  "results": [
    {
      "country": "United States",
      "city": "Lake Forest",
      "address_1": "275 N Field Dr",
      "reason_for_recall": "Hospira has received reports of PCA pumps not detecting distal occlusions, one of which resulted in a serious injury. The issue is caused by normal wear and tear on the Half Nut (the component/nut that travels up and down the lead screw) which prevents it from properly detecting the pressure build-up associated with a distal occlusion.",
      "address_2": "",
      "product_quantity": "8,216 infusion pump systems",
      "code_info": "List No. 12384; Serial Numbers: 12920041 ;  12920048 ;  12920052 ;  12920056 ;  12920066 ;  12920069 ;  12920079 ;  12920082 ;  12920101 ;  12920102 ;  12920104 ;  12920105 ;  12920106 ;  12920107 ;  12920117 ;  12920121 ;  12920122 ;  12920152 ;  12920158 ;  12920159 ;  12920160 ;  12920174 ;  12920246 ;  12920248 ;  12920251 ;  12920252 ;  12920255 ;  12920257 ;  12920264 ;  12920265 ;  12920268 ;  12920270 ;  12920275 ;  12920276 ;  12920277 ;  12920278 ;  12920282 ;  12920287 ;  12920289 ;  12920291 ;  12920294 ;  12920304 ;  12920305 ;  12920311 ;  12920312 ;  12920320 ;  12920328 ;  12920330 ;  12920332 ;  12920340 ;  12920343 ;  12920346 ;  12920347 ;  12920352 ;  12920353 ;  12920357 ;  12920358 ;  12920359 ;  12920360 ;  12920362 ;  12920364 ;  12920369 ;  12920371 ;  12920372 ;  12920377 ;  12920381 ;  12920382 ;  12920383 ;  12920384 ;  12920387 ;  12920388 ;  12920390 ;  12920392 ;  12920398 ;  12920401 ;  12920403 ;  12920404 ;  12920405 ;  12920410 ;  12920412 ;  12920413 ;  12920415 ;  12920416 ;  12920420 ;  12920424 ;  12920425 ;  12920428 ;  12920429 ;  12920432 ;  12920433 ;  12920434 ;  12920435 ;  12920436 ;  12920441 ;  12920444 ;  12920446 ;  12920448 ;  12920449 ;  12920453 ;  12920455 ;  12920457 ;  12920458 ;  12920460 ;  12920461 ;  12920462 ;  12920466 ;  12920467 ;  12920468 ;  12920469 ;  12920472 ;  12920473 ;  12920476 ;  12920477 ;  12920478 ;  12920480 ;  12920481 ;  12920482 ;  12920484 ;  12920485 ;  12920486 ;  12920487 ;  12920488 ;  12920489 ;  12920490 ;  12920492 ;  12920496 ;  12920497 ;  12920498 ;  12920499 ;  12920501 ;  12920505 ;  12920506 ;  12920507 ;  12920508 ;  12920509 ;  12920510 ;  12920514 ;  12920516 ;  12920517 ;  12920518 ;  12920519 ;  12920520 ;  12920521 ;  12920522 ;  12920523 ;  12920524 ;  12920525 ;  12920526 ;  12920530 ;  12920533 ;  12920534 ;  12920535 ;  12920537 ;  12920540 ;  12920541 ;  12920542 ;  12920544 ;  12920545 ;  12920546 ;  12920548 ;  12920555 ;  12920558 ;  12920559 ;  12920560 ;  12920562 ;  12920565 ;  12920566 ;  12920567 ;  12920568 ;  12920569 ;  12920570 ;  12920571 ;  12920574 ;  12920576 ;  12920577 ;  12920578 ;  12920581 ;  12920582 ;  12920585 ;  12920586 ;  12920587 ;  12920589 ;  12920590 ;  12920591 ;  12920592 ;  12920593 ;  12920596 ;  12920598 ;  12920599 ;  12920600 ;  12920601 ;  12920602 ;  12920603 ;  12920604 ;  12920605 ;  12920606 ;  12920607 ;  12920608 ;  12920610 ;  12920611 ;  12920612 ;  12920614 ;  12920615 ;  12920620 ;  12920621 ;  12920623 ;  12920626 ;  12920627 ;  12920629 ;  12920632 ;  12920633 ;  12920634 ;  12920635 ;  12920639 ;  12920641 ;  12920644 ;  12920646 ;  12920647 ;  12920648 ;  12920649 ;  12920651 ;  12920654 ;  12920657 ;  12920659 ;  12920660 ;  12920663 ;  12920664 ;  12920665 ;  12920666 ;  12920668 ;  12920669 ;  12920670 ;  12920671 ;  12920674 ;  12920675 ;  12920677 ;  12920679 ;  12920680 ;  12920681 ;  12920682 ;  12920686 ;  12920688 ;  12920690 ;  12920691 ;  12920692 ;  12920696 ;  12920699 ;  12920702 ;  12920703 ;  12920706 ;  12920708 ;  12920710 ;  12920714 ;  12920715 ;  12920716 ;  12920717 ;  12920720 ;  12920724 ;  12920726 ;  12920727 ;  12920728 ;  12920729 ;  12920731 ;  12920733 ;  12920734 ;  12920736 ;  12920743 ;  12920744 ;  12920746 ;  12920747 ;  12920749 ;  12920750 ;  12920756 ;  12920759 ;  12920761 ;  12920762 ;  12920763 ;  12920766 ;  12920767 ;  12920770 ;  12920771 ;  12920773 ;  12920774 ;  12920778 ;  12920779 ;  12920782 ;  12920783 ;  12920788 ;  12920792 ;  12920793 ;  12920795 ;  12920796 ;  12920797 ;  12920798 ;  12920800 ;  12920802 ;  12920808 ;  12920809 ;  12920811 ;  12920813 ;  12920815 ;  12920819 ;  12920820 ;  12920821 ;  12920826 ;  12920827 ;  12920828 ;  12920834 ;  12920836 ;  12920837 ;  12920839 ;  12920840 ;  12920841 ;  12920844 ;  12920848 ;  12920850 ;  12920851 ;  12920852 ;  12920853 ;  12920856 ;  12920857 ;  12920858 ;  12920859 ;  12920861 ;  12920863 ;  12920865 ;  12920866 ;  12920868 ;  12920871 ;  12920872 ;  12920873 ;  12920874 ;  12920875 ;  12920879 ;  12920880 ;  12920881 ;  12920884 ;  12920885 ;  12920887 ;  12920888 ;  12920890 ;  12920891 ;  12920894 ;  12920895 ;  12920896 ;  12920900 ;  12920902 ;  12920903 ;  12920904 ;  12920908 ;  12920909 ;  12920910 ;  12920911 ;  12920916 ;  12920917 ;  12920918 ;  12920924 ;  12920925 ;  12920926 ;  12920928 ;  12920931 ;  12920932 ;  12920933 ;  12920935 ;  12920936 ;  12920938 ;  12920940 ;  12920941 ;  12920945 ;  12920946 ;  12920947 ;  12920948 ;  12920950 ;  12920951 ;  12920954 ;  12920956 ;  12920957 ;  12920958 ;  12920960 ;  12920961 ;  12920963 ;  12920966 ;  12920968 ;  12920969 ;  12920970 ;  12920971 ;  12920973 ;  12920974 ;  12920978 ;  12920979 ;  12920980 ;  12920982 ;  12920983 ;  12920984 ;  12920986 ;  12920990 ;  12920991 ;  12920997 ;  12921003 ;  12921005 ;  12921006 ;  12921008 ;  12921010 ;  12921011 ;  12921012 ;  12921017 ;  12921018 ;  12921020 ;  12921021 ;  12921023 ;  12921024 ;  12921025 ;  12921026 ;  12921030 ;  12921032 ;  12921040 ;  12921042 ;  12921044 ;  12921050 ;  12921053 ;  12921054 ;  12921055 ;  12921056 ;  12921057 ;  12921060 ;  12921063 ;  12921064 ;  12921065 ;  12921066 ;  12921067 ;  12921069 ;  12921070 ;  12921072 ;  12921074 ;  12921076 ;  12921082 ;  12921083 ;  12921084 ;  12921086 ;  12921090 ;  12921091 ;  12921093 ;  12921095 ;  12921096 ;  12921097 ;  12921098 ;  12921099 ;  12921100 ;  12921102 ;  12921106 ;  12921107 ;  12921109 ;  12921110 ;  12921111 ;  12921115 ;  12921116 ;  12921117 ;  12921118 ;  12921119 ;  12921121 ;  12921123 ;  12921124 ;  12921125 ;  12921126 ;  12921128 ;  12921130 ;  12921135 ;  12921137 ;  12921142 ;  12921144 ;  12921145 ;  12921146 ;  12921147 ;  12921151 ;  12921155 ;  12921158 ;  12921159 ;  12921160 ;  12921161 ;  12921163 ;  12921168 ;  12921172 ;  12921173 ;  12921174 ;  12921175 ;  12921177 ;  12921179 ;  12921180 ;  12921181 ;  12921183 ;  12921184 ;  12921185 ;  12921186 ;  12921187 ;  12921188 ;  12921189 ;  12921192 ;  12921193 ;  12921194 ;  12921195 ;  12921197 ;  12921199 ;  12921202 ;  12921204 ;  12921205 ;  12921207 ;  12921212 ;  12921217 ;  12921219 ;  12921221 ;  12921223 ;  12921226 ;  12921227 ;  12921230 ;  12921232 ;  12921234 ;  12921235 ;  12921236 ;  12921238 ;  12921239 ;  12921241 ;  12921242 ;  12921243 ;  12921247 ;  12921249 ;  12921250 ;  12921254 ;  12921255 ;  12921256 ;  12921259 ;  12921260 ;  12921262 ;  12921263 ;  12921264 ;  12921265 ;  12921266 ;  12921267 ;  12921269 ;  12921270 ;  12921271 ;  12921272 ;  12921273 ;  12921274 ;  12921276 ;  12921280 ;  12921281 ;  12921282 ;  12921288 ;  12921290 ;  12921294 ;  12921296 ;  12921297 ;  12921298 ;  12921299 ;  12921300 ;  12921302 ;  12921304 ;  12921305 ;  12921306 ;  12921307 ;  12921309 ;  12921311 ;  12921312 ;  12921314 ;  12921315 ;  12921316 ;  12921317 ;  12921319 ;  12921324 ;  12921325 ;  12921326 ;  12921328 ;  12921330 ;  12921331 ;  12921332 ;  12921335 ;  12921338 ;  12921340 ;  12921342 ;  12921343 ;  12921344 ;  12921347 ;  12921348 ;  12921349 ;  12921351 ;  12921352 ;  12921353 ;  12921354 ;  12921356 ;  12921359 ;  12921360 ;  12921361 ;  12921363 ;  12921366 ;  12921368 ;  12921370 ;  12921372 ;  12921373 ;  12921374 ;  12921376 ;  12921377 ;  12921378 ;  12921379 ;  12921380 ;  12921381 ;  12921385 ;  12921386 ;  12921387 ;  12921389 ;  12921390 ;  12921391 ;  12921392 ;  12921394 ;  12921395 ;  12921396 ;  12921399 ;  12921400 ;  12921403 ;  12921404 ;  12921405 ;  12921406 ;  12921408 ;  12921410 ;  12921413 ;  12921414 ;  12921415 ;  12921417 ;  12921418 ;  12921419 ;  12921420 ;  12921421 ;  12921422 ;  12921423 ;  12921425 ;  12921426 ;  12921427 ;  12921428 ;  12921432 ;  12921439 ;  12921441 ;  12921442 ;  12921444 ;  12921445 ;  12921446 ;  12921447 ;  12921448 ;  12921449 ;  12921450 ;  12921451 ;  12921459 ;  12921462 ;  12921463 ;  12921469 ;  12921475 ;  12921477 ;  12921478 ;  12921479 ;  12921480 ;  12921482 ;  12921483 ;  12921485 ;  12921486 ;  12921487 ;  12921488 ;  12921489 ;  12921490 ;  12921491 ;  12921493 ;  12921495 ;  12921497 ;  12921498 ;  12921499 ;  12921500 ;  12921501 ;  12921502 ;  12921504 ;  12921507 ;  12921508 ;  12921509 ;  12921510 ;  12921512 ;  12921514 ;  12921517 ;  12921519 ;  12921521 ;  12921525 ;  12921527 ;  12921530 ;  12921531 ;  12921532 ;  12921533 ;  12921534 ;  12921535 ;  12921538 ;  12921539 ;  12921540 ;  12921541 ;  12921542 ;  12921544 ;  12921545 ;  12921546 ;  12921547 ;  12921548 ;  12921550 ;  12921551 ;  12921552 ;  12921554 ;  12921555 ;  12921556 ;  12921558 ;  12921560 ;  12921561 ;  12921562 ;  12921563 ;  12921564 ;  12921565 ;  12921569 ;  12921570 ;  12921576 ;  12921577 ;  12921578 ;  12921580 ;  12921583 ;  12921584 ;  12921585 ;  12921587 ;  12921588 ;  12921591 ;  12921592 ;  12921597 ;  12921603 ;  12921606 ;  12921607 ;  12921613 ;  12921615 ;  12921616 ;  12921618 ;  12921621 ;  12921626 ;  12921636 ;  12921637 ;  12921638 ;  12921639 ;  12921640 ;  12921641 ;  12921643 ;  12921644 ;  12921646 ;  12921647 ;  12921648 ;  12921650 ;  12921652 ;  12921654 ;  12921655 ;  12921656 ;  12921657 ;  12921658 ;  12921661 ;  12921663 ;  12921665 ;  12921669 ;  12921671 ;  12921672 ;  12921676 ;  12921680 ;  12921686 ;  12921689 ;  12921690 ;  12921693 ;  12921694 ;  12921695 ;  12921700 ;  12921701 ;  12921702 ;  12921703 ;  12921704 ;  12921705 ;  12921706 ;  12921710 ;  12921712 ;  12921716 ;  12921717 ;  12921719 ;  12921720 ;  12921721 ;  12921723 ;  12921724 ;  12921728 ;  12921730 ;  12921733 ;  12921734 ;  12921738 ;  12921748 ;  12921750 ;  12921756 ;  12921757 ;  12921761 ;  12921769 ;  12921770 ;  12921774 ;  12921776 ;  12921777 ;  12921778 ;  12921784 ;  12921785 ;  12921786 ;  12921787 ;  12921788 ;  12921789 ;  12921791 ;  12921794 ;  12921797 ;  12921800 ;  12921803 ;  12921805 ;  12921806 ;  12921807 ;  12921812 ;  12921814 ;  12921817 ;  12921818 ;  12921819 ;  12921824 ;  12921825 ;  12921827 ;  12921828 ;  12921829 ;  12921830 ;  12921831 ;  12921833 ;  12921834 ;  12921839 ;  12921840 ;  12921841 ;  12921846 ;  12921850 ;  12921851 ;  12921854 ;  12921855 ;  12921857 ;  12921858 ;  12921860 ;  12921861 ;  12921862 ;  12921863 ;  12921864 ;  12921866 ;  12921867 ;  12921868 ;  12921871 ;  12921872 ;  12921876 ;  12921877 ;  12921878 ;  12921881 ;  12921882 ;  12921883 ;  12921886 ;  12921889 ;  12921892 ;  12921895 ;  12921896 ;  12921897 ;  12921898 ;  12921902 ;  12921908 ;  12921912 ;  12921913 ;  12921919 ;  12921924 ;  12921925 ;  12921926 ;  12921929 ;  12921931 ;  12921932 ;  12921938 ;  12921941 ;  12921942 ;  12921944 ;  12921947 ;  12921948 ;  12921950 ;  12921951 ;  12921952 ;  12921953 ;  12921956 ;  12921957 ;  12921958 ;  12921963 ;  12921964 ;  12921965 ;  12921967 ;  12921968 ;  12921972 ;  12921975 ;  12921976 ;  12921977 ;  12921978 ;  12921983 ;  12921984 ;  12921985 ;  12921986 ;  12921987 ;  12921990 ;  12921995 ;  12921998 ;  12922001 ;  12922020 ;  12922024 ;  12922030 ;  12922034 ;  12922036 ;  12922041 ;  12922043 ;  12922044 ;  12922045 ;  12922047 ;  12922051 ;  12922052 ;  12922054 ;  12922055 ;  12922056 ;  12922058 ;  12922059 ;  12922060 ;  12922061 ;  12922062 ;  12922065 ;  12922066 ;  12922067 ;  12922069 ;  12922071 ;  12922072 ;  12922074 ;  12922075 ;  12922076 ;  12922077 ;  12922078 ;  12922079 ;  12922081 ;  12922083 ;  12922084 ;  12922085 ;  12922087 ;  12922088 ;  12922092 ;  12922093 ;  12922094 ;  12922096 ;  12922097 ;  12922098 ;  12922099 ;  12922100 ;  12922101 ;  12922103 ;  12922105 ;  12922106 ;  12922107 ;  12922108 ;  12922110 ;  12922111 ;  12922113 ;  12922114 ;  12922116 ;  12922117 ;  12922119 ;  12922120 ;  12922121 ;  12922122 ;  12922123 ;  12922126 ;  12922127 ;  12922128 ;  12922129 ;  12922131 ;  12922132 ;  12922133 ;  12922135 ;  12922139 ;  12922141 ;  12922142 ;  12922143 ;  12922144 ;  12922145 ;  12922148 ;  12922150 ;  12922153 ;  12922155 ;  12922156 ;  12922157 ;  12922158 ;  12922161 ;  12922162 ;  12922163 ;  12922164 ;  12922165 ;  12922167 ;  12922168 ;  12922170 ;  12922171 ;  12922172 ;  12922174 ;  12922175 ;  12922176 ;  12922177 ;  12922178 ;  12922179 ;  12922180 ;  12922181 ;  12922182 ;  12922183 ;  12922184 ;  12922185 ;  12922186 ;  12922187 ;  12922188 ;  12922189 ;  12922190 ;  12922191 ;  12922192 ;  12922193 ;  12922194 ;  12922195 ;  12922196 ;  12922198 ;  12922199 ;  12922200 ;  12922201 ;  12922202 ;  12922203 ;  12922204 ;  12922205 ;  12922206 ;  12922207 ;  12922208 ;  12922209 ;  12922210 ;  12922211 ;  12922212 ;  12922213 ;  12922215 ;  12922216 ;  12922219 ;  12922220 ;  12922221 ;  12922222 ;  12922224 ;  12922225 ;  12922226 ;  12922227 ;  12922229 ;  12922230 ;  12922232 ;  12922233 ;  12922235 ;  12922236 ;  12922240 ;  12922241 ;  12922242 ;  12922244 ;  12922250 ;  12922251 ;  12922252 ;  12922253 ;  12922254 ;  12922256 ;  12922257 ;  12922258 ;  12922260 ;  12922261 ;  12922262 ;  12922263 ;  12922264 ;  12922266 ;  12922267 ;  12922269 ;  12922270 ;  12922272 ;  12922273 ;  12922274 ;  12922277 ;  12922278 ;  12922279 ;  12922280 ;  12922281 ;  12922283 ;  12922284 ;  12922285 ;  12922286 ;  12922288 ;  12922289 ;  12922290 ;  12922292 ;  12922293 ;  12922294 ;  12922295 ;  12922297 ;  12922298 ;  12922299 ;  12922300 ;  12922301 ;  12922302 ;  12922303 ;  12922304 ;  12922306 ;  12922307 ;  12922309 ;  12922310 ;  12922312 ;  12922313 ;  12922314 ;  12922315 ;  12922316 ;  12922317 ;  12922318 ;  12922319 ;  12922320 ;  12922321 ;  12922322 ;  12922324 ;  12922325 ;  12922326 ;  12922327 ;  12922331 ;  12922333 ;  12922334 ;  12922336 ;  12922338 ;  12922340 ;  12922341 ;  12922342 ;  12922343 ;  12922344 ;  12922345 ;  12922346 ;  12922348 ;  12922350 ;  12922351 ;  12922352 ;  12922353 ;  12922356 ;  12922357 ;  12922358 ;  12922359 ;  12922360 ;  12922361 ;  12922363 ;  12922366 ;  12922367 ;  12922368 ;  12922369 ;  12922370 ;  12922371 ;  12922373 ;  12922376 ;  12922378 ;  12922380 ;  12922382 ;  12922383 ;  12922384 ;  12922385 ;  12922387 ;  12922388 ;  12922389 ;  12922393 ;  12922395 ;  12922398 ;  12922404 ;  12922405 ;  12922406 ;  12922412 ;  12922420 ;  12922421 ;  12922422 ;  12922424 ;  12922426 ;  12922427 ;  12922429 ;  12922430 ;  12922433 ;  12922434 ;  12922436 ;  12922437 ;  12922439 ;  12922440 ;  12922442 ;  12922444 ;  12922450 ;  12922451 ;  12922452 ;  12922456 ;  12922461 ;  12922463 ;  12922464 ;  12922465 ;  12922466 ;  12922468 ;  12922470 ;  12922471 ;  12922472 ;  12922474 ;  12922475 ;  12922477 ;  12922478 ;  12922481 ;  12922482 ;  12922483 ;  12922485 ;  12922486 ;  12922487 ;  12922489 ;  12922494 ;  12922495 ;  12922496 ;  12922497 ;  12922498 ;  12922501 ;  12922507 ;  12922509 ;  12922510 ;  12922514 ;  12922515 ;  12922520 ;  12922523 ;  12922526 ;  12922527 ;  12922528 ;  12922532 ;  12922534 ;  12922536 ;  12922537 ;  12922539 ;  12922540 ;  12922542 ;  12922543 ;  12922546 ;  12922547 ;  12922548 ;  12922549 ;  12922551 ;  12922552 ;  12922554 ;  12922555 ;  12922556 ;  12922558 ;  12922559 ;  12922560 ;  12922561 ;  12922562 ;  12922563 ;  12922564 ;  12922567 ;  12922569 ;  12922571 ;  12922572 ;  12922573 ;  12922575 ;  12922576 ;  12922580 ;  12922581 ;  12922582 ;  12922584 ;  12922585 ;  12922589 ;  12922592 ;  12922593 ;  12922596 ;  12922597 ;  12922602 ;  12922607 ;  12922608 ;  12922609 ;  12922610 ;  12922612 ;  12922614 ;  12922616 ;  12922617 ;  12922619 ;  12922620 ;  12922622 ;  12922623 ;  12922625 ;  12922631 ;  12922635 ;  12922636 ;  12922639 ;  12922640 ;  12922641 ;  12922645 ;  12922646 ;  12922647 ;  12922648 ;  12922649 ;  12922650 ;  12922654 ;  12922658 ;  12922659 ;  12922660 ;  12922663 ;  12922664 ;  12922665 ;  12922668 ;  12922669 ;  12922675 ;  12922677 ;  12922678 ;  12922679 ;  12922680 ;  12922683 ;  12922686 ;  12922689 ;  12922690 ;  12922691 ;  12922693 ;  12922695 ;  12922696 ;  12922697 ;  12922698 ;  12922699 ;  12922700 ;  12922701 ;  12922702 ;  12922703 ;  12922704 ;  12922705 ;  12922708 ;  12922710 ;  12922711 ;  12922712 ;  12922713 ;  12922714 ;  12922715 ;  12922715 ;  12922715 ;  12922716 ;  12922717 ;  12922718 ;  12922719 ;  12922720 ;  12922721 ;  12922722 ;  12922724 ;  12922725 ;  12922726 ;  12922727 ;  12922731 ;  12922732 ;  12922736 ;  12922737 ;  12922738 ;  12922741 ;  12922742 ;  12922743 ;  12922744 ;  12922745 ;  12922749 ;  12922750 ;  12922751 ;  12922754 ;  12922755 ;  12922756 ;  12922757 ;  12922758 ;  12922759 ;  12922761 ;  12922763 ;  12922765 ;  12922766 ;  12922767 ;  12922771 ;  12922773 ;  12922776 ;  12922780 ;  12922782 ;  12922783 ;  12922786 ;  12922788 ;  12922791 ;  12922795 ;  12922796 ;  12922798 ;  12922799 ;  12922800 ;  12922801 ;  12922802 ;  12922803 ;  12922805 ;  12922806 ;  12922808 ;  12922809 ;  12922810 ;  12922811 ;  12922813 ;  12922815 ;  12922816 ;  12922817 ;  12922818 ;  12922820 ;  12922821 ;  12922823 ;  12922824 ;  12922825 ;  12922826 ;  12922827 ;  12922828 ;  12922829 ;  12922830 ;  12922832 ;  12922833 ;  12922834 ;  12922835 ;  12922837 ;  12922838 ;  12922841 ;  12922842 ;  12922843 ;  12922846 ;  12922847 ;  12922848 ;  12922850 ;  12922852 ;  12922854 ;  12922861 ;  12922863 ;  12922864 ;  12922867 ;  12922872 ;  12922873 ;  12922884 ;  12922885 ;  12922887 ;  12922889 ;  12922891 ;  12922892 ;  12922893 ;  12922896 ;  12922897 ;  12922910 ;  12922915 ;  12922918 ;  12922919 ;  12922921 ;  12922922 ;  12922926 ;  12922927 ;  12922930 ;  12922931 ;  12922933 ;  12922937 ;  12922939 ;  12922945 ;  12922947 ;  12922949 ;  12922950 ;  12922951 ;  12922952 ;  12922953 ;  12922955 ;  12922956 ;  12922957 ;  12922958 ;  12922960 ;  12922962 ;  12922964 ;  12922965 ;  12922966 ;  12922967 ;  12922968 ;  12922971 ;  12922972 ;  12922973 ;  12922975 ;  12922976 ;  12922977 ;  12922978 ;  12922979 ;  12922980 ;  12922981 ;  12922982 ;  12922984 ;  12922985 ;  12922986 ;  12922987 ;  12922988 ;  12922989 ;  12922990 ;  12922991 ;  12922993 ;  12922994 ;  12922995 ;  12922998 ;  12923001 ;  12923002 ;  12923005 ;  12923006 ;  12923007 ;  12923008 ;  12923012 ;  12923015 ;  12923017 ;  12923020 ;  12923021 ;  12923023 ;  12923027 ;  12923030 ;  12923031 ;  12923032 ;  12923033 ;  12923035 ;  12923036 ;  12923040 ;  12923042 ;  12923049 ;  12923051 ;  12923055 ;  12923057 ;  12923058 ;  12923059 ;  12923062 ;  12923063 ;  12923064 ;  12923065 ;  12923067 ;  12923068 ;  12923069 ;  12923072 ;  12923073 ;  12923074 ;  12923075 ;  12923076 ;  12923077 ;  12923079 ;  12923080 ;  12923081 ;  12923082 ;  12923083 ;  12923084 ;  12923085 ;  12923087 ;  12923090 ;  12923091 ;  12923092 ;  12923095 ;  12923096 ;  12923097 ;  12923098 ;  12923101 ;  12923102 ;  12923103 ;  12923105 ;  12923106 ;  12923108 ;  12923109 ;  12923110 ;  12923111 ;  12923114 ;  12923115 ;  12923116 ;  12923117 ;  12923118 ;  12923120 ;  12923121 ;  12923122 ;  12923124 ;  12923125 ;  12923126 ;  12923127 ;  12923128 ;  12923129 ;  12923132 ;  12923133 ;  12923134 ;  12923136 ;  12923137 ;  12923138 ;  12923139 ;  12923140 ;  12923142 ;  12923143 ;  12923144 ;  12923146 ;  12923147 ;  12923148 ;  12923149 ;  12923150 ;  12923152 ;  12923153 ;  12923155 ;  12923156 ;  12923157 ;  12923158 ;  12923159 ;  12923160 ;  12923161 ;  12923162 ;  12923163 ;  12923164 ;  12923167 ;  12923168 ;  12923169 ;  12923171 ;  12923172 ;  12923173 ;  12923174 ;  12923177 ;  12923178 ;  12923179 ;  12923181 ;  12923183 ;  12923184 ;  12923186 ;  12923187 ;  12923189 ;  12923190 ;  12923191 ;  12923192 ;  12923193 ;  12923194 ;  12923195 ;  12923197 ;  12923199 ;  12923201 ;  12923202 ;  12923204 ;  12923205 ;  12923206 ;  12923207 ;  12923208 ;  12923211 ;  12923212 ;  12923216 ;  12923219 ;  12923221 ;  12923224 ;  12923228 ;  