  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lugano",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85444",
      "recalling_firm": "Inpeco S.A.",
      "address_1": "Via San Gottardo 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide including in the states of IL, NY.",
      "recall_number": "Z-2140-2020",
      "product_description": "FlexLab Automation System with Advia2120LAS Interface Module (P/N FLX-219-00, FLX-219-10), ImmunoCAP 1000 Interface Module (P/N FLX-226-01, FLX-226-10), or StaRRsed Interface Module (P/N FLX-268-00)is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.",
      "product_quantity": "37 systems",
      "reason_for_recall": "The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.",
      "recall_initiation_date": "20200407",
      "center_classification_date": "20200528",
      "termination_date": "20240206",
      "report_date": "20200603",
      "code_info": "Serial Numbers:   FLX.005  FLX.006  FLX.011  FLX.0117  FLX.0121  FLX.0124  FLX.0126  FLX.0129  FLX.0133  FLX.0136  FLX.0139  FLX.0140  FLX.0147  FLX.0149  FLX.0153  FLX.0157  FLX.016  FLX.0165  FLX.0168  FLX.0173  FLX.0189  FLX.0191  FLX.0192  FLX.022  FLX.030  FLX.032  FLX.035  FLX.036  FLX.037  FLX.050  FLX.056  FLX.060  FLX.062  FLX.065  FLX.082  FLX.096  SHD.002"
    }
  ]
}