  "results": [
    {
      "status": "Terminated",
      "city": "Pomona",
      "state": "CA",
      "country": "United States",
      "classification": "Class III",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80096",
      "recalling_firm": "Alere San Diego Inc. DBA Immunalysis Corporation",
      "address_1": "829 Towne Center Dr",
      "address_2": "N/A",
      "postal_code": "91767-5901",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : CA, CO, CT, FL, LA, MD, NC, NJ, OH, OK, TN, TX, WI, and WV",
      "recall_number": "Z-2140-2018",
      "product_description": "lmmunalysis EDDP Urine Control Set 1    The EDDP Urine Controls (High 125 ng/ml and Low 75 ng/ml) are used as control materials in the ammonolysis EDDP Specific Urine Enzyme Immunoassay, which has a cutoff of 100 ng/ml. This test provides only a preliminary analytical test result and a more specific alternate chemical method, such as Gas Chromatography/Mass Spectrometry or Liquid Chromatography/Tandem Mass Spectrometry, must be used to obtain a confirmed analytical result.",
      "product_quantity": "37 kits",
      "reason_for_recall": "The High (125 ng/ml) Control does not meet specification - it has been confirmed to produce values below 100 ng/mL when run on validated clinical chemistry analyzers, which is below the 100 ng/mL cutoff of the EDDP Specific Urine Enzyme Immunoassay with which this control is utilized.",
      "recall_initiation_date": "20171128",
      "center_classification_date": "20180608",
      "termination_date": "20210129",
      "report_date": "20180620",
      "code_info": "Part Number: 3006-5, Lot Number: EK16135"
    }
  ]
}