12923230 ;  12923235 ;  12923237 ;  12923241 ;  12923244 ;  12923247 ;  12923251 ;  12923256 ;  12923258 ;  12923259 ;  12923260 ;  12923262 ;  12923264 ;  12923265 ;  12923269 ;  12923271 ;  12923272 ;  12923274 ;  12923275 ;  12923277 ;  12923278 ;  12923279 ;  12923282 ;  12923284 ;  12923285 ;  12923286 ;  12923287 ;  12923288 ;  12923289 ;  12923292 ;  12923293 ;  12923295 ;  12923297 ;  12923299 ;  12923300 ;  12923301 ;  12923302 ;  12923303 ;  12923304 ;  12923305 ;  12923308 ;  12923311 ;  12923312 ;  12923314 ;  12923315 ;  12923316 ;  12923322 ;  12923323 ;  12923332 ;  12923333 ;  12923336 ;  12923340 ;  12923342 ;  12923344 ;  12923350 ;  12923353 ;  12923355 ;  12923356 ;  12923357 ;  12923360 ;  12923373 ;  12923378 ;  12923379 ;  12923382 ;  12923383 ;  12923385 ;  12923388 ;  12923389 ;  12923396 ;  12923402 ;  12923405 ;  12923406 ;  12923408 ;  12923414 ;  12923415 ;  12923416 ;  12923417 ;  12923418 ;  12923420 ;  12923421 ;  12923423 ;  12923425 ;  12923426 ;  12923427 ;  12923428 ;  12923431 ;  12923433 ;  12923435 ;  12923437 ;  12923438 ;  12923440 ;  12923441 ;  12923446 ;  12923447 ;  12923448 ;  12923451 ;  12923452 ;  12923453 ;  12923454 ;  12923458 ;  12923460 ;  12923461 ;  12923465 ;  12923466 ;  12923467 ;  12923468 ;  12923470 ;  12923472 ;  12923474 ;  12923475 ;  12923477 ;  12923478 ;  12923480 ;  12923481 ;  12923483 ;  12923484 ;  12923496 ;  12923502 ;  12923506 ;  12923507 ;  12923508 ;  12923511 ;  12923518 ;  12923533 ;  12923548 ;  12923552 ;  12923565 ;  12923571 ;  12923576 ;  12923579 ;  12923584 ;  12923589 ;  12923593 ;  12923594 ;  12923599 ;  12923606 ;  12923609 ;  12923613 ;  12923614 ;  12923615 ;  12923617 ;  12923618 ;  12923621 ;  12923622 ;  12923623 ;  12923625 ;  12923627 ;  12923628 ;  12923630 ;  12923632 ;  12923635 ;  12923637 ;  12923640 ;  12923645 ;  12923648 ;  12923653 ;  12923657 ;  12923658 ;  12923659 ;  12923660 ;  12923661 ;  12923663 ;  12923664 ;  12923665 ;  12923666 ;  12923667 ;  12923669 ;  12923671 ;  12923673 ;  12923676 ;  12923677 ;  12923680 ;  12923681 ;  12923684 ;  12923685 ;  12923687 ;  12923689 ;  12923690 ;  12923691 ;  12923692 ;  12923695 ;  12923697 ;  12923699 ;  12923700 ;  12923703 ;  12923704 ;  12923705 ;  12923710 ;  12923711 ;  12923712 ;  12923714 ;  12923715 ;  12923716 ;  12923718 ;  12923719 ;  12923720 ;  12923721 ;  12923722 ;  12923725 ;  12923726 ;  12923728 ;  12923730 ;  12923731 ;  12923732 ;  12923734 ;  12923735 ;  12923736 ;  12923737 ;  12923738 ;  12923741 ;  12923742 ;  12923746 ;  12923747 ;  12923748 ;  12923749 ;  12923750 ;  12923751 ;  12923753 ;  12923754 ;  12923757 ;  12923760 ;  12923774 ;  12923777 ;  12923779 ;  12923784 ;  12923786 ;  12923792 ;  12923794 ;  12923795 ;  12923796 ;  12923799 ;  12923800 ;  12923802 ;  12923803 ;  12923805 ;  12923811 ;  12923815 ;  12923816 ;  12923818 ;  12923820 ;  12923822 ;  12923823 ;  12923824 ;  12923825 ;  12923826 ;  12923829 ;  12923830 ;  12923834 ;  12923836 ;  12923838 ;  12923839 ;  12923840 ;  12923841 ;  12923842 ;  12923843 ;  12923845 ;  12923847 ;  12923848 ;  12923850 ;  12923851 ;  12923853 ;  12923854 ;  12923855 ;  12923856 ;  12923857 ;  12923858 ;  12923859 ;  12923860 ;  12923861 ;  12923862 ;  12923863 ;  12923864 ;  12923865 ;  12923866 ;  12923869 ;  12923870 ;  12923871 ;  12923872 ;  12923873 ;  12923874 ;  12923876 ;  12923877 ;  12923878 ;  12923879 ;  12923880 ;  12923881 ;  12923882 ;  12923883 ;  12923884 ;  12923885 ;  12923887 ;  12923888 ;  12923889 ;  12923890 ;  12923891 ;  12923892 ;  12923893 ;  12923894 ;  12923895 ;  12923896 ;  12923897 ;  12923899 ;  12923900 ;  12923901 ;  12923902 ;  12923904 ;  12923905 ;  12923906 ;  12923907 ;  12923908 ;  12923909 ;  12923910 ;  12923911 ;  12923912 ;  12923913 ;  12923914 ;  12923915 ;  12923916 ;  12923917 ;  12923918 ;  12923919 ;  12923920 ;  12923921 ;  12923922 ;  12923924 ;  12923925 ;  12923926 ;  12923927 ;  12923928 ;  12923929 ;  12923930 ;  12923931 ;  12923932 ;  12923933 ;  12923934 ;  12923935 ;  12923936 ;  12923937 ;  12923938 ;  12923939 ;  12923940 ;  12923941 ;  12923942 ;  12923943 ;  12923944 ;  12923945 ;  12923946 ;  12923949 ;  12923950 ;  12923951 ;  12923952 ;  12923954 ;  12923955 ;  12923956 ;  12923957 ;  12923958 ;  12923959 ;  12923960 ;  12923962 ;  12923963 ;  12923964 ;  12923966 ;  12923967 ;  12923968 ;  12923969 ;  12923970 ;  12923971 ;  12923972 ;  12923973 ;  12923974 ;  12923975 ;  12923977 ;  12923978 ;  12923979 ;  12923980 ;  12923981 ;  12923982 ;  12923983 ;  12923984 ;  12923985 ;  12923986 ;  12923987 ;  12923988 ;  12923989 ;  12923990 ;  12923992 ;  12923993 ;  12923994 ;  12923995 ;  12923996 ;  12923997 ;  12923998 ;  12923999 ;  12924000 ;  12924001 ;  12924002 ;  12924003 ;  12924004 ;  12924005 ;  12924006 ;  12924007 ;  12924008 ;  12924010 ;  12924011 ;  12924012 ;  12924013 ;  12924014 ;  12924015 ;  12924017 ;  12924019 ;  12924020 ;  12924021 ;  12924022 ;  12924023 ;  12924024 ;  12924025 ;  12924026 ;  12924027 ;  12924028 ;  12924029 ;  12924032 ;  12924033 ;  12924034 ;  12924035 ;  12924036 ;  12924037 ;  12924038 ;  12924039 ;  12924040 ;  12924041 ;  12924042 ;  12924043 ;  12924045 ;  12924048 ;  12924052 ;  12924053 ;  12924058 ;  12924059 ;  12924060 ;  12924062 ;  12924063 ;  12924065 ;  12924066 ;  12924067 ;  12924068 ;  12924069 ;  12924070 ;  12924071 ;  12924072 ;  12924073 ;  12924074 ;  12924075 ;  12924076 ;  12924077 ;  12924078 ;  12924080 ;  12924081 ;  12924082 ;  12924083 ;  12924085 ;  12924086 ;  12924088 ;  12924089 ;  12924091 ;  12924092 ;  12924093 ;  12924094 ;  12924095 ;  12924096 ;  12924097 ;  12924098 ;  12924099 ;  12924100 ;  12924101 ;  12924102 ;  12924103 ;  12924104 ;  12924105 ;  12924106 ;  12924107 ;  12924108 ;  12924111 ;  12924112 ;  12924113 ;  12924114 ;  12924115 ;  12924116 ;  12924117 ;  12924118 ;  12924119 ;  12924121 ;  12924122 ;  12924123 ;  12924125 ;  12924127 ;  12924128 ;  12924129 ;  12924130 ;  12924131 ;  12924132 ;  12924133 ;  12924134 ;  12924135 ;  12924136 ;  12924137 ;  12924138 ;  12924139 ;  12924140 ;  12924141 ;  12924142 ;  12924145 ;  12924146 ;  12924147 ;  12924148 ;  12924149 ;  12924150 ;  12924151 ;  12924152 ;  12924153 ;  12924154 ;  12924155 ;  12924156 ;  12924157 ;  12924158 ;  12924159 ;  12924162 ;  12924163 ;  12924165 ;  12924166 ;  12924167 ;  12924168 ;  12924171 ;  12924174 ;  12924175 ;  12924176 ;  12924177 ;  12924178 ;  12924179 ;  12924180 ;  12924181 ;  12924182 ;  12924183 ;  12924185 ;  12924186 ;  12924187 ;  12924189 ;  12924192 ;  12924193 ;  12924194 ;  12924195 ;  12924196 ;  12924197 ;  12924198 ;  12924199 ;  12924200 ;  12924201 ;  12924202 ;  12924203 ;  12924204 ;  12924205 ;  12924206 ;  12924207 ;  12924209 ;  12924210 ;  12924211 ;  12924212 ;  12924213 ;  12924214 ;  12924216 ;  12924217 ;  12924218 ;  12924219 ;  12924221 ;  12924222 ;  12924223 ;  12924224 ;  12924225 ;  12924226 ;  12924227 ;  12924228 ;  12924229 ;  12924230 ;  12924231 ;  12924232 ;  12924233 ;  12924235 ;  12924237 ;  12924238 ;  12924239 ;  12924241 ;  12924242 ;  12924243 ;  12924244 ;  12924245 ;  12924247 ;  12924248 ;  12924249 ;  12924250 ;  12924251 ;  12924252 ;  12924253 ;  12924255 ;  12924257 ;  12924258 ;  12924259 ;  12924260 ;  12924261 ;  12924262 ;  12924263 ;  12924264 ;  12924265 ;  12924266 ;  12924267 ;  12924269 ;  12924270 ;  12924271 ;  12924272 ;  12924273 ;  12924274 ;  12924275 ;  12924277 ;  12924278 ;  12924279 ;  12924281 ;  12924282 ;  12924283 ;  12924284 ;  12924286 ;  12924287 ;  12924288 ;  12924290 ;  12924291 ;  12924293 ;  12924295 ;  12924296 ;  12924298 ;  12924299 ;  12924301 ;  12924302 ;  12924303 ;  12924307 ;  12924308 ;  12924309 ;  12924311 ;  12924313 ;  12924314 ;  12924315 ;  12924317 ;  12924321 ;  12924322 ;  12924324 ;  12924326 ;  12924328 ;  12924329 ;  12924332 ;  12924335 ;  12924337 ;  12924344 ;  12924347 ;  12924348 ;  12924349 ;  12924352 ;  12924353 ;  12924355 ;  12924357 ;  12924358 ;  12924359 ;  12924363 ;  12924364 ;  12924367 ;  12924370 ;  12924371 ;  12924372 ;  12924373 ;  12924377 ;  12924378 ;  12924379 ;  12924380 ;  12924384 ;  12924385 ;  12924386 ;  12924388 ;  12924389 ;  12924391 ;  12924392 ;  12924397 ;  12924398 ;  12924399 ;  12924403 ;  12924404 ;  12924407 ;  12924408 ;  12924409 ;  12924411 ;  12924417 ;  12924421 ;  12924428 ;  12924429 ;  12924430 ;  12924431 ;  12924432 ;  12924433 ;  12924435 ;  12924437 ;  12924443 ;  12924446 ;  12924449 ;  12924452 ;  12924459 ;  12924464 ;  12924467 ;  12924469 ;  12924473 ;  12924480 ;  12924481 ;  12924483 ;  12924489 ;  12924490 ;  12924498 ;  12924501 ;  12924504 ;  12924513 ;  12924515 ;  12924518 ;  12924520 ;  12924521 ;  12924525 ;  12924529 ;  12924530 ;  12924531 ;  12924536 ;  12924537 ;  12924538 ;  12924541 ;  12924542 ;  12924543 ;  12924546 ;  12924547 ;  12924550 ;  12924554 ;  12924556 ;  12924560 ;  12924562 ;  12924564 ;  12924565 ;  12924566 ;  12924567 ;  12924572 ;  12924575 ;  12924576 ;  12924578 ;  12924579 ;  12924580 ;  12924581 ;  12924584 ;  12924585 ;  12924586 ;  12924587 ;  12924588 ;  12924589 ;  12924590 ;  12924592 ;  12924593 ;  12924594 ;  12924595 ;  12924596 ;  12924599 ;  12924600 ;  12924603 ;  12924605 ;  12924606 ;  12924609 ;  12924610 ;  12924611 ;  12924613 ;  12924614 ;  12924615 ;  12924617 ;  12924618 ;  12924619 ;  12924620 ;  12924621 ;  12924625 ;  12924627 ;  12924628 ;  12924629 ;  12924630 ;  12924632 ;  12924635 ;  12924636 ;  12924637 ;  12924638 ;  12924640 ;  12924642 ;  12924643 ;  12924644 ;  12924645 ;  12924646 ;  12924648 ;  12924649 ;  12924650 ;  12924651 ;  12924652 ;  12924653 ;  12924654 ;  12924656 ;  12924657 ;  12924658 ;  12924659 ;  12924664 ;  12924667 ;  12924668 ;  12924670 ;  12924671 ;  12924673 ;  12924674 ;  12924675 ;  12924676 ;  12924677 ;  12924678 ;  12924679 ;  12924680 ;  12924681 ;  12924682 ;  12924684 ;  12924685 ;  12924686 ;  12924688 ;  12924689 ;  12924691 ;  12924693 ;  12924694 ;  12924695 ;  12924697 ;  12924698 ;  12924699 ;  12924700 ;  12924703 ;  12924705 ;  12924709 ;  12924711 ;  12924713 ;  12924715 ;  12924716 ;  12924719 ;  12924723 ;  12924724 ;  12924727 ;  12924728 ;  12924729 ;  12924733 ;  12924734 ;  12924735 ;  12924736 ;  12924740 ;  12924742 ;  12924743 ;  12924747 ;  12924748 ;  12924749 ;  12924751 ;  12924752 ;  12924754 ;  12924756 ;  12924757 ;  12924759 ;  12924760 ;  12924761 ;  12924764 ;  12924766 ;  12924767 ;  12924768 ;  12924769 ;  12924770 ;  12924771 ;  12924772 ;  12924773 ;  12924774 ;  12924775 ;  12924776 ;  12924777 ;  12924778 ;  12924779 ;  12924780 ;  12924781 ;  12924782 ;  12924784 ;  12924785 ;  12924786 ;  12924787 ;  12924788 ;  12924790 ;  12924791 ;  12924792 ;  12924793 ;  12924794 ;  12924795 ;  12924796 ;  12924798 ;  12924799 ;  12924800 ;  12924801 ;  12924802 ;  12924803 ;  12924804 ;  12924805 ;  12924806 ;  12924807 ;  12924809 ;  12924811 ;  12924812 ;  12924813 ;  12924814 ;  12924815 ;  12924816 ;  12924817 ;  12924818 ;  12924819 ;  12924821 ;  12924822 ;  12924824 ;  12924825 ;  12924826 ;  12924827 ;  12924828 ;  12924832 ;  12924834 ;  12924835 ;  12924836 ;  12924837 ;  12924842 ;  12924844 ;  12924845 ;  12924846 ;  12924847 ;  12924848 ;  12924852 ;  12924853 ;  12924858 ;  12924859 ;  12924861 ;  12924863 ;  12924864 ;  12924865 ;  12924868 ;  12924869 ;  12924871 ;  12924873 ;  12924875 ;  12924877 ;  12924878 ;  12924881 ;  12924882 ;  12924886 ;  12924888 ;  12924891 ;  12924892 ;  12924893 ;  12924894 ;  12924895 ;  12924896 ;  12924899 ;  12924902 ;  12924906 ;  12924907 ;  12924908 ;  12924911 ;  12924917 ;  12924918 ;  12924919 ;  12924920 ;  12924925 ;  12924926 ;  12924928 ;  12924929 ;  12924930 ;  12924933 ;  12924936 ;  12924937 ;  12924938 ;  12924939 ;  12924940 ;  12924950 ;  12924952 ;  12924953 ;  12924954 ;  12924964 ;  12924965 ;  12924966 ;  12924971 ;  12924974 ;  12924976 ;  12924977 ;  12924978 ;  12924979 ;  12924981 ;  12924983 ;  12924984 ;  12924985 ;  12924986 ;  12924987 ;  12924988 ;  12924990 ;  12924993 ;  12924994 ;  12924995 ;  12924997 ;  12925000 ;  12925005 ;  12925007 ;  12925008 ;  12925009 ;  12925010 ;  12925012 ;  12925014 ;  12925017 ;  12925018 ;  12925025 ;  12925026 ;  12925029 ;  12925031 ;  12925032 ;  12925034 ;  12925035 ;  12925036 ;  12925037 ;  12925041 ;  12925043 ;  12925045 ;  12925046 ;  12925047 ;  12925048 ;  12925049 ;  12925050 ;  12925051 ;  12925052 ;  12925053 ;  12925054 ;  12925056 ;  12925058 ;  12925059 ;  12925060 ;  12925062 ;  12925063 ;  12925064 ;  12925065 ;  12925066 ;  12925067 ;  12925068 ;  12925069 ;  12925070 ;  12925071 ;  12925072 ;  12925073 ;  12925074 ;  12925076 ;  12925077 ;  12925078 ;  12925079 ;  12925080 ;  12925082 ;  12925084 ;  12925085 ;  12925086 ;  12925089 ;  12925091 ;  12925092 ;  12925095 ;  12925097 ;  12925098 ;  12925100 ;  12925106 ;  12925108 ;  12925109 ;  12925111 ;  12925112 ;  12925121 ;  12925122 ;  12925125 ;",
      "center_classification_date": "20160708",
      "distribution_pattern": "Nationwide Distribution to AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY) + District of Columbia + Puerto Rico + Virgin Islands; *** FOREIGN:",
      "state": "IL",
      "product_description": "The PCA 3 system cam be used in a wide range of clinical settings, including but not limited to: general floor, labor/delivery,post-partum,  burn unit, medical surgical, operating room, oncology, critical care units, post-anesthesia care unit (PACU), and pediatrics. The PCA 3 infusion pump allows clinicians to administer or patients to self-administer analgesia safely and effectively within clinician programmed limits. The epidural route can be used to provide anesthesia or analgesia.",
      "report_date": "20160720",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hospira Inc.",
      "recall_number": "Z-2140-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68318",
      "termination_date": "20161107",
      "more_code_info": "12925127 ;  12925129 ;  12925134 ;  12925137 ;  12925139 ;  12925141 ;  12925142 ;  12925143 ;  12925144 ;  12925145 ;  12925146 ;  12925147 ;  12925148 ;  12925150 ;  12925152 ;  12925154 ;  12925155 ;  12925156 ;  12925157 ;  12925158 ;  12925159 ;  12925160 ;  12925161 ;  12925162 ;  12925163 ;  12925164 ;  12925165 ;  12925167 ;  12925168 ;  12925169 ;  12925170 ;  12925171 ;  12925173 ;  12925174 ;  12925175 ;  12925176 ;  12925177 ;  12925178 ;  12925179 ;  12925181 ;  12925183 ;  12925184 ;  12925185 ;  12925187 ;  12925188 ;  12925190 ;  12925191 ;  12925192 ;  12925193 ;  12925194 ;  12925195 ;  12925196 ;  12925197 ;  12925198 ;  12925199 ;  12925200 ;  12925201 ;  12925202 ;  12925203 ;  12925204 ;  12925205 ;  12925206 ;  12925207 ;  12925208 ;  12925209 ;  12925210 ;  12925211 ;  12925212 ;  12925213 ;  12925214 ;  12925215 ;  12925217 ;  12925218 ;  12925219 ;  12925220 ;  12925221 ;  12925222 ;  12925223 ;  12925224 ;  12925225 ;  12925226 ;  12925227 ;  12925228 ;  12925229 ;  12925230 ;  12925231 ;  12925232 ;  12925234 ;  12925235 ;  12925236 ;  12925237 ;  12925238 ;  12925239 ;  12925240 ;  12925241 ;  12925242 ;  12925243 ;  12925244 ;  12925245 ;  12925246 ;  12925247 ;  12925248 ;  12925249 ;  12925250 ;  12925251 ;  12925252 ;  12925253 ;  12925254 ;  12925255 ;  12925256 ;  12925258 ;  12925259 ;  12925260 ;  12925262 ;  12925263 ;  12925264 ;  12925265 ;  12925266 ;  12925268 ;  12925269 ;  12925270 ;  12925272 ;  12925274 ;  12925276 ;  12925277 ;  12925280 ;  12925282 ;  12925286 ;  12925296 ;  12925297 ;  12925300 ;  12925302 ;  12925304 ;  12925305 ;  12925313 ;  12925314 ;  12925315 ;  12925319 ;  12925321 ;  12925322 ;  12925324 ;  12925326 ;  12925327 ;  12925328 ;  12925329 ;  12925331 ;  12925333 ;  12925334 ;  12925335 ;  12925336 ;  12925340 ;  12925341 ;  12925342 ;  12925343 ;  12925344 ;  12925346 ;  12925347 ;  12925350 ;  12925352 ;  12925353 ;  12925354 ;  12925355 ;  12925356 ;  12925357 ;  12925358 ;  12925359 ;  12925360 ;  12925361 ;  12925362 ;  12925363 ;  12925364 ;  12925365 ;  12925366 ;  12925367 ;  12925369 ;  12925370 ;  12925371 ;  12925372 ;  12925373 ;  12925374 ;  12925375 ;  12925376 ;  12925377 ;  12925378 ;  12925379 ;  12925380 ;  12925381 ;  12925382 ;  12925383 ;  12925384 ;  12925385 ;  12925388 ;  12925389 ;  12925390 ;  12925392 ;  12925393 ;  12925394 ;  12925395 ;  12925396 ;  12925397 ;  12925398 ;  12925401 ;  12925402 ;  12925403 ;  12925405 ;  12925409 ;  12925410 ;  12925411 ;  12925412 ;  12925413 ;  12925414 ;  12925416 ;  12925419 ;  12925420 ;  12925422 ;  12925423 ;  12925424 ;  12925425 ;  12925426 ;  12925427 ;  12925429 ;  12925431 ;  12925437 ;  12925441 ;  12925445 ;  12925446 ;  12925447 ;  12925451 ;  12925456 ;  12925457 ;  12925458 ;  12925460 ;  12925462 ;  12925464 ;  12925465 ;  12925466 ;  12925466 ;  12925466 ;  12925467 ;  12925468 ;  12925473 ;  12925474 ;  12925478 ;  12925480 ;  12925482 ;  12925483 ;  12925485 ;  12925491 ;  12925493 ;  12925494 ;  12925497 ;  12925501 ;  12925505 ;  12925506 ;  12925508 ;  12925510 ;  12925512 ;  12925515 ;  12925516 ;  12925518 ;  12925519 ;  12925520 ;  12925523 ;  12925524 ;  12925533 ;  12925535 ;  12925536 ;  12925538 ;  12925540 ;  12925542 ;  12925544 ;  12925548 ;  12925549 ;  12925550 ;  12925551 ;  12925554 ;  12925555 ;  12925557 ;  12925563 ;  12925564 ;  12925566 ;  12925567 ;  12925575 ;  12925575 ;  12925575 ;  12925577 ;  12925578 ;  12925579 ;  12925580 ;  12925581 ;  12925583 ;  12925589 ;  12925590 ;  12925591 ;  12925595 ;  12925600 ;  12925601 ;  12925603 ;  12925605 ;  12925607 ;  12925608 ;  12925610 ;  12925611 ;  12925612 ;  12925613 ;  12925614 ;  12925617 ;  12925618 ;  12925620 ;  12925621 ;  12925624 ;  12925626 ;  12925627 ;  12925628 ;  12925630 ;  12925631 ;  12925633 ;  12925635 ;  12925637 ;  12925638 ;  12925639 ;  12925641 ;  12925642 ;  12925643 ;  12925644 ;  12925645 ;  12925647 ;  12925648 ;  12925649 ;  12925651 ;  12925652 ;  12925653 ;  12925654 ;  12925655 ;  12925656 ;  12925657 ;  12925658 ;  12925660 ;  12925661 ;  12925662 ;  12925663 ;  12925664 ;  12925665 ;  12925666 ;  12925669 ;  12925671 ;  12925672 ;  12925673 ;  12925677 ;  12925678 ;  12925679 ;  12925680 ;  12925681 ;  12925682 ;  12925683 ;  12925684 ;  12925685 ;  12925686 ;  12925687 ;  12925688 ;  12925689 ;  12925690 ;  12925691 ;  12925692 ;  12925693 ;  12925694 ;  12925695 ;  12925696 ;  12925697 ;  12925698 ;  12925699 ;  12925700 ;  12925701 ;  12925702 ;  12925703 ;  12925704 ;  12925705 ;  12925708 ;  12925710 ;  12925711 ;  12925716 ;  12925717 ;  12925721 ;  12925723 ;  12925728 ;  12925729 ;  12925731 ;  12925733 ;  12925735 ;  12925737 ;  12925738 ;  12925741 ;  12925742 ;  12925744 ;  12925745 ;  12925746 ;  12925747 ;  12925748 ;  12925750 ;  12925752 ;  12925753 ;  12925754 ;  12925755 ;  12925756 ;  12925758 ;  12925759 ;  12925760 ;  12925761 ;  12925761 ;  12925761 ;  12925762 ;  12925764 ;  12925765 ;  12925766 ;  12925767 ;  12925768 ;  12925770 ;  12925771 ;  12925772 ;  12925773 ;  12925773 ;  12925773 ;  12925774 ;  12925775 ;  12925776 ;  12925779 ;  12925779 ;  12925779 ;  12925780 ;  12925781 ;  12925783 ;  12925784 ;  12925787 ;  12925788 ;  12925789 ;  12925792 ;  12925793 ;  12925794 ;  12925795 ;  12925798 ;  12925799 ;  12925802 ;  12925803 ;  12925804 ;  12925807 ;  12925808 ;  12925810 ;  12925811 ;  12925812 ;  12925813 ;  12925814 ;  12925817 ;  12925820 ;  12925821 ;  12925824 ;  12925825 ;  12925827 ;  12925828 ;  12925830 ;  12925831 ;  12925832 ;  12925833 ;  12925834 ;  12925837 ;  12925839 ;  12925840 ;  12925844 ;  12925845 ;  12925848 ;  12925851 ;  12925854 ;  12925857 ;  12925858 ;  12925859 ;  12925862 ;  12925863 ;  12925864 ;  12925866 ;  12925867 ;  12925868 ;  12925869 ;  12925872 ;  12925876 ;  12925880 ;  12925881 ;  12925887 ;  12925888 ;  12925889 ;  12925891 ;  12925892 ;  12925893 ;  12925896 ;  12925897 ;  12925898 ;  12925899 ;  12925900 ;  12925901 ;  12925902 ;  12925906 ;  12925907 ;  12925910 ;  12925911 ;  12925915 ;  12925917 ;  12925920 ;  12925921 ;  12925922 ;  12925923 ;  12925924 ;  12925926 ;  12925928 ;  12925929 ;  12925930 ;  12925931 ;  12925932 ;  12925933 ;  12925936 ;  12925937 ;  12925939 ;  12925940 ;  12925941 ;  12925942 ;  