injury or drug abuse.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hospira Inc.",
      "recall_number": "Z-2142-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68318",
      "termination_date": "20161107",
      "more_code_info": "5590389 ;  95590390 ;  95590391 ;  95590392 ;  95590393 ;  95590394 ;  95590395 ;  95590396 ;  95590397 ;  95590398 ;  95590399 ;  95590400 ;  95590401 ;  95590402 ;  95590403 ;  95590404 ;  95590405 ;  95590406 ;  95590407 ;  95590408 ;  95590409 ;  95590410 ;  95590411 ;  95590412 ;  95590413 ;  95590414 ;  95590415 ;  95590416 ;  95590417 ;  95590418 ;  95590419 ;  95590420 ;  95590421 ;  95590422 ;  95590423 ;  95590424 ;  95590425 ;  95590426 ;  95590427 ;  95590428 ;  95590429 ;  95590430 ;  95590431 ;  95590432 ;  95590433 ;  95590434 ;  95590435 ;  95590436 ;  95590437 ;  95590438 ;  95590439 ;  95590440 ;  95590441 ;  95590442 ;  95590443 ;  95590444 ;  95590445 ;  95590446 ;  95590447 ;  95590448 ;  95590449 ;  95590450 ;  95590451 ;  95590452 ;  95590453 ;  95590454 ;  95590455 ;  95590456 ;  95590457 ;  95590458 ;  95590459 ;  95590460 ;  95590461 ;  95590462 ;  95590463 ;  95590464 ;  95590465 ;  95590466 ;  95590467 ;  95590468 ;  95590469 ;  95590470 ;  95590471 ;  95590472 ;  95590473 ;  95590474 ;  95590475 ;  95590476 ;  95590477 ;  95590478 ;  95590479 ;  95590480 ;  95590481 ;  95590482 ;  95590483 ;  95590484 ;  95590485 ;  95590486 ;  95590487 ;  95590488 ;  95590489 ;  95590490 ;  95590491 ;  95590492 ;  95590493 ;  95590494 ;  95590495 ;  95590496 ;  95590497 ;  95590498 ;  95590499 ;  95590500 ;  95590501 ;  95590502 ;  95590503 ;  95590504 ;  95590505 ;  95590506 ;  95590507 ;  95590508 ;  95590509 ;  95590510 ;  95590511 ;  95590512 ;  95590513 ;  95590514 ;  95590515 ;  95590516 ;  95590517 ;  95590518 ;  95590519 ;  95590520 ;  95590521 ;  95590522 ;  95590523 ;  95590524 ;  95590525 ;  95590526 ;  95590527 ;  95590528 ;  95590529 ;  95590530 ;  95590531 ;  95590532 ;  95590533 ;  95590534 ;  95590535 ;  95590536 ;  95590537 ;  95590538 ;  95590539 ;  95590540 ;  95590541 ;  95590542 ;  95590543 ;  95590544 ;  95590545 ;  95590546 ;  95590547 ;  95590548 ;  95590549 ;  95590550 ;  95590551 ;  95590552 ;  95590553 ;  95590554 ;  95590555 ;  95590556 ;  95590557 ;  95590558 ;  95590559 ;  95590560 ;  95590561 ;  95590562 ;  95590563 ;  95590564 ;  95590566 ;  95590567 ;  95590568 ;  95590569 ;  95590570 ;  95590571 ;  95590572 ;  95590573 ;  95590574 ;  95590575 ;  95590576 ;  95590577 ;  95590578 ;  95590579 ;  95590580 ;  95590581 ;  95590582 ;  95590583 ;  95590584 ;  95590585 ;  95590586 ;  95590587 ;  95590588 ;  95590589 ;  95590590 ;  95590591 ;  95590592 ;  95590593 ;  95590594 ;  95590595 ;  95590596 ;  95590597 ;  95590598 ;  95590599 ;  95590600 ;  95590601 ;  95590602 ;  95590603 ;  95590604 ;  95590605 ;  95590606 ;  95590607 ;  95590608 ;  95590609 ;  95590610 ;  95590611 ;  95590612 ;  95590613 ;  95590614 ;  95590615 ;  95590616 ;  95590617 ;  95590618 ;  95590619 ;  95590620 ;  95590621 ;  95590622 ;  95590623 ;  95590624 ;  95590625 ;  95590626 ;  95590627 ;  95590628 ;  95590629 ;  95590630 ;  95590631 ;  95590632 ;  95590633 ;  95590634 ;  95590635 ;  95590636 ;  95590637 ;  95590638 ;  95590639 ;  95590640 ;  95590641 ;  95590642 ;  95590643 ;  95590644 ;  95590645 ;  95590646 ;  95590647 ;  95590648 ;  95590649 ;  95590650 ;  95590651 ;  95590652 ;  95590653 ;  95590654 ;  95590655 ;  95590656 ;  95590657 ;  95590658 ;  95590659 ;  95590660 ;  95590661 ;  95590662 ;  95590663 ;  95590664 ;  95590665 ;  95590666 ;  95590667 ;  95590668 ;  95590669 ;  95590670 ;  95590671 ;  95590672 ;  95590673 ;  95590674 ;  95590675 ;  95590676 ;  95590677 ;  95590678 ;  95590679 ;  95590680 ;  95590681 ;  95590682 ;  95590683 ;  95590684 ;  