12925943 ;  12925944 ;  12925950 ;  12925951 ;  12925952 ;  12925953 ;  12925954 ;  12925955 ;  12925958 ;  12925960 ;  12925961 ;  12925963 ;  12925964 ;  12925966 ;  12925968 ;  12925972 ;  12925973 ;  12925976 ;  12925977 ;  12925979 ;  12925981 ;  12925983 ;  12925984 ;  12925985 ;  12925986 ;  12925987 ;  12925989 ;  12925991 ;  12925993 ;  12925994 ;  12925995 ;  12925996 ;  12925999 ;  12926002 ;  12926003 ;  12926004 ;  12926006 ;  12926008 ;  12926009 ;  12926010 ;  12926011 ;  12926012 ;  12926015 ;  12926017 ;  12926019 ;  12926020 ;  12926021 ;  12926022 ;  12926024 ;  12926026 ;  12926027 ;  12926029 ;  12926030 ;  12926031 ;  12926033 ;  12926035 ;  12926036 ;  12926037 ;  12926038 ;  12926039 ;  12926040 ;  12926041 ;  12926043 ;  12926044 ;  12926045 ;  12926047 ;  12926048 ;  12926049 ;  12926065 ;  12926066 ;  12926067 ;  12926068 ;  12926069 ;  12926070 ;  12926072 ;  12926074 ;  12926075 ;  12926076 ;  12926077 ;  12926078 ;  12926079 ;  12926080 ;  12926081 ;  12926082 ;  12926083 ;  12926084 ;  12926085 ;  12926086 ;  12926087 ;  12926088 ;  12926089 ;  12926090 ;  12926091 ;  12926092 ;  12926093 ;  12926094 ;  12926095 ;  12926096 ;  12926097 ;  12926098 ;  12926099 ;  12926100 ;  12926101 ;  12926103 ;  12926104 ;  12926108 ;  12926111 ;  12926112 ;  12926113 ;  12926116 ;  12926121 ;  12926123 ;  12926124 ;  12926127 ;  12926129 ;  12926130 ;  12926132 ;  12926133 ;  12926135 ;  12926136 ;  12926137 ;  12926138 ;  12926139 ;  12926140 ;  12926141 ;  12926142 ;  12926143 ;  12926144 ;  12926145 ;  12926146 ;  12926149 ;  12926150 ;  12926153 ;  12926154 ;  12926155 ;  12926157 ;  12926158 ;  12926159 ;  12926160 ;  12926161 ;  12926162 ;  12926164 ;  12926165 ;  12926167 ;  12926169 ;  12926174 ;  12926176 ;  12926177 ;  12926183 ;  12926184 ;  12926185 ;  12926188 ;  12926191 ;  12926194 ;  12926195 ;  12926196 ;  12926197 ;  12926199 ;  12926201 ;  12926202 ;  12926203 ;  12926205 ;  12926207 ;  12926209 ;  12926210 ;  12926211 ;  12926212 ;  12926213 ;  12926214 ;  12926216 ;  12926218 ;  12926220 ;  12926222 ;  12926224 ;  12926225 ;  12926226 ;  12926227 ;  12926232 ;  12926233 ;  12926235 ;  12926237 ;  12926240 ;  12926241 ;  12926243 ;  12926244 ;  12926245 ;  12926246 ;  12926247 ;  12926248 ;  12926249 ;  12926251 ;  12926252 ;  12926254 ;  12926255 ;  12926257 ;  12926259 ;  12926261 ;  12926262 ;  12926263 ;  12926271 ;  12926273 ;  12926277 ;  12926278 ;  12926279 ;  12926280 ;  12926281 ;  12926282 ;  12926286 ;  12926289 ;  12926295 ;  12926296 ;  12926299 ;  12926301 ;  12926305 ;  12926306 ;  12926311 ;  12926312 ;  12926313 ;  12926315 ;  12926318 ;  12926323 ;  12926326 ;  12926330 ;  12926333 ;  12926336 ;  12926343 ;  12926344 ;  12926347 ;  12926348 ;  12926351 ;  12926354 ;  12926355 ;  12926358 ;  12926362 ;  12926365 ;  12926366 ;  12926369 ;  12926370 ;  12926373 ;  12926374 ;  12926375 ;  12926377 ;  12926380 ;  12926381 ;  12926383 ;  12926384 ;  12926388 ;  12926392 ;  12926394 ;  12926395 ;  12926396 ;  12926398 ;  12926399 ;  12926400 ;  12926401 ;  12926402 ;  12926403 ;  12926407 ;  12926408 ;  12926410 ;  12926411 ;  12926419 ;  12926421 ;  12926422 ;  12926426 ;  12926428 ;  12926430 ;  12926431 ;  12926434 ;  12926435 ;  12926437 ;  12926438 ;  12926439 ;  12926441 ;  12926448 ;  12926451 ;  12926452 ;  12926453 ;  12926456 ;  12926457 ;  12926458 ;  12926459 ;  12926460 ;  12926461 ;  12926463 ;  12926464 ;  12926465 ;  12926469 ;  12926470 ;  12926473 ;  12926474 ;  12926475 ;  12926477 ;  12926477 ;  12926477 ;  12926478 ;  12926479 ;  12926481 ;  12926481 ;  12926481 ;  12926482 ;  12926483 ;  12926483 ;  12926483 ;  12926486 ;  12926489 ;  12926489 ;  12926489 ;  12926490 ;  12926491 ;  12926491 ;  12926491 ;  12926492 ;  12926492 ;  12926492 ;  12926493 ;  12926493 ;  12926493 ;  12926494 ;  12926495 ;  12926495 ;  12926495 ;  12926496 ;  12926497 ;  12926500 ;  12926502 ;  12926504 ;  12926504 ;  12926504 ;  12926506 ;  12926507 ;  12926508 ;  12926510 ;  12926511 ;  12926511 ;  12926511 ;  12926512 ;  12926513 ;  12926514 ;  12926515 ;  12926515 ;  12926515 ;  12926516 ;  12926516 ;  12926516 ;  12926518 ;  12926518 ;  12926518 ;  12926519 ;  12926520 ;  12926520 ;  12926520 ;  12926521 ;  12926522 ;  12926523 ;  12926523 ;  12926523 ;  12926525 ;  12926526 ;  12926527 ;  12926527 ;  12926527 ;  12926528 ;  12926528 ;  12926528 ;  12926529 ;  12926530 ;  12926531 ;  12926532 ;  12926532 ;  12926532 ;  12926534 ;  12926534 ;  12926534 ;  12926536 ;  12926536 ;  12926536 ;  12926537 ;  12926537 ;  12926537 ;  12926538 ;  12926538 ;  12926538 ;  12926540 ;  12926540 ;  12926540 ;  12926541 ;  12926541 ;  12926541 ;  12926542 ;  12926543 ;  12926544 ;  12926545 ;  12926546 ;  12926548 ;  12926548 ;  12926548 ;  12926549 ;  12926550 ;  12926551 ;  12926552 ;  12926553 ;  12926555 ;  12926557 ;  12926558 ;  12926559 ;  12926560 ;  12926561 ;  12926562 ;  12926563 ;  12926564 ;  12926565 ;  12926569 ;  12926570 ;  12926572 ;  12926573 ;  12926574 ;  12926575 ;  12926576 ;  12926578 ;  12926580 ;  12926581 ;  12926582 ;  12926584 ;  12926585 ;  12926587 ;  12926588 ;  12926589 ;  12926590 ;  12926591 ;  12926592 ;  12926593 ;  12926594 ;  12926595 ;  12926596 ;  12926597 ;  12926598 ;  12926600 ;  12926602 ;  12926603 ;  12926604 ;  12926606 ;  12926608 ;  12926613 ;  12926615 ;  12926622 ;  12926624 ;  12926630 ;  12926638 ;  12926639 ;  12926642 ;  12926645 ;  12926646 ;  12926647 ;  12926652 ;  12926656 ;  12926659 ;  12926660 ;  12926661 ;  12926662 ;  12926664 ;  12926669 ;  12926670 ;  12926672 ;  12926673 ;  12926677 ;  12926679 ;  12926680 ;  12926683 ;  12926684 ;  12926687 ;  12926688 ;  12926689 ;  12926690 ;  12926692 ;  12926694 ;  12926696 ;  12926697 ;  12926703 ;  12926706 ;  12926709 ;  12926710 ;  12926711 ;  12926712 ;  12926713 ;  12926715 ;  12926716 ;  12926718 ;  12926722 ;  12926723 ;  12926725 ;  12926726 ;  12926727 ;  12926728 ;  12926729 ;  12926730 ;  12926732 ;  12926733 ;  12926734 ;  12926735 ;  12926736 ;  12926737 ;  12926738 ;  12926739 ;  12926740 ;  12926741 ;  12926742 ;  12926743 ;  12926744 ;  12926745 ;  12926746 ;  12926748 ;  12926749 ;  12926750 ;  12926751 ;  12926752 ;  12926753 ;  12926754 ;  12926755 ;  12926757 ;  12926759 ;  12926761 ;  12926762 ;  12926763 ;  12926764 ;  12926766 ;  12926767 ;  12926768 ;  12926769 ;  12926770 ;  12926771 ;  12926772 ;  12926775 ;  12926776 ;  12926777 ;  12926781 ;  12926782 ;  12926784 ;  12926785 ;  12926787 ;  12926788 ;  12926789 ;  12926791 ;  12926792 ;  12926793 ;  12926794 ;  12926795 ;  12926796 ;  12926800 ;  12926809 ;  12926813 ;  12926815 ;  12926816 ;  12926819 ;  12926824 ;  12926830 ;  12926832 ;  12926837 ;  12926840 ;  12926849 ;  12926853 ;  12926854 ;  12926855 ;  12926858 ;  12926859 ;  12926862 ;  12926863 ;  12926864 ;  12926866 ;  12926872 ;  12926873 ;  12926877 ;  12926879 ;  12926880 ;  12926883 ;  12926885 ;  12926887 ;  12926890 ;  12926891 ;  12926898 ;  12926899 ;  12926900 ;  12926907 ;  12926910 ;  12926911 ;  12926915 ;  12926916 ;  12926925 ;  12926928 ;  12926929 ;  12926932 ;  12926938 ;  12926941 ;  12926946 ;  12926948 ;  12926949 ;  12926953 ;  12926954 ;  12926956 ;  12926958 ;  12926962 ;  12926963 ;  12926965 ;  12926966 ;  12926967 ;  12926969 ;  12926970 ;  12926971 ;  12926972 ;  12926973 ;  12926974 ;  12926975 ;  12926976 ;  12926978 ;  12926981 ;  12926982 ;  12926983 ;  12926984 ;  12926985 ;  12926986 ;  12926987 ;  12926988 ;  12926989 ;  12926990 ;  12926991 ;  12926993 ;  12926994 ;  12926995 ;  12926996 ;  12926997 ;  12926998 ;  12926999 ;  12927000 ;  12927001 ;  12927002 ;  12927003 ;  12927004 ;  12927005 ;  12927008 ;  12927009 ;  12927010 ;  12927012 ;  12927013 ;  12927014 ;  12927015 ;  12927016 ;  12927017 ;  12927018 ;  12927019 ;  12927020 ;  12927021 ;  12927022 ;  12927023 ;  12927024 ;  12927025 ;  12927026 ;  12927027 ;  12927028 ;  12927029 ;  12927030 ;  12927031 ;  12927032 ;  12927033 ;  12927034 ;  12927035 ;  12927036 ;  12927037 ;  12927041 ;  12927042 ;  12927043 ;  12927044 ;  12927045 ;  12927046 ;  12927047 ;  12927049 ;  12927051 ;  12927052 ;  12927053 ;  12927055 ;  12927056 ;  12927057 ;  12927058 ;  12927059 ;  12927061 ;  12927062 ;  12927063 ;  12927064 ;  12927065 ;  12927067 ;  12927069 ;  12927070 ;  12927071 ;  12927072 ;  12927073 ;  12927074 ;  12927075 ;  12927077 ;  12927078 ;  12927079 ;  12927080 ;  12927081 ;  12927082 ;  12927083 ;  12927084 ;  12927085 ;  12927086 ;  12927087 ;  12927089 ;  12927090 ;  12927091 ;  12927092 ;  12927093 ;  12927095 ;  12927097 ;  12927099 ;  12927100 ;  12927101 ;  12927102 ;  12927103 ;  12927104 ;  12927106 ;  12927107 ;  12927108 ;  12927109 ;  12927110 ;  12927111 ;  12927112 ;  12927113 ;  12927114 ;  12927115 ;  12927117 ;  12927118 ;  12927119 ;  12927121 ;  12927123 ;  12927125 ;  12927126 ;  12927127 ;  12927130 ;  12927131 ;  12927132 ;  12927134 ;  12927136 ;  12927137 ;  12927138 ;  12927141 ;  12927142 ;  12927143 ;  12927144 ;  12927145 ;  12927146 ;  12927147 ;  12927148 ;  12927149 ;  12927150 ;  12927151 ;  12927153 ;  12927155 ;  12927156 ;  12927157 ;  12927158 ;  12927160 ;  12927161 ;  12927162 ;  12927164 ;  12927165 ;  12927166 ;  12927167 ;  12927169 ;  12927170 ;  12927171 ;  12927172 ;  12927173 ;  12927174 ;  12927175 ;  12927176 ;  12927177 ;  12927178 ;  12927179 ;  12927180 ;  12927181 ;  12927182 ;  12927184 ;  12927185 ;  12927186 ;  12927187 ;  12927188 ;  12927189 ;  12927190 ;  12927191 ;  12927192 ;  12927194 ;  12927195 ;  12927196 ;  12927197 ;  12927198 ;  12927200 ;  12927201 ;  12927202 ;  12927203 ;  12927204 ;  12927205 ;  12927206 ;  12927207 ;  12927208 ;  12927210 ;  12927211 ;  12927212 ;  12927213 ;  12927214 ;  12927215 ;  12927219 ;  12927223 ;  12927224 ;  12927226 ;  12927227 ;  12927228 ;  12927229 ;  12927230 ;  12927231 ;  12927232 ;  12927233 ;  12927234 ;  12927235 ;  12927237 ;  12927238 ;  12927239 ;  12927240 ;  12927242 ;  12927244 ;  12927245 ;  12927246 ;  12927248 ;  12927251 ;  12927252 ;  12927253 ;  12927254 ;  12927262 ;  12927263 ;  12927264 ;  12927265 ;  12927266 ;  12927268 ;  12927269 ;  12927270 ;  12927271 ;  12927272 ;  12927273 ;  12927274 ;  12927277 ;  12927278 ;  12927279 ;  12927280 ;  12927281 ;  12927282 ;  12927283 ;  12927285 ;  12927286 ;  12927288 ;  12927289 ;  12927290 ;  12927291 ;  12927292 ;  12927293 ;  12927295 ;  12927296 ;  12927297 ;  12927298 ;  12927299 ;  12927301 ;  12927302 ;  12927303 ;  12927306 ;  12927307 ;  12927308 ;  12927309 ;  12927311 ;  12927312 ;  12927313 ;  12927314 ;  12927315 ;  12927316 ;  12927317 ;  12927318 ;  12927320 ;  12927321 ;  12927322 ;  12927323 ;  12927324 ;  12927326 ;  12927327 ;  12927328 ;  12927342 ;  12927344 ;  12927346 ;  12927347 ;  12927348 ;  12927349 ;  12927351 ;  12927352 ;  12927354 ;  12927356 ;  12927357 ;  12927358 ;  12927359 ;  12927360 ;  12927361 ;  12927363 ;  12927364 ;  12927367 ;  12927368 ;  12927369 ;  12927370 ;  12927371 ;  12927372 ;  12927373 ;  12927375 ;  12927376 ;  12927377 ;  12927378 ;  12927380 ;  12927381 ;  12927382 ;  12927383 ;  12927384 ;  12927385 ;  12927386 ;  12927387 ;  12927388 ;  12927389 ;  12927390 ;  12927391 ;  12927392 ;  12927395 ;  12927396 ;  12927397 ;  12927398 ;  12927399 ;  12927400 ;  12927401 ;  12927402 ;  12927403 ;  12927404 ;  12927405 ;  12927407 ;  12927408 ;  12927409 ;  12927410 ;  12927411 ;  12927412 ;  12927413 ;  12927414 ;  12927415 ;  12927416 ;  12927417 ;  12927418 ;  12927420 ;  12927421 ;  12927422 ;  12927423 ;  12927424 ;  12927425 ;  12927426 ;  12927427 ;  12927428 ;  12927429 ;  12927430 ;  12927431 ;  12927432 ;  12927433 ;  12927434 ;  12927435 ;  12927437 ;  12927439 ;  12927440 ;  12927441 ;  12927445 ;  12927447 ;  12927450 ;  12927451 ;  12927452 ;  12927454 ;  12927455 ;  12927456 ;  12927458 ;  12927465 ;  12927466 ;  12927467 ;  12927468 ;  12927469 ;  12927470 ;  12927471 ;  12927472 ;  12927476 ;  12927479 ;  12927480 ;  12927490 ;  12927494 ;  12927507 ;  12927508 ;  12927514 ;  12927515 ;  12927516 ;  12927517 ;  12927520 ;  12927524 ;  12927525 ;  12927526 ;  12927527 ;  12927529 ;  12927531 ;  12927533 ;  12927534 ;  12927535 ;  12927536 ;  12927537 ;  12927538 ;  12927541 ;  12927544 ;  12927548 ;  12927549 ;  12927550 ;  12927553 ;  12927554 ;  12927556 ;  12927557 ;  12927558 ;  12927559 ;  12927560 ;  12927561 ;  12927562 ;  12927563 ;  12927564 ;  12927565 ;  12927566 ;  12927567 ;  12927570 ;  12927571 ;  12927572 ;  12927573 ;  12927574 ;  12927575 ;  12927576 ;  12927577 ;  12927578 ;  12927579 ;  12927580 ;  12927581 ;  12927582 ;  12927583 ;  12927585 ;  12927586 ;  12927587 ;  12927589 ;  12927590 ;  12927591 ;  12927596 ;  12927603 ;  12927605 ;  12927606 ;  12927613 ;  12927615 ;  12927617 ;  12927619 ;  12927620 ;  12927623 ;  12927626 ;  12927650 ;  12927674 ;  12927675 ;  12927677 ;  12927678 ;  12927679 ;  12927680 ;  12927681 ;  12927682 ;  12927683 ;  12927684 ;  12927685 ;  12927687 ;  12927688 ;  12927689 ;  12927690 ;  12927691 ;  12927692 ;  12927693 ;  12927694 ;  12927696 ;  12927698 ;  12927699 ;  12927700 ;  12927702 ;  12927704 ;  12927706 ;  12927707 ;  12927708 ;  12927709 ;  12927710 ;  12927711 ;  12927712 ;  12927713 ;  12927714 ;  12927715 ;  12927716 ;  12927717 ;  12927718 ;  12927719 ;  12927720 ;  12927721 ;  12927722 ;  12927723 ;  12927724 ;  12927725 ;  12927726 ;  12927727 ;  12927728 ;  12927729 ;  12927730 ;  12927731 ;  12927732 ;  12927733 ;  12927734 ;  12927736 ;  12927737 ;  12927738 ;  12927739 ;  12927740 ;  12927741 ;  12927742 ;  12927743 ;  12927744 ;  12927745 ;  12927748 ;  12927749 ;  12927752 ;  12927753 ;  12927754 ;  12927756 ;  12927757 ;  12927758 ;  12927762 ;  12927763 ;  12927764 ;  12927765 ;  12927766 ;  12927766 ;  12927766 ;  12927767 ;  12927769 ;  12927770 ;  12927771 ;  12927780 ;  12927787 ;  12927789 ;  12927793 ;  12927805 ;  12927829 ;  12927856 ;  12927857 ;  12927867 ;  12927874 ;  12927879 ;  12927881 ;  12927889 ;  12927898 ;  12927900 ;  12927902 ;  12927903 ;  12927906 ;  12927908 ;  12927909 ;  12927912 ;  12927917 ;  12927918 ;  12927921 ;  12927923 ;  12927925 ;  12927926 ;  12927931 ;  12927932 ;  12927933 ;  12927934 ;  12927935 ;  12927936 ;  12927937 ;  12927938 ;  12927939 ;  12927941 ;  12927944 ;  12927945 ;  12927946 ;  12927947 ;  12927948 ;  12927951 ;  12927952 ;  12927955 ;  12927956 ;  12927957 ;  12927958 ;  12927962 ;  12927964 ;  12927967 ;  12927968 ;  12927970 ;  12927971 ;  12927972 ;  12927973 ;  12927974 ;  12927978 ;  12927979 ;  12927980 ;  12927981 ;  12927983 ;  12927984 ;  12927985 ;  12927986 ;  12927987 ;  12927988 ;  12927989 ;  12927990 ;  12927991 ;  12927993 ;  12927994 ;  12927995 ;  12927997 ;  12927998 ;  12927999 ;  12928000 ;  12928001 ;  12928002 ;  12928003 ;  12928004 ;  12928005 ;  12928006 ;  12928007 ;  12928009 ;  12928010 ;  12928011 ;  12928012 ;  12928013 ;  12928014 ;  12928015 ;  12928016 ;  12928017 ;  12928018 ;  12928019 ;  12928020 ;  12928024 ;  12928025 ;  12928026 ;  12928027 ;  12928028 ;  12928029 ;  12928030 ;  12928031 ;  12928032 ;  12928033 ;  12928034 ;  12928035 ;  12928036 ;  12928037 ;  12928038 ;  12928039 ;  12928040 ;  12928041 ;  12928042 ;  12928043 ;  12928044 ;  12928045 ;  12928046 ;  12928047 ;  12928048 ;  12928051 ;  12928052 ;  12928053 ;  12928055 ;  12928056 ;  12928058 ;  12928059 ;  12928060 ;  12928061 ;  12928062 ;  12928063 ;  12928064 ;  12928065 ;  12928066 ;  12928067 ;  12928068 ;  12928069 ;  12928070 ;  12928071 ;  12928072 ;  12928073 ;  12928074 ;  12928076 ;  12928077 ;  12928080 ;  12928082 ;  12928084 ;  12928085 ;  12928086 ;  12928087 ;  12928090 ;  12928091 ;  12928092 ;  12928094 ;  12928095 ;  12928096 ;  12928097 ;  12928098 ;  12928099 ;  12928102 ;  12928103 ;  12928104 ;  12928105 ;  12928106 ;  12928107 ;  12928109 ;  12928110 ;  12928111 ;  12928112 ;  12928113 ;  12928115 ;  12928121 ;  12928123 ;  12928130 ;  12928135 ;  12928137 ;  12928138 ;  12928142 ;  12928147 ;  12928151 ;  12928154 ;  12928156 ;  12928159 ;  12928160 ;  12928161 ;  12928162 ;  12928163 ;  12928164 ;  12928166 ;  12928167 ;  12928168 ;  12928175 ;  12928178 ;  12928179 ;  12928180 ;  12928182 ;  12928183 ;  12928185 ;  12928188 ;  12928191 ;  12928194 ;  12928197 ;  12928198 ;  12928200 ;  12928201 ;  12928206 ;  12928208 ;  12928209 ;  12928210 ;  12928211 ;  12928212 ;  12928215 ;  12928216 ;  12928218 ;  12928221 ;  12928222 ;  12928224 ;  12928225 ;  12928226 ;  12928227 ;  12928229 ;  12928230 ;  12928231 ;  12928234 ;  12928236 ;  12928241 ;  12928250 ;  12928258 ;  12928260 ;  12928261 ;  12928262 ;  12928264 ;  12928265 ;  12928267 ;  12928268 ;  12928269 ;  12928270 ;  12928271 ;  12928273 ;  12928274 ;  12928275 ;  12928276 ;  12928277 ;  12928278 ;  12928281 ;  12928282 ;  12928285 ;  12928286 ;  12928290 ;  12928291 ;  12928296 ;  12928298 ;  12928300 ;  12928303 ;  12928304 ;  12928305 ;  12928306 ;  12928307 ;  12928308 ;  12928309 ;  12928310 ;  12928311 ;  12928312 ;  12928313 ;  12928314 ;  12928316 ;  12928317 ;  12928318 ;  12928319 ;  12928320 ;  12928322 ;  12928323 ;  12928324 ;  12928325 ;  12928326 ;  12928327 ;  12928329 ;  12928331 ;  12928332 ;  12928333 ;  12928334 ;  12928337 ;  12928338 ;  12928339 ;  12928340 ;  12928341 ;  12928342 ;  12928344 ;  12928350 ;  12928351 ;  12928353 ;  12928354 ;  12928355 ;  12928356 ;  12928357 ;  12928358 ;  12928361 ;  12928362 ;  12928363 ;  12928364 ;  12928366 ;  12928367 ;  12928368 ;  12928369 ;  12928371 ;  12928372 ;  12928378 ;  12928379 ;  12928382 ;  12928384 ;  12928385 ;  12928386 ;  12928388 ;  12928390 ;  12928391 ;  12928392 ;  12928394 ;  12928396 ;  12928398 ;  12928399 ;  12928402 ;  12928404 ;  12928407 ;  12928408 ;  12928409 ;  12928410 ;  12928411 ;  12928412 ;  12928414 ;  12928415 ;  12928416 ;  12928418 ;  12928419 ;  12928420 ;  12928421 ;  12928422 ;  12928423 ;  12928424 ;  12928427 ;  12928428 ;  12928429 ;  12928430 ;  12928432 ;  12928437 ;  12928443 ;  12928445 ;  12928446 ;  12928447 ;  12928450 ;  12928453 ;  12928456 ;  12928459 ;  12928461 ;  12928462 ;  12928463 ;  12928464 ;  12928465 ;  12928466 ;  12928467 ;  12928468 ;  12928469 ;  12928470 ;  12928471 ;  12928472 ;  12928473 ;  12928474 ;  12928475 ;  12928476 ;  12928477 ;  12928479 ;  12928480 ;  12928482 ;  12928483 ;  12928486 ;  12928487 ;  12928489 ;  12928493 ;  12928496 ;  12928499 ;  12928500 ;  12928507 ;  12928512 ;  12928513 ;  12928514 ;  12928515 ;  12928516 ;  12928517 ;  12928518 ;  12928519 ;  12928520 ;  12928521 ;  12928522 ;  12928523 ;  12928524 ;  12928525 ;  12928526 ;  12928527 ;  12928528 ;  12928529 ;  12928530 ;  12928531 ;  12928532 ;  12928533 ;  12928534 ;  12928535 ;  12928536 ;  12928537 ;  12928538 ;  12928539 ;  12928540 ;  12928541 ;  12928542 ;  12928543 ;  12928544 ;  12928545 ;  12928546 ;  12928548 ;  12928549 ;  12928550 ;  12928551 ;  12928552 ;  12928553 ;  12928554 ;  12928555 ;  12928556 ;  12928557 ;  12928558 ;  12928560 ;  12928561 ;  12928562 ;  12928563 ;  12928564 ;  12928565 ;  12928566 ;  12928567 ;  12928568 ;  12928569 ;  12928570 ;  12928571 ;  12928572 ;  12928573 ;  12928574 ;  12928575 ;  12928576 ;  12928577 ;  12928578 ;  12928579 ;  12928580 ;  12928581 ;  12928582 ;  12928583 ;  12928584 ;  12928585 ;  12928586 ;  12928587 ;  12928589 ;  12928590 ;  12928591 ;  12928592 ;  12928593 ;  12928594 ;  12928595 ;  12928596 ;  12928598 ;  12928602 ;  12928603 ;  12928606 ;  12928607 ;  12928608 ;  12928609 ;  12928610 ;  12928611 ;  12928612 ;  12928613 ;  12928614 ;  12928616 ;  12928617 ;  12928618 ;  12928619 ;  12928620 ;  12928621 ;  12928622 ;  12928623 ;  12928624 ;  12928625 ;  12928626 ;  12928627 ;  12928628 ;  12928629 ;  12928630 ;  12928631 ;  12928632 ;  12928633 ;  12928634 ;  12928636 ;  12928637 ;  12928640 ;  12928641 ;  12928642 ;  12928643 ;  12928644 ;  12928645 ;  12928646 ;  12928647 ;  12928648 ;  12928649 ;  12928651 ;  12928652 ;  12928653 ;  12928654 ;  12928655 ;  12928656 ;  12928657 ;  12928659 ;  12928660 ;  12928661 ;  12928662 ;  12928663 ;  12928664 ;  12928666 ;  12928670 ;  12928673 ;  12928674 ;  12928675 ;  12928676 ;  12928677 ;  12928682 ;  12928684 ;  12928685 ;  12928686 ;  12928687 ;  12928688 ;  12928689 ;  12928690 ;  12928693 ;  12928694 ;  12928695 ;  12928696 ;  12928698 ;  12928700 ;  12928703 ;  12928706 ;  12928708 ;  12928710 ;  12928712 ;  12928713 ;  12928717 ;  12928718 ;  12928722 ;  12928723 ;  12928725 ;  12928726 ;  12928731 ;  12928732 ;  12928734 ;  12928735 ;  12928740 ;  12928741 ;  12928742 ;  12928744 ;  12928745 ;  12928746 ;  12928749 ;  12928750 ;  12928751 ;  12928752 ;  12928753 ;  12928754 ;  12928755 ;  12928757 ;  12928758 ;  12928759 ;  12928761 ;  12928762 ;  12928763 ;  12928764 ;  12928766 ;  12928767 ;  12928768 ;  12928769 ;  12928770 ;  12928771 ;  12928774 ;  12928776 ;  12928777 ;  12928778 ;  12928779 ;  12928781 ;  12928782 ;  12928787 ;  12928788 ;  12928789 ;  12928790 ;  12928791 ;  12928792 ;  12928793 ;  12928794 ;  12928796 ;  12928797 ;  12928798 ;  12928799 ;  12928800 ;  12928801 ;  12928802 ;  12928804 ;  12928805 ;  12928806 ;  12928807 ;  12928809 ;  12928810 ;  12928811 ;  12928812 ;  12928813 ;  12928816 ;  12928819 ;  12928820 ;  12928822 ;  12928823 ;  12928824 ;  12928825 ;  12928826 ;  12928828 ;  12928832 ;  12928833 ;  12928834 ;  12928835 ;  12928836 ;  12928838 ;  12928839 ;  12928840 ;  12928841 ;  12928843 ;  12928844 ;  12928845 ;  12928846 ;  12928847 ;  12928848 ;  12928849 ;  12928851 ;  12928852 ;  12928853 ;  12928854 ;  12928855 ;  12928856 ;  12928857 ;  12928858 ;  12928859 ;  12928860 ;  12928861 ;  12928864 ;  12928865 ;  12928866 ;  12928867 ;  12928868 ;  12928870 ;  12928871 ;  12928872 ;  12928873 ;  12928874 ;  12928875 ;  12928876 ;  12928877 ;  12928878 ;  12928879 ;  12928880 ;  12928882 ;  12928883 ;  12928884 ;  12928885 ;  12928886 ;  12928887 ;  12928889 ;  12928890 ;  12928893 ;  12928894 ;  12928895 ;  12928896 ;  12928897 ;  12928898 ;  12928899 ;  12928900 ;  12928901 ;  12928902 ;  12928903 ;  12928904 ;  12928905 ;  12928906 ;  12928907 ;  12928908 ;  12928909 ;  12928911 ;  12928913 ;  12928914 ;  12928915 ;  12928917 ;  12928918 ;  12928919 ;  12928920 ;  12928921 ;  12928922 ;  12928923 ;  12928924 ;  12928925 ;  12928926 ;  12928928 ;  12928929 ;  12928930 ;  12928931 ;  12928932 ;  12928933 ;  12928934 ;  12928935 ;  12928936 ;  12928937 ;  12928938 ;  12928939 ;  12928940 ;  12928941 ;  12928942 ;  12928943 ;  12928944 ;  12928945 ;  12928946 ;  12928947 ;  12928948 ;  12928950 ;  12928951 ;  12928952 ;  12928953 ;  12928954 ;  12928955 ;  12928957 ;  12928958 ;  12928959 ;  12928960 ;  12928961 ;  12928962 ;  12928963 ;  12928964 ;  12928965 ;  12928967 ;  12928968 ;  12928969 ;  12928970 ;  12928971 ;  12928972 ;  12928973 ;  12928974 ;  12928975 ;  12928976 ;  12928977 ;  12928978 ;  12928979 ;  12928980 ;  12928981 ;  12928983 ;  12928984 ;  12928985 ;  12928987 ;  12928988 ;  12928989 ;  12928991 ;  12928992 ;  12928993 ;  12928994 ;  12928995 ;  12928997 ;  12928998 ;  12928999 ;  12929000 ;  12929001 ;  12929002 ;  12929003 ;  12929004 ;  12929005 ;  12929006 ;  12929007 ;  12929008 ;  12929009 ;  12929012 ;  12929013 ;  12929014 ;  12929015 ;  12929016 ;  12929017 ;  12929018 ;  12929019 ;  12929020 ;  12929021 ;  12929022 ;  12929023 ;  12929024 ;  12929025 ;  12929026 ;  12929027 ;  12929028 ;  12929029 ;  12929030 ;  12929031 ;  12929033 ;  12929034 ;  12929036 ;  12929037 ;  12929038 ;  12929039 ;  12929040 ;  12929041 ;  12929042 ;  12929043 ;  12929044 ;  12929045 ;  12929047 ;  12929048 ;  12929049 ;  12929050 ;  12929051 ;  12929052 ;  12929053 ;  12929054 ;  12929055 ;  12929056 ;  12929057 ;  12929058 ;  12929059 ;  12929060 ;  12929061 ;  12929062 ;  12929063 ;  12929064 ;  12929066 ;  12929067 ;  12929068 ;  12929069 ;  12929070 ;  12929072 ;  12929075 ;  12929076 ;  12929078 ;  12929079 ;  12929081 ;  12929083 ;  12929084 ;  12929087 ;  12929089 ;  12929090 ;  12929091 ;  12929092 ;  12929093 ;  12929094 ;  12929095 ;  12929097 ;  12929099 ;  12929103 ;  12929105 ;  12929106 ;  12929107 ;  12929109 ;  12929111 ;  12929116 ;  12929117 ;  12929119 ;  12929120 ;  12929121 ;  12929122 ;  12929123 ;  12929124 ;  12929125 ;  12929126 ;  12929128 ;  12929131 ;  12929132 ;  12929135 ;  12929138 ;  12929139 ;  12929140 ;  12929141 ;  12929142 ;  12929143 ;  12929145 ;  12929146 ;  12929147 ;  12929148 ;  12929149 ;  12929151 ;  12929156 ;  12929160 ;  12929161 ;  12929162 ;  12929163 ;  12929164 ;  12929165 ;  12929166 ;  12929167 ;  12929168 ;  12929169 ;  12929170 ;  12929171 ;  12929172 ;  12929175 ;  12929176 ;  12929177 ;  12929178 ;  12929179 ;  12929180 ;  12929181 ;  12929182 ;  12929183 ;  12929184 ;  12929185 ;  12929188 ;  12929190 ;  12929191 ;  12929192 ;  12929194 ;  12929195 ;  12929196 ;  12929197 ;  12929198 ;  12929201 ;  12929202 ;  12929205 ;  12929206 ;  12929208 ;  12929209 ;  12929210 ;  12929212 ;  12929213 ;  12929214 ;  12929216 ;  12929217 ;  12929220 ;  12929221 ;  12929222 ;  12929223 ;  12929224 ;  12929225 ;  12929227 ;  12929228 ;  12929229 ;  12929230 ;  12929231 ;  12929232 ;  12929233 ;  12929234 ;  12929236 ;  12929240 ;  12929241 ;  12929242 ;  12929244 ;  12929247 ;  12929249 ;  12929252 ;  12929253 ;  12929255 ;  12929259 ;  12929260 ;  12929263 ;  12929264 ;  12929270 ;  12929271 ;  12929272 ;  12929274 ;  12929276 ;  12929277 ;  12929283 ;  12929284 ;  12929285 ;  12929286 ;  12929288 ;  12929289 ;  12929290 ;  12929291 ;  12929292 ;  12929295 ;  12929298 ;  12929299 ;  12929300 ;  12929301 ;  12929302 ;  12929303 ;  12929306 ;  12929307 ;  12929309 ;  12929310 ;  12929311 ;  12929312 ;  12929313 ;  12929314 ;  12929315 ;  12929316 ;  12929317 ;  12929318 ;  12929319 ;  12929321 ;  12929323 ;  12929324 ;  12929325 ;  12929327 ;  12929329 ;  12929331 ;  12929332 ;  12929333 ;  12929334 ;  12929335 ;  12929336 ;  12929337 ;  12929338 ;  12929339 ;  12929340 ;  12929341 ;  12929342 ;  12929343 ;  12929344 ;  12929345 ;  12929349 ;  12929350 ;  12929351 ;  12929352 ;  12929353 ;  12929354 ;  12929355 ;  12929356 ;  12929357 ;  12929358 ;  12929359 ;  12929360 ;  12929361 ;  12929362 ;  12929363 ;  12929364 ;  12929365 ;  12929366 ;  12929369 ;  12929370 ;  12929371 ;  12929373 ;  12929374 ;  12929375 ;  12929376 ;  12929377 ;  12929378 ;  12929380 ;  12929381  ;  12929382 ;  12929385 ;  12929386 ;  12929387 ;  12929388 ;  12929389 ;  12929390 ;  12929391 ;  12929392 ;  12929393 ;  12929394 ;  12929395 ;  12929396 ;  12929400 ;  12929401 ;  12929402 ;  12929403 ;  12929404 ;  12929405 ;  12929406 ;  12929407 ;  12929408 ;  12929409 ;  12929410 ;  12929414 ;  12929420 ;  12929421 ;  12929422 ;  12929423 ;  12929426 ;  12929427 ;  12929428 ;  12929429 ;  12929430 ;  12929431 ;  12929432 ;  12929433 ;  12929434 ;  12929435 ;  12929436 ;  12929437 ;  12929438 ;  12929439 ;  12929441 ;  12929442 ;  12929443 ;  12929444 ;  12929446 ;  12929447 ;  12929448 ;  12929450 ;  12929451 ;  12929452 ;  12929453 ;  12929454 ;  12929455 ;  12929456 ;  12929457 ;  12929458 ;  12929459 ;  12929460 ;  12929461 ;  12929462 ;  12929463 ;  12929464 ;  12929465 ;  12929466 ;  12929467 ;  12929468 ;  12929470 ;  12929471 ;  12929472 ;  12929473 ;  12929474 ;  12929475 ;  12929476 ;  12929477 ;  12929478 ;  12929487 ;  12929488 ;  12929489 ;  12929490 ;  12929492 ;  12929493 ;  12929494 ;  12929495 ;  12929496 ;  12929497 ;  12929499 ;  12929500 ;  12929501 ;  12929502 ;  12929503 ;  12929504 ;  12929505 ;  12929506 ;  12929507 ;  12929508 ;  12929509 ;  12929510 ;  12929511 ;  12929512 ;  12929513 ;  12929515 ;  12929517 ;  12929518 ;  12929521 ;  12929522 ;  12929523 ;  12929524 ;  12929525 ;  12929526 ;  12929529 ;  12929531 ;  12929533 ;  12929534 ;  12929535 ;  12929537 ;  12929538 ;  12929539 ;  12929540 ;  12929541 ;  12929542 ;  12929543 ;  12929544 ;  12929545 ;  12929546 ;  12929547 ;  12929548 ;  12929549 ;  12929550 ;  12929551 ;  12929554 ;  12929555 ;  12929556 ;  12929557 ;  12929558 ;  12929559 ;  12929560 ;  12929561 ;  12929563 ;  12929564 ;  12929565 ;  12929566 ;  12929568 ;  12929569 ;  12929570 ;  12929571 ;  12929572 ;  12929575 ;  12929576 ;  12929577 ;  12929579 ;  12929582 ;  12929583 ;  12929584 ;  12929586 ;  12929587 ;  12929588 ;  12929590 ;  12929591 ;  12929592 ;  12929593 ;  12929594 ;  12929595 ;  12929596 ;  12929597 ;  12929599 ;  12929600 ;  12929601 ;  12929603 ;  12929604 ;  12929605 ;  12929606 ;  12929607 ;  12929608 ;  12929609 ;  12929610 ;  12929611 ;  12929612 ;  12929613 ;  12929614 ;  12929615 ;  12929616 ;  12929618 ;  12929619 ;  12929620 ;  12929621 ;  12929622 ;  12929623 ;  12929624 ;  12929625 ;  12929626 ;  12929627 ;  12929628 ;  12929630 ;  12929631 ;  12929632 ;  12929633 ;  12929634 ;  12929635 ;  12929636 ;  12929637 ;  12929638 ;  12929640 ;  12929641 ;  12929642 ;  12929643 ;  12929644 ;  12929645 ;  12929646 ;  12929647 ;  12929648 ;  12929649 ;  12929650 ;  12929651 ;  12929652 ;  12929653 ;  12929654 ;  12929655 ;  12929656 ;  12929657 ;  12929658 ;  12929659 ;  12929660 ;  12929661 ;  12929662 ;  12929663 ;  12929664 ;  12929665 ;  12929666 ;  12929667 ;  12929668 ;  12929670 ;  12929671 ;  12929672 ;  12929673 ;  12929674 ;  12929675 ;  12929676 ;  12929677 ;  12929678 ;  12929680 ;  12929681 ;  12929682 ;  12929683 ;  12929684 ;  12929685 ;  12929686 ;  12929687 ;  12929688 ;  12929689 ;  12929690 ;  12929691 ;  12929692 ;  12929693 ;  12929694 ;  12929695 ;  12929696 ;  12929697 ;  12929698 ;  12929699 ;  12929700 ;  12929701 ;  12929702 ;  12929704 ;  12929705 ;  12929706 ;  12929707 ;  12929709 ;  12929711 ;  12929712 ;  12929713 ;  12929714 ;  12929715 ;  12929716 ;  12929717 ;  12929718 ;  12929719 ;  12929721 ;  12929722 ;  12929723 ;  12929724 ;  12929725 ;  12929726 ;  12929728 ;  12929729 ;  12929731 ;  12929733 ;  12929734 ;  12929735 ;  12929736 ;  12929738 ;  12929739 ;  12929740 ;  12929741 ;  12929742 ;  12929743 ;  12929744 ;  12929745 ;  12929746 ;  12929747 ;  12929748 ;  12929750 ;  12929751 ;  12929752 ;  12929753 ;  12929755 ;  12929756 ;  12929757 ;  12929758 ;  12929759 ;  12929760 ;  12929761 ;  12929763 ;  12929764 ;  12929765 ;  12929766 ;  12929768 ;  12929769 ;  12929770 ;  12929771 ;  12929772 ;  12929774 ;  12929775 ;  12929776 ;  12929777 ;  12929778 ;  12929780 ;  12929781 ;  12929782 ;  12929784 ;  12929785 ;  12929786 ;  12929788 ;  