95590685 ;  95590686 ;  95590687 ;  95590688 ;  95590689 ;  95590690 ;  95590691 ;  95590692 ;  95590693 ;  95590694 ;  95590695 ;  95590696 ;  95590697 ;  95590698 ;  95590699 ;  95590700 ;  95590701 ;  95590702 ;  95590703 ;  95590704 ;  95590705 ;  95590706 ;  95590707 ;  95590708 ;  95590709 ;  95590710 ;  95590711 ;  95590712 ;  95590713 ;  95590714 ;  95590715 ;  95590716 ;  95590717 ;  95590718 ;  95590719 ;  95590720 ;  95590721 ;  95590722 ;  95590723 ;  95590724 ;  95590725 ;  95590726 ;  95590727 ;  95590728 ;  95590729 ;  95590730 ;  95590731 ;  95590732 ;  95590733 ;  95590734 ;  95590735 ;  95590736 ;  95590737 ;  95590738 ;  95590739 ;  95590740 ;  95590741 ;  95590742 ;  95590743 ;  95590744 ;  95590745 ;  95590746 ;  95590747 ;  95590748 ;  95590749 ;  95590750 ;  95590751 ;  95590752 ;  95590753 ;  95590754 ;  95590755 ;  95590756 ;  95590757 ;  95590758 ;  95590759 ;  95590760 ;  95590761 ;  95590762 ;  95590763 ;  95590764 ;  95590765 ;  95590766 ;  95590767 ;  95590768 ;  95590769 ;  95590770 ;  95590771 ;  95590772 ;  95590773 ;  95590774 ;  95590775 ;  95590776 ;  95590777 ;  95590778 ;  95590779 ;  95590780 ;  95590781 ;  95590782 ;  95590783 ;  95590784 ;  95590785 ;  95590786 ;  95590787 ;  95590788 ;  95590789 ;  95590790 ;  95590791 ;  95590792 ;  95590793 ;  95590794 ;  95590795 ;  95590796 ;  95590797 ;  95590798 ;  95590799 ;  95590803 ;  95590804 ;  95590807 ;  95590965 ;  95610001 ;  95610002 ;  95610003 ;  95610004 ;  95610005 ;  95610006 ;  95610007 ;  95610008 ;  95610009 ;  95610010 ;  95610011 ;  95610012 ;  95610013 ;  95610014 ;  95610015 ;  95610016 ;  95610017 ;  95610018 ;  95610019 ;  95610020 ;  95610021 ;  95610022 ;  95610023 ;  95610024 ;  95610025 ;  95610026 ;  95610027 ;  95610028 ;  95610029 ;  95610030 ;  95610031 ;  95610032 ;  95610033 ;  95610034 ;  95610035 ;  95610036 ;  95610037 ;  95610038 ;  95610039 ;  95610040 ;  95610041 ;  95610042 ;  95610043 ;  95610044 ;  95610045 ;  95610046 ;  95610047 ;  95610048 ;  95610049 ;  95610050 ;  95610051 ;  95610052 ;  95610053 ;  95610054 ;  95610055 ;  95610056 ;  95610057 ;  95610058 ;  95610059 ;  95610060 ;  95610061 ;  95610062 ;  95610063 ;  95610064 ;  95610065 ;  95610066 ;  95610067 ;  95610068 ;  95610069 ;  95610070 ;  95610071 ;  95610072 ;  95610073 ;  95610074 ;  95610075 ;  95610076 ;  95610077 ;  95610078 ;  95610079 ;  95610080 ;  95610081 ;  95610082 ;  95610083 ;  95610084 ;  95610085 ;  95610086 ;  95610087 ;  95610088 ;  95610089 ;  95610090 ;  95610091 ;  95610092 ;  95610093 ;  95610094 ;  95610095 ;  95610096 ;  95610097 ;  95610098 ;  95610099 ;  95610100 ;  95610101 ;  95610102 ;  95610103 ;  95610104 ;  95610105 ;  95610106 ;  95610107 ;  95610108 ;  95610109 ;  95610110 ;  95610111 ;  95610112 ;  95610113 ;  95610114 ;  95610115 ;  95610116 ;  95610117 ;  95610118 ;  95610119 ;  95610120 ;  95610121 ;  95610122 ;  95610123 ;  95610124 ;  95610125 ;  95610126 ;  95610127 ;  95610128 ;  95610129 ;  95610130 ;  95610131 ;  95610132 ;  95610133 ;  95610134 ;  95610135 ;  95610136 ;  95610137 ;  95610138 ;  95610139 ;  95610140 ;  95610141 ;  95610142 ;  95610143 ;  95610144 ;  95610145 ;  95610146 ;  95610147 ;  95610148 ;  95610149 ;  95610150 ;  95610151 ;  95610152 ;  95610153 ;  95610154 ;  95610155 ;  95610156 ;  95610157 ;  95610158 ;  95610159 ;  95610160 ;  95610161 ;  95610162 ;  95610163 ;  95610164 ;  95610165 ;  95610166 ;  95610167 ;  95610168 ;  95610169 ;  95610170 ;  95610171 ;  95610172 ;  95610173 ;  95610174 ;  95610175 ;  95610176 ;  95610177 ;  95610178 ;  95610179 ;  95610180 ;  95610181 ;  95610182 ;  95620001 ;  95620002 ;  95620003 ;  95620004 ;  95620005 ;  95620006 ;  95620007 ;  95620008 ;  95620009 ;  95620010 ;  95620011 ;  95620012 ;  95620013 ;  95620014 ;  95620015 ;  95620016 ;  95620017 ;  95620018 ;  95620019 ;  95620020 ;  95620021 ;  95620022 ;  95620023 ;  95620024 ;  95620025 ;  95620026 ;  95620027 ;  95620028 ;  95620029 ;  95620030 ;  95620031 ;  95620032 ;  95620033 ;  95620034 ;  95620035 ;  95620036 ;  95620037 ;  95620038 ;  95620039 ;  95620040 ;  95620041 ;  95620042 ;  95620043 ;  95620044 ;  95620045 ;  95620046 ;  95620047 ;  95620048 ;  95620049 ;  95620050 ;  95620051 ;  95620052 ;  95620053 ;  95620054 ;  95620055 ;  95620056 ;  95620057 ;  95620058 ;  95620059 ;  95620060 ;  95620061 ;  95620062 ;  95620063 ;  95620064 ;  95620065 ;  95620066 ;  95620067 ;  95620068 ;  95620069 ;  95620070 ;  95620071 ;  95620072 ;  95620073 ;  95620074 ;  95620075 ;  95620076 ;  95620077 ;  95620078 ;  95620079 ;  95620080 ;  95620081 ;  95620082 ;  95620083 ;  95620084 ;  95620085 ;  95620086 ;  95620087 ;  95620088 ;  95620089 ;  95620090 ;  95620091 ;  95620092 ;  95620093 ;  95620094 ;  95620095 ;  95620096 ;  95620097 ;  95620098 ;  95620099 ;  95620100 ;  95620101 ;  95620102 ;  95620103 ;  95620104 ;  95620105 ;  95620106 ;  95620107 ;  95620108 ;  95620109 ;  95620110 ;  95620111 ;  95620112 ;  95620113 ;  95620114 ;  95620115 ;  95620116 ;  95620117 ;  95620118 ;  95620119 ;  95620120 ;  95620121 ;  95620122 ;  95620123 ;  95620124 ;  95620125 ;  95620126 ;  95620127 ;  95620128 ;  95620129 ;  95620130 ;  95620131 ;  95620132 ;  95620133 ;  95620134 ;  95620135 ;  95620136 ;  95620137 ;  95620138 ;  95620139 ;  95620140 ;  95620141 ;  95620142 ;  95620143 ;  95620144 ;  95620145 ;  95620146 ;  95620147 ;  95620148 ;  95620149 ;  95620150 ;  95620151 ;  95620152 ;  95620153 ;  95620154 ;  95620155 ;  95620156 ;  95620157 ;  95620158 ;  95620159 ;  95620160 ;  95620161 ;  95620162 ;  95620163 ;  95620164 ;  95620165 ;  95620166 ;  95620167 ;  95620169 ;  95620170 ;  95620171 ;  95620172 ;  95620173 ;  95620174 ;  95620176 ;  95620177 ;  95620178 ;  95620179 ;  95620180 ;  95620181 ;  95620182 ;  95620183 ;  95620185 ;  95620186 ;  95620187 ;  95620188 ;  95620189 ;  95620190 ;  95620191 ;  95620192 ;  95620193 ;  95620194 ;  95620195 ;  95620196 ;  95620197 ;  95620198 ;  95620199 ;  95620200 ;  95620201 ;  95620202 ;  95620203 ;  95620204 ;  95620205 ;  95620206 ;  95620207 ;  95620208 ;  95620209 ;  95620210 ;  95620211 ;  95620212 ;  95620213 ;  95620214 ;  95620215 ;  95620216 ;  95620217 ;  95620218 ;  95620219 ;  95620220 ;  95620221 ;  95620222 ;  95620223 ;  95620224 ;  95620225 ;  95620226 ;  95620227 ;  95620228 ;  95620229 ;  95620230 ;  95620231 ;  95620232 ;  95620233 ;  95620234 ;  95620235 ;  95620236 ;  95620237 ;  95620238 ;  95620239 ;  95620240 ;  95620241 ;  95620242 ;  95620243 ;  95620244 ;  95620245 ;  95620246 ;  95620247 ;  95620248 ;  95620249 ;  95620250 ;  95620251 ;  95620252 ;  95620288 ;  95620289 ;  95620290 ;  95620291 ;  95620292 ;  95620293 ;  95620294 ;  95620295 ;  95620296 ;  95620297 ;  95620298 ;  95620299 ;  95620300 ;  95620301 ;  95620302 ;  95620303 ;  95620304 ;  95620305 ;  95620306 ;  95620307 ;  95620308 ;  95620309 ;  95620310 ;  95620311 ;  95620312 ;  95620313 ;  95620314 ;  95620315 ;  95620316 ;  95620317 ;  95620318 ;  95620319 ;  95620320 ;  95620321 ;  95620322 ;  95620323 ;  95620324 ;  95620325 ;  95620326 ;  95620327 ;  95620328 ;  95620329 ;  95620330 ;  95620331 ;  95620332 ;  95620333 ;  95620334 ;  95620335 ;  95620336 ;  95620337 ;  95620338 ;  95620339 ;  95620340 ;  95620341 ;  95620342 ;  95620343 ;  95620344 ;  95620345 ;  95620346 ;  95620347 ;  95620348 ;  95620349 ;  95620350 ;  95620351 ;  95620352 ;  95620353 ;  95620354 ;  95620355 ;  95620356 ;  95620357 ;  95620358 ;  95620359 ;  95620360 ;  95620361 ;  95620362 ;  95620363 ;  95620364 ;  95620365 ;  95620366 ;  95620367 ;  95620368 ;  95620369 ;  95620370 ;  95620371 ;  95620372 ;  95620373 ;  95620374 ;  95620375 ;  95620376 ;  95620377 ;  95620378 ;  95620379 ;  95620380 ;  95620381 ;  95620382 ;  95620383 ;  95620384 ;  95620385 ;  95620386 ;  95620387 ;  95620388 ;  95620389 ;  95620390 ;  95620391 ;  95620392 ;  95620393 ;  95620394 ;  95620395 ;  95620396 ;  95620397 ;  95620398 ;  95620399 ;  95620400 ;  95620401 ;  95620402 ;  95620403 ;  95620404 ;  95620405 ;  95620406 ;  95620407 ;  95620408 ;  95620409 ;  95620410 ;  95620411 ;  95620412 ;  95620413 ;  95620414 ;  95620415 ;  95620416 ;  95620417 ;  95620418 ;  95620419 ;  95620420 ;  95620421 ;  95620457 ;  95620458 ;  95620459 ;  95620460 ;  95620461 ;  95620462 ;  95620463 ;  95620464 ;  95620465 ;  95620466 ;  95620467 ;  95620468 ;  95620469 ;  95620470 ;  95620471 ;  95620472 ;  95620473 ;  95620474 ;  95620475 ;  95620476 ;  95620477 ;  95620478 ;  95620479 ;  95620480 ;  95620481 ;  95620482 ;  95620483 ;  95620484 ;  95620485 ;  95620486 ;  95620487 ;  95620488 ;  95620489 ;  95620490 ;  95620491 ;  95620492 ;  95620493 ;  95620494 ;  95620495 ;  95620496 ;  95620497 ;  95620498 ;  95620499 ;  95620500 ;  95620501 ;  95620502 ;  95620503 ;  95620504 ;  95620505 ;  95620506 ;  95620507 ;  95620508 ;  95620509 ;  95620510 ;  95620511 ;  95620512 ;  95620513 ;  95620514 ;  95620515 ;  95620516 ;  95620517 ;  95620518 ;  95620519 ;  95620520 ;  95620521 ;  95620522 ;  95620523 ;  95620524 ;  95620525 ;  95620526 ;  95620527 ;  95620528 ;  95620529 ;  95620530 ;  95620531 ;  95620532 ;  95620533 ;  95620534 ;  95620535 ;  95620536 ;  95620537 ;  95620538 ;  95620539 ;  95620540 ;  95620541 ;  95620542 ;  95620543 ;  95620544 ;  95620545 ;  95620546 ;  95620547 ;  95620548 ;  95620549 ;  95620550 ;  95620551 ;  95620552 ;  95620553 ;  95620554 ;  95620555 ;  95620556 ;  95620557 ;  95620558 ;  95620559 ;  95620560 ;  95620561 ;  95620562 ;  95620563 ;  95620564 ;  95620565 ;  95620566 ;  95620567 ;  95620568 ;  95620569 ;  95620570 ;  95620571 ;  95620572 ;  95620573 ;  95620574 ;  95620575 ;  95620576 ;  95620577 ;  95620578 ;  95620579 ;  95620580 ;  95620581 ;  95620582 ;  95620583 ;  95620584 ;  95620585 ;  95620586 ;  95620587 ;  95620588 ;  95620589 ;  95620590 ;  95620591 ;  95620592 ;  95620593 ;  95620594 ;  95620595 ;  95620596 ;  95620597 ;  95620598 ;  95620599 ;  95620600 ;  95620601 ;  95620602 ;  95620603 ;  95620604 ;  95620605 ;  95620606 ;  95620607 ;  95620608 ;  95620609 ;  95620610 ;  95620611 ;  95620612 ;  95620613 ;  95620614 ;  95620615 ;  95620616 ;  95620617 ;  95620618 ;  95620619 ;  95620620 ;  95620621 ;  95620622 ;  95620623 ;  95620624 ;  95620625 ;  95620626 ;  95620627 ;  95620628 ;  95620629 ;  95620630 ;  95620631 ;  95620632 ;  95620633 ;  95620634 ;  95620635 ;  95620636 ;  95620637 ;  95620638 ;  95620639 ;  95620640 ;  95620641 ;  95620642 ;  95620643 ;  95620644 ;  95620645 ;  95620646 ;  95620647 ;  95620648 ;  95620649 ;  95620650 ;  95620651 ;  95620652 ;  95620653 ;  95620654 ;  95620656 ;  95620657 ;  95620658 ;  95620659 ;  95620660 ;  95620661 ;  95620662 ;  95620663 ;  95620664 ;  95620665 ;  95620666 ;  95620667 ;  95620668 ;  95620669 ;  95620670 ;  95620671 ;  95620672 ;  95620673 ;  95620674 ;  95620675 ;  95620676 ;  95620677 ;  95620678 ;  95620679 ;  95620680 ;  95620681 ;  95620682 ;  95620683 ;  95620684 ;  95620685 ;  95620686 ;  95620687 ;  95620688 ;  95620689 ;  95620690 ;  95620691 ;  95620692 ;  95620693 ;  95620694 ;  95620695 ;  95620696 ;  95620697 ;  95620716 ;  95620717 ;  95620718 ;  95620719 ;  95620720 ;  95620721 ;  95620722 ;  95620723 ;  95620724 ;  95620725 ;  95620726 ;  95620727 ;  95620728 ;  95620729 ;  95620730 ;  95620731 ;  95620732 ;  95620733 ;  95620734 ;  95620735 ;  95620736 ;  95620737 ;  95620738 ;  95620739 ;  95620740 ;  95620741 ;  95620742 ;  95620743 ;  95620744 ;  95620745 ;  95620746 ;  95620747 ;  95620748 ;  95620749 ;  95620750 ;  95620751 ;  95620752 ;  95620753 ;  95620754 ;  95620755 ;  95620756 ;  95620757 ;  95620758 ;  95620759 ;  95620760 ;  95620761 ;  95620762 ;  95620763 ;  95620764 ;  95620765 ;  95620766 ;  95620767 ;  95620768 ;  95620769 ;  95620770 ;  95620771 ;  95620772 ;  95620773 ;  95620774 ;  95620775 ;  95620776 ;  95620777 ;  95620778 ;  95620779 ;  95620780 ;  95620781 ;  95620782 ;  95620783 ;  95620784 ;  95620785 ;  95620786 ;  95620787 ;  95620788 ;  95620789 ;  95620790 ;  95620791 ;  95620792 ;  95620793 ;  95620794 ;  95620795 ;  95620796 ;  95620797 ;  95620798 ;  95620799 ;  95620800 ;  95620801 ;  95620802 ;  95620803 ;  95620804 ;  95620805 ;  95620806 ;  95620807 ;  95620808 ;  95620809 ;  95620810 ;  95620811 ;  95620812 ;  95620813 ;  95620814 ;  95620815 ;  95620816 ;  95620817 ;  95620818 ;  95620819 ;  95620820 ;  95620821 ;  95620822 ;  95620823 ;  95620824 ;  95620825 ;  95620826 ;  95620827 ;  95620828 ;  95620829 ;  95620830 ;  95620831 ;  95620832 ;  95620833 ;  95620834 ;  95620835 ;  95620836 ;  95620837 ;  95620838 ;  95620839 ;  95620840 ;  95620841 ;  95620842 ;  95620843 ;  95620844 ;  95620845 ;  95620846 ;  95620847 ;  95620848 ;  95620849 ;  95620850 ;  95620851 ;  95620852 ;  95620853 ;  95620854 ;  95620855 ;  95620856 ;  95620857 ;  95620858 ;  95620859 ;  95620860 ;  95620861 ;  95620862 ;  95620863 ;  95620864 ;  95620865 ;  95620866 ;  95620867 ;  95620868 ;  95620869 ;  95620870 ;  95620871 ;  95620872 ;  95620873 ;  95620874 ;  95620875 ;  95620876 ;  95620877 ;  95620878 ;  95620879 ;  95620880 ;  95620881 ;  95620882 ;  95620883 ;  95620884 ;  95620885 ;  95620886 ;  95620887 ;  95620888 ;  95620889 ;  95620890 ;  95620891 ;  95620892 ;  95620893 ;  95620894 ;  95620895 ;  95620896 ;  95620897 ;  95620898 ;  95620899 ;  95620900 ;  95620901 ;  95620902 ;  95620903 ;  95620904 ;  95620905 ;  95620906 ;  95620907 ;  95620908 ;  95620909 ;  95620910 ;  95620911 ;  95620912 ;  95620913 ;  95620914 ;  95620915 ;  95620916 ;  95620917 ;  95620918 ;  95620919 ;  95620920 ;  95620921 ;  95620922 ;  95620923 ;  95620924 ;  95620925 ;  95620926 ;  95620927 ;  95620928 ;  95620929 ;  95620930 ;  95620931 ;  95620932 ;  95620933 ;  95620934 ;  95620935 ;  95620936 ;  95620937 ;  95620938 ;  95620939 ;  95620940 ;  95620941 ;  95620942 ;  95620943 ;  95620944 ;  95620945 ;  95620946 ;  95620947 ;  95620948 ;  95620949 ;  95620950 ;  95620951 ;  95620952 ;  95620953 ;  95620954 ;  95620955 ;  95620956 ;  95620957 ;  95620958 ;  95620959 ;  95620960 ;  95620961 ;  95620962 ;  95620970 ;  95620971 ;  95620972 ;  95620973 ;  95620975 ;  95620977 ;  95620978 ;  95620979 ;  95620980 ;  95620981 ;  95620984 ;  95620986 ;  95620987 ;  95620988 ;  95620989 ;  95620990 ;  95620994 ;  95620995 ;  95620996 ;  95620997 ;  95620998 ;  95620999 ;  95621000 ;  95621001 ;  95621002 ;  95621003 ;  95630003 ;  95630029 ;  95630032 ;  95630038 ;  95630041 ;  95630049 ;  95630051 ;  95630052 ;  95630055 ;  95630056 ;  95630063 ;  95630065 ;  95630070 ;  95630075 ;  95630078 ;  95630080 ;  95630081 ;  95630086 ;  95630087 ;  95630088 ;  95630091 ;  95630092 ;  95630094 ;  95630102 ;  95630104 ;  95630106 ;  95630107 ;  95630108 ;  95630110 ;  95630111 ;  95630112 ;  95630116 ;  95630117 ;  95630121 ;  95630122 ;  95630123 ;  95630129 ;  95630130 ;  95630131 ;  95630134 ;  95630137 ;  95630138 ;  95630143 ;  95630148 ;  95630149 ;  95630150 ;  95630151 ;  95630152 ;  95630153 ;  95630154 ;  95630155 ;  95630156 ;  95630159 ;  95630160 ;  95630161 ;  95630162 ;  95630163 ;  95630164 ;  95630165 ;  95630166 ;  95630168 ;  95630169 ;  95630172 ;  95630173 ;  95630174 ;  95630175 ;  95630176 ;  95630178 ;  95630179 ;  95630180 ;  95630183 ;  95630184 ;  95630185 ;  95630187 ;  95630188 ;  95630189 ;  95630190 ;  95630191 ;  95630193 ;  95630196 ;  95630197 ;  95630198 ;  95630199 ;  95630200 ;  95630202 ;  95630203 ;  95630204 ;  95630205 ;  95630206 ;  95630207 ;  95630209 ;  95630210 ;  95630213 ;  95630217 ;  95630218 ;  95630219 ;  95630220 ;  95630221 ;  95630222 ;  95630225 ;  95630226 ;  95630227 ;  95630229 ;  95630230 ;  95630231 ;  95630232 ;  95630233 ;  95630234 ;  95630236 ;  95630238 ;  95630244 ;  95630246 ;  95630248 ;  95630250 ;  95630251 ;  95630252 ;  95630255 ;  95630256 ;  95630257 ;  95630260 ;  95630262 ;  95630270 ;  95630274 ;  95630285 ;  95630286 ;  95630290 ;  95630296 ;  95630300 ;  95630301 ;  95630302 ;  95630303 ;  95630306 ;  95630308 ;  95630309 ;  95630314 ;  95630315 ;  95630316 ;  95630317 ;  95630318 ;  95630319 ;  95630322 ;  95630324 ;  95630326 ;  95630328 ;  95630330 ;  95630331 ;  95630334 ;  95630337 ;  95630339 ;  95630342 ;  95630343 ;  95630349 ;  95630358 ;  95630359 ;  95630360 ;  95630372 ;  95630373 ;  95630374 ;  95630379 ;  95630383 ;  95630386 ;  95630387 ;  95630388 ;  95630389 ;  95630390 ;  95630393 ;  95630394 ;  95630395 ;  95630396 ;  95630397 ;  95630398 ;  95630403 ;  95630405 ;  95630406 ;  95630407 ;  95630408 ;  95630411 ;  95630414 ;  95630429 ;  95630437 ;  95630439 ;  95630440 ;  95630442 ;  95630444 ;  95630447 ;  95630448 ;  95630451 ;  95630452 ;  95630457 ;  95630460 ;  95630461 ;  95630466 ;  95630468 ;  95630469 ;  95630471 ;  95630473 ;  95630474 ;  95630475 ;  95630476 ;  95630481 ;  95630482 ;  95630484 ;  95630487 ;  95630493 ;  95630497 ;  95630498 ;  95630499 ;  95630500 ;  95630502 ;  95630503 ;  95630504 ;  95630505 ;  95630506 ;  95630507 ;  95630508 ;  95630511 ;  95630517 ;  95630518 ;  95630519 ;  95630520 ;  95630523 ;  95630525 ;  95630526 ;  95630527 ;  95630528 ;  95630530 ;  95630531 ;  95630532 ;  95630533 ;  95630534 ;  95630536 ;  95630537 ;  95630538 ;  95630539 ;  95630540 ;  95630541 ;  95630542 ;  95630544 ;  95630546 ;  95630548 ;  95630549 ;  95630550 ;  95630556 ;  95630557 ;  95630558 ;  95630559 ;  95630561 ;  95630588 ;  95630589 ;  95630591 ;  95630592 ;  95630593 ;  95630594 ;  95630596 ;  95630600 ;  95630602 ;  95630605 ;  95630614 ;  95630624 ;  95630625 ;  95630626 ;  95630628 ;  95630629 ;  95630630 ;  95630634 ;  95630635 ;  95630636 ;  95630637 ;  95630638 ;  95630639 ;  95630640 ;  95630641 ;  95630652 ;  95630661 ;  95630662 ;  95630663 ;  95630664 ;  95630668 ;  95630670 ;  95630671 ;  95630672 ;  95630674 ;  95630675 ;  95630681 ;  95630686 ;  95630691 ;  95630692 ;  95630695 ;  95630696 ;  95630698 ;  95630699 ;  95630700 ;  95630701 ;  95630706 ;  95630707 ;  95630708 ;  95630714 ;  95630717 ;  95630719 ;  95630720 ;  95630723 ;  95630724 ;  95630728 ;  95630731 ;  95630739 ;  95630740 ;  95630741 ;  95630744 ;  95630747 ;  95630751 ;  95630754 ;  95630760 ;  95630765 ;  95630766 ;  95630768 ;  95630776 ;  95630785 ;  95630793 ;  95630794 ;  95630800 ;  95630803 ;  95630804 ;  95630806 ;  95630811 ;  95630814 ;  95630832 ;  95630834 ;  95630835 ;  95630837 ;  95630844 ;  95630845 ;  95630846 ;  95630848 ;  95630849 ;  95630850 ;  95630852 ;  95630862 ;  95630863 ;  95630869 ;  95630873 ;  95630879 ;  95630881 ;  95630900 ;  95630901 ;  95630902 ;  95630903 ;  95630906 ;  95630907 ;  95630908 ;  95630912 ;  95630918 ;  95630919 ;  95630926 ;  95630927 ;  95630928 ;  95630929 ;  95630930 ;  95630933 ;  95630934 ;  95630935 ;  95630936 ;  95630938 ;  95630939 ;  95630940 ;  95630946 ;  95630947 ;  95630950 ;  95630952 ;  95630959 ;  95630965 ;  95630970 ;  95630971 ;  95630975 ;  95630976 ;  95630979 ;  95630980 ;  95630981 ;  95630982 ;  95630983 ;  95630984 ;  95630985 ;  95630990 ;  95630991 ;  95630992 ;  95630993 ;  95630995 ;  95631000 ;  95631003 ;  95631004 ;  95631005 ;  95631007 ;  95631009 ;  95631011 ;  95631012 ;  95631014 ;  95631018 ;  95631019 ;  95631020 ;  95631021 ;  95631023 ;  95631024 ;  95631026 ;  95631027 ;  95631028 ;  95631034 ;  95631037 ;  95631038 ;  95631039 ;  95631041 ;  95631044 ;  95631046 ;  95631048 ;  95631049 ;  95631055 ;  95631056 ;  95631058 ;  95631060 ;  95631061 ;  95631076 ;  95631077 ;  95631082 ;  95631083 ;  95631093 ;  95631094 ;  95631095 ;  95631097 ;  95631098 ;  95631099 ;  95631100 ;  95631101 ;  95631102 ;  95631103 ;  95631104 ;  95631105 ;  95631106 ;  95631107 ;  95631108 ;  95631109 ;  95631110 ;  95631111 ;  95631112 ;  95631113 ;  95631114 ;  95631116 ;  95631117 ;  95631119 ;  95631124 ;  95631126 ;  95631128 ;  95631131 ;  95631133 ;  95640013 ;  95640017 ;  95640019 ;  95640020 ;  95640022 ;  95640025 ;  95640032 ;  95640041 ;  95640043 ;  95640044 ;  95640045 ;  95640047 ;  95640049 ;  95640053 ;  95640056 ;  95640058 ;  95640060 ;  95640063 ;  95640064 ;  95640067 ;  95640068 ;  95640069 ;  95640070 ;  95640071 ;  95640072 ;  95640073 ;  95640074 ;  95640076 ;  95640077 ;  95640078 ;  95640080 ;  95640081 ;  95640082 ;  95640083 ;  95640084 ;  95640085 ;  95640086 ;  95640087 ;  95640088 ;  95640089 ;  95640090 ;  95640091 ;  95640092 ;  95640093 ;  95640094 ;  95640095 ;  95640096 ;  95640097 ;  95640098 ;  95640099 ;  95640101 ;  95640102 ;  95640103 ;  95640105 ;  95640106 ;  95640107 ;  95640108 ;  95640109 ;  95640110 ;  95640111 ;  95640112 ;  95640113 ;  95640114 ;  95640117 ;  95640118 ;  95640119 ;  95640122 ;  95640123 ;  95640124 ;  95640125 ;  95640126 ;  95640127 ;  95640129 ;  95640130 ;  95640131 ;  95640132 ;  95640133 ;  95640134 ;  95640135 ;  95640136 ;  95640137 ;  95640138 ;  95640140 ;  95640141 ;  95640142 ;  95640143 ;  95640144 ;  95640145 ;  95640146 ;  95640147 ;  95640148 ;  95640149 ;  95640150 ;  95640151 ;  95640152 ;  95640154 ;  95640155 ;  95640156 ;  95640158 ;  95640159 ;  95640160 ;  95640161 ;  95640162 ;  95640167 ;  95640168 ;  95640170 ;  95640171 ;  95640172 ;  95640173 ;  95640174 ;  95640175 ;  95640176 ;  95640177 ;  95640178 ;  95640179 ;  95640180 ;  95640181 ;  95640182 ;  95640183 ;  95640184 ;  95640190 ;  95640195 ;  95640196 ;  95640197 ;  95640202 ;  95640203 ;  95640204 ;  95640205 ;  95640206 ;  95640209 ;  95640210 ;  95640211 ;  95640212 ;  95640213 ;  95640214 ;  95640215 ;  95640216 ;  95640217 ;  95640218 ;  95640219 ;  95640220 ;  95640221 ;  95640222 ;  95640223 ;  95640224 ;  95640225 ;  95640226 ;  95640227 ;  95640228 ;  95640229 ;  95640230 ;  95640231 ;  95640232 ;  95640234 ;  95640238 ;  95640239 ;  95640240 ;  95640241 ;  95640242 ;  95640243 ;  95640244 ;  95640245 ;  95640246 ;  95640247 ;  95640248 ;  95640249 ;  95640250 ;  95640251 ;  95640253 ;  95640254 ;  95640255 ;  95640256 ;  95640257 ;  95640258 ;  95640259 ;  95640260 ;  95640261 ;  95640262 ;  95640265 ;  95640266 ;  95640268 ;  95640270 ;  95640271 ;  95640272 ;  95640273 ;  95640274 ;  95640275 ;  95640276 ;  95640277 ;  95640278 ;  95640279 ;  95640280 ;  95640281 ;  95640282 ;  95640283 ;  95640284 ;  95640289 ;  95640290 ;  95640291 ;  95640292 ;  95640293 ;  95640294 ;  95640295 ;  95640296 ;  95640297 ;  95640298 ;  95640299 ;  95640300 ;  95640301 ;  95640302 ;  95640303 ;  95640305 ;  95640306 ;  95640311 ;  95640312 ;  95640315 ;  95640316 ;  95640317 ;  95640318 ;  95640319 ;  95640320 ;  95640321 ;  95640349 ;  95640350 ;  95640352 ;  95640353 ;  95650001 ;  95650002 ;  95650003 ;  95650004 ;  95650005 ;  95650006 ;  95650007 ;  95650008 ;  95650009 ;  95650010 ;  95650011 ;  95650012 ;  95650013 ;  95650014 ;  95650015 ;  95650016 ;  95650017 ;  95650018 ;  95650019 ;  95650020 ;  95650021 ;  95650022 ;  95650023 ;  95650024 ;  95650025 ;  95650026 ;  95650027 ;  95650028 ;  95650029 ;  95650030 ;  95650031 ;  95650032 ;  95650033 ;  95650034 ;  95650035 ;  95650036 ;  95650037 ;  95650038 ;  95650039 ;  95650040 ;  95650041 ;  95650042 ;  95650043 ;  95650044 ;  95650045 ;  95650046 ;  95650047 ;  95650048 ;  95650049 ;  95650050 ;  95650051 ;  95650052 ;  95650053 ;  95650054 ;  95650055 ;  95650056 ;  95650057 ;  95650058 ;  95650059 ;  95650060 ;  95650061 ;  95650062 ;  95650063 ;  95650064 ;  95650065 ;  95650066 ;  95650067 ;  95650068 ;  95650069 ;  95650070 ;  95650071 ;  95650072 ;  95650073 ;  95650074 ;  95650075 ;  95650076 ;  95650077 ;  95650078 ;  95650079 ;  95650080 ;  95650081 ;  95650082 ;  95650083 ;  95650084 ;  95650085 ;  95650086 ;  95650087 ;  95650088 ;  95650089 ;  95650090 ;  95650091 ;  95650092 ;  95650093 ;  95650094 ;  95650095 ;  95650096 ;  95650097 ;  95650098 ;  95650099 ;  95650100 ;  95650101 ;  95650102 ;  95650103 ;  95650104 ;  95650105 ;  95650106 ;  95650107 ;  95650108 ;  95650109 ;  95650110 ;  95650111 ;  95650112 ;  95650113 ;  95650114 ;  95650115 ;  95650116 ;  95650117 ;  95650118 ;  95650119 ;  95650120 ;  95650121 ;  95650122 ;  95650123 ;  95650124 ;  95650125 ;  95650126 ;  95650127 ;  95650128 ;  95650129 ;  95650130 ;  95650131 ;  95650132 ;  95650133 ;  95650134 ;  95650135 ;  95650136 ;  95650137 ;  95650138 ;  95650139 ;  95650140 ;  95650141 ;  95650142 ;  95670001 ;  95670002 ;  95670003 ;  95670005 ;  95670007 ;  95670009 ;  95670011 ;  95670012 ;  95670013 ;  95670014 ;  95670015 ;  95670016 ;  95670017 ;  95670019 ;  95670020 ;  95670021 ;  95670022 ;  95670023 ;  95670024 ;  95670026 ;  95670028 ;  95670029 ;  95670030 ;  95670031 ;  95670033 ;  95670034 ;  95670035 ;  95670036 ;  95670037 ;  95670038 ;  95670039 ;  95670040 ;  95670041 ;  95670042 ;  95670043 ;  95670044 ;  95670045 ;  95670046 ;  95670047 ;  95670049 ;  95670050 ;  95670052 ;  95670053 ;  95670054 ;  95670055 ;  95670057 ;  95670058 ;  95670061 ;  95670062 ;  95670063 ;  95670064 ;  95670065 ;  95670067 ;  95670070 ;  95670072 ;  95670073 ;  95670074 ;  95670075 ;  95670076 ;  95670077 ;  95670078 ;  95670079 ;  95670080 ;  95670081 ;  95670082 ;  95670083 ;  95670084 ;  95670085 ;  95670086 ;  95670087 ;  95670088 ;  95670089 ;  95670090 ;  95670091 ;  95670092 ;  95670093 ;  95670094 ;  95670097 ;  95670098 ;  95670099 ;  95670100 ;  95670101 ;  95670102 ;  95670103 ;  95670104 ;  95670105 ;  95670106 ;  95670107 ;  95670108 ;  95670109 ;  95670110 ;  95670111 ;  95670112 ;  95670113 ;  95670114 ;  95670115 ;  95670116 ;  95670117 ;  95670118 ;  95670119 ;  95670121 ;  95670122 ;  95670123 ;  95670124 ;  95670125 ;  95670126 ;  95670127 ;  95670128 ;  95670129 ;  95670130 ;  95670131 ;  95670132 ;  95670133 ;  95670134 ;  95670135 ;  95670136 ;  95670137 ;  95670138 ;  95670139 ;  95670140 ;  95670141 ;  95670142 ;  95670143 ;  95670144 ;  95670145 ;  95670146 ;  95670147 ;  95670148 ;  95670149 ;  95670150 ;  95670151 ;  95670152 ;  95670153 ;  95670154 ;  95670155 ;  95670156 ;  95670157 ;  95670158 ;  95670159 ;  95670160 ;  95670161 ;  95670162 ;  95670163 ;  95670164 ;  95670165 ;  95670166 ;  95670167 ;  95670169 ;  95670171 ;  95670172 ;  95670173 ;  95670174 ;  95670175 ;  95670176 ;  95670177 ;  95670178 ;  95670179 ;  95670180 ;  95670181 ;  95670182 ;  95670183 ;  95670184 ;  95670185 ;  95670186 ;  95670187 ;  95670188 ;  95670189 ;  95670190 ;  95670191 ;  95670192 ;  95670193 ;  95670194 ;  95670195 ;  95670196 ;  95670197 ;  95670198 ;  95670199 ;  95670200 ;  95670201 ;  95670202 ;  95670203 ;  95670204",
      "recall_initiation_date": "20130308",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}