12929790 ;  12929791 ;  12929792 ;  12929793 ;  12929794 ;  12929795 ;  12929796 ;  12929797 ;  12929798 ;  12929799 ;  12929800 ;  12929801 ;  12929802 ;  12929803 ;  12929804 ;  12929805 ;  12929808 ;  12929809 ;  12929810 ;  12929811 ;  12929812 ;  12929813 ;  12929814 ;  12929815 ;  12929816 ;  12929817 ;  12929818 ;  12929819 ;  12929820 ;  12929821 ;  12929822 ;  12929823 ;  12929824 ;  12929825 ;  12929827 ;  12929828 ;  12929830 ;  12929831 ;  12929832 ;  12929833 ;  12929834 ;  12929835 ;  12929837 ;  12929840 ;  12929844 ;  12929845 ;  12929850 ;  12929853 ;  12929854 ;  12929855 ;  12929857 ;  12929858 ;  12929860 ;  12929861 ;  12929862 ;  12929863 ;  12929864 ;  12929865 ;  12929866 ;  12929867 ;  12929869 ;  12929871 ;  12929872 ;  12929873 ;  12929874 ;  12929876 ;  12929877 ;  12929878 ;  12929879 ;  12929882 ;  12929883 ;  12929888 ;  12929891 ;  12929892 ;  12929893 ;  12929894 ;  12929895 ;  12929896 ;  12929897 ;  12929898 ;  12929899 ;  12929900 ;  12929901 ;  12929902 ;  12929903 ;  12929904 ;  12929905 ;  12929906 ;  12929907 ;  12929909 ;  12929912 ;  12929913 ;  12929914 ;  12929915 ;  12929916 ;  12929918 ;  12929919 ;  12929920 ;  12929921 ;  12929923 ;  12929925 ;  12929926 ;  12929927 ;  12929928 ;  12929931 ;  12929932 ;  12929933 ;  12929934 ;  12929935 ;  12929936 ;  12929937 ;  12929938 ;  12929939 ;  12929940 ;  12929942 ;  12929943 ;  12929944 ;  12929945 ;  12929946 ;  12929947 ;  12929948 ;  12929949 ;  12929950 ;  12929951 ;  12929953 ;  12929954 ;  12929955 ;  12929956 ;  12929957 ;  12929958 ;  12929959 ;  12929960 ;  12929961 ;  12929962 ;  12929963 ;  12929964 ;  12929965 ;  12929967 ;  12929968 ;  12929969 ;  12929973 ;  12929976 ;  12929977 ;  12929978 ;  12929980 ;  12929983 ;  12929987 ;  12929989 ;  12929990 ;  12929994 ;  12929995 ;  12929996 ;  12929997 ;  12929998 ;  12929999 ;  12930000 ;  12930001 ;  12930003 ;  12930004 ;  12930008 ;  12930009 ;  12930011 ;  12930012 ;  12930014 ;  12930015 ;  12930016 ;  12930017 ;  12930018 ;  12930019 ;  12930020 ;  12930021 ;  12930022 ;  12930023 ;  12930025 ;  12930026 ;  12930030 ;  12930031 ;  12930036 ;  12930037 ;  12930038 ;  12930039 ;  12930040 ;  12930042 ;  12930043 ;  12930044 ;  12930045 ;  12930046 ;  12930047 ;  12930048 ;  12930049 ;  12930050 ;  12930053 ;  12930054 ;  12930055 ;  12930056 ;  12930057 ;  12930058 ;  12930060 ;  12930062 ;  12930063 ;  12930064 ;  12930066 ;  12930068 ;  12930069 ;  12930070 ;  12930071 ;  12930072 ;  12930074 ;  12930075 ;  12930076 ;  12930077 ;  12930078 ;  12930080 ;  12930081 ;  12930083 ;  12930084 ;  12930085 ;  12930086 ;  12930087 ;  12930088 ;  12930089 ;  12930091 ;  12930092 ;  12930093 ;  12930094 ;  12930095 ;  12930096 ;  12930097 ;  12930098 ;  12930099 ;  12930101 ;  12930102 ;  12930103 ;  12930104 ;  12930105 ;  12930106 ;  12930107 ;  12930108 ;  12930109 ;  12930110 ;  12930111 ;  12930112 ;  12930113 ;  12930115 ;  12930118 ;  12930119 ;  12930120 ;  12930123 ;  12930128 ;  12930129 ;  12930130 ;  12930131 ;  12930132 ;  12930134 ;  12930136 ;  12930138 ;  12930139 ;  12930140 ;  12930141 ;  12930143 ;  12930144 ;  12930145 ;  12930146 ;  12930148 ;  12930149 ;  12930150 ;  12930152 ;  12930153 ;  12930155 ;  12930156 ;  12930157 ;  12930158 ;  12930160 ;  12930161 ;  12930162 ;  12930163 ;  12930164 ;  12930165 ;  12930167 ;  12930168 ;  12930169 ;  12930170 ;  12930172 ;  12930173 ;  12930174 ;  12930175 ;  12930176 ;  12930177 ;  12930178 ;  12930180 ;  12930182 ;  12930183 ;  12930184 ;  12930185 ;  12930187 ;  12930190 ;  12930191 ;  12930192 ;  12930193 ;  12930194 ;  12930195 ;  12930196 ;  12930197 ;  12930199 ;  12930200 ;  12930201 ;  12930202 ;  12930203 ;  12930204 ;  12930205 ;  12930206 ;  12930207 ;  12930208 ;  12930209 ;  12930210 ;  12930211 ;  12930212 ;  12930213 ;  12930214 ;  12930215 ;  12930216 ;  12930217 ;  12930218 ;  12930219 ;  12930220 ;  12930221 ;  12930222 ;  12930223 ;  12930224 ;  12930225 ;  12930226 ;  12930227 ;  12930228 ;  12930229 ;  12930231 ;  12930232 ;  12930235 ;  12930236 ;  12930237 ;  12930238 ;  12930239 ;  12930240 ;  12930241 ;  12930242 ;  12930243 ;  12930244 ;  12930245 ;  12930246 ;  12930247 ;  12930248 ;  12930249 ;  12930250 ;  12930251 ;  12930252 ;  12930253 ;  12930254 ;  12930255 ;  12930256 ;  12930257 ;  12930258 ;  12930260 ;  12930261 ;  12930262 ;  12930263 ;  12930264 ;  12930265 ;  12930266 ;  12930267 ;  12930268 ;  12930269 ;  12930270 ;  12930271 ;  12930272 ;  12930273 ;  12930274 ;  12930275 ;  12930276 ;  12930277 ;  12930278 ;  12930279 ;  12930280 ;  12930281 ;  12930282 ;  12930283 ;  12930284 ;  12930285 ;  12930286 ;  12930287 ;  12930288 ;  12930289 ;  12930290 ;  12930291 ;  12930292 ;  12930293 ;  12930294 ;  12930295 ;  12930296 ;  12930298 ;  12930299 ;  12930300 ;  12930301 ;  12930302 ;  12930303 ;  12930304 ;  12930305 ;  12930306 ;  12930307 ;  12930308 ;  12930309 ;  12930310 ;  12930311 ;  12930312 ;  12930313 ;  12930314 ;  12930315 ;  12930316 ;  12930317 ;  12930318 ;  12930320 ;  12930321 ;  12930322 ;  12930323 ;  12930324 ;  12930325 ;  12930326 ;  12930328 ;  12930329 ;  12930331 ;  12930332 ;  12930333 ;  12930334 ;  12930335 ;  12930336 ;  12930337 ;  12930338 ;  12930339 ;  12930340 ;  12930341 ;  12930342 ;  12930343 ;  12930344 ;  12930345 ;  12930346 ;  12930347 ;  12930348 ;  12930349 ;  12930350 ;  12930352 ;  12930353 ;  12930354 ;  12930355 ;  12930356 ;  12930357 ;  12930358 ;  12930359 ;  12930360 ;  12930361 ;  12930362 ;  12930363 ;  12930364 ;  12930365 ;  12930366 ;  12930367 ;  12930368 ;  12930369 ;  12930370 ;  12930371 ;  12930372 ;  12930373 ;  12930374 ;  12930375 ;  12930376 ;  12930377 ;  12930378 ;  12930379 ;  12930380 ;  12930381 ;  12930382 ;  12930383 ;  12930384 ;  12930385 ;  12930386 ;  12930387 ;  12930388 ;  12930389 ;  12930390 ;  12930391 ;  12930392 ;  12930393 ;  12930394 ;  12930396 ;  12930398 ;  12930399 ;  12930401 ;  12930402 ;  12930403 ;  12930404 ;  12930405 ;  12930406 ;  12930407 ;  12930408 ;  12930409 ;  12930410 ;  12930411 ;  12930412 ;  12930413 ;  12930414 ;  12930416 ;  12930417 ;  12930421 ;  12930422 ;  12930424 ;  12930425 ;  12930426 ;  12930428 ;  12930429 ;  12930430 ;  12930431 ;  12930432 ;  12930433 ;  12930434 ;  12930435 ;  12930436 ;  12930437 ;  12930438 ;  12930439 ;  12930440 ;  12930441 ;  12930442 ;  12930443 ;  12930444 ;  12930445 ;  12930446 ;  12930447 ;  12930448 ;  12930449 ;  12930450 ;  12930451 ;  12930452 ;  12930453 ;  12930454 ;  12930455 ;  12930456 ;  12930458 ;  12930460 ;  12930462 ;  12930463 ;  12930464 ;  12930465 ;  12930466 ;  12930467 ;  12930468 ;  12930469 ;  12930470 ;  12930471 ;  12930472 ;  12930473 ;  12930474 ;  12930475 ;  12930476 ;  12930477 ;  12930478 ;  12930479 ;  12930480 ;  12930481 ;  12930482 ;  12930483 ;  12930485 ;  12930486 ;  12930488 ;  12930489 ;  12930490 ;  12930491 ;  12930492 ;  12930495 ;  12930496 ;  12930497 ;  12930498 ;  12930499 ;  12930501 ;  12930502 ;  12930503 ;  12930506 ;  12930507 ;  12930508 ;  12930509 ;  12930510 ;  12930511 ;  12930512 ;  12930513 ;  12930514 ;  12930515 ;  12930517 ;  12930518 ;  12930519 ;  12930520 ;  12930523 ;  12930524 ;  12930525 ;  12930526 ;  12930527 ;  12930528 ;  12930530 ;  12930531 ;  12930532 ;  12930534 ;  12930537 ;  12930538 ;  12930539 ;  12930540 ;  12930541 ;  12930542 ;  12930543 ;  12930545 ;  12930546 ;  12930547 ;  12930548 ;  12930550 ;  12930552 ;  12930553 ;  12930554 ;  12930555 ;  12930556 ;  12930557 ;  12930561 ;  12930564 ;  12930565 ;  12930566 ;  12930567 ;  12930568 ;  12930569 ;  12930570 ;  12930571 ;  12930572 ;  12930573 ;  12930575 ;  12930576 ;  12930577 ;  12930578 ;  12930579 ;  12930581 ;  12930582 ;  12930583 ;  12930584 ;  12930585 ;  12930588 ;  12930589 ;  12930590 ;  12930592 ;  12930593 ;  12930594 ;  12930596 ;  12930597 ;  12930598 ;  12930599 ;  12930600 ;  12930601 ;  12930602 ;  12930603 ;  12930604 ;  12930605 ;  12930606 ;  12930607 ;  12930610 ;  12930611 ;  12930613 ;  12930614 ;  12930615 ;  12930616 ;  12930617 ;  12930618 ;  12930620 ;  12930621 ;  12930622 ;  12930623 ;  12930624 ;  12930625 ;  12930626 ;  12930627 ;  12930628 ;  12930632 ;  12930635 ;  12930637 ;  12930638 ;  12930639 ;  12930640 ;  12930642 ;  12930644 ;  12930647 ;  12930648 ;  12930649 ;  12930653 ;  12930657 ;  12930658 ;  12930659 ;  12930660 ;  12930661 ;  12930668 ;  12930669 ;  12930670 ;  12930671 ;  12930672 ;  12930673 ;  12930674 ;  12930676 ;  12930678 ;  12930682 ;  12930683 ;  12930689 ;  12930691 ;  12930692 ;  12930694 ;  12930696 ;  12930697 ;  12930704 ;  12930711 ;  12930712 ;  12930714 ;  12930720 ;  12930721 ;  12930722 ;  12930723 ;  12930725 ;  12930726 ;  12930727 ;  12930728 ;  12930732 ;  12930749 ;  12930758 ;  12930761 ;  12930762 ;  12930766 ;  12930767 ;  12930768 ;  12930769 ;  12930770 ;  12930771 ;  12930772 ;  12930773 ;  12930774 ;  12930775 ;  12930776 ;  12930778 ;  12930779 ;  12930780 ;  12930781 ;  12930782 ;  12930783 ;  12930784 ;  12930785 ;  12930786 ;  12930787 ;  12930788 ;  12930790 ;  12930791 ;  12930792 ;  12930793 ;  12930794 ;  12930795 ;  12930796 ;  12930797 ;  12930798 ;  12930799 ;  12930801 ;  12930803 ;  12930804 ;  12930805 ;  12930806 ;  12930807 ;  12930808 ;  12930809 ;  12930810 ;  12930811 ;  12930813 ;  12930815 ;  12930816 ;  12930817 ;  12930818 ;  12930821 ;  12930822 ;  12930823 ;  12930824 ;  12930825 ;  12930826 ;  12930827 ;  12930828 ;  12930829 ;  12930830 ;  12930831 ;  12930832 ;  12930834 ;  12930841 ;  12930845 ;  12930847 ;  12930848 ;  12930850 ;  12930851 ;  12930852 ;  12930853 ;  12930854 ;  12930857 ;  12930859 ;  12930862 ;  12930866 ;  12930868 ;  12930869 ;  12930871 ;  12930872 ;  12930873 ;  12930875 ;  12930878 ;  12930879 ;  12930880 ;  12930881 ;  12930883 ;  12930886 ;  12930887 ;  12930890 ;  12930891 ;  12930892 ;  12930893 ;  12930894 ;  12930895 ;  12930896 ;  12930898 ;  12930899 ;  12930900 ;  12930902 ;  12930904 ;  12930908 ;  12930909 ;  12930910 ;  12930911 ;  12930912 ;  12930913 ;  12930914 ;  12930915 ;  12930916 ;  12930917 ;  12930918 ;  12930919 ;  12930920 ;  12930921 ;  12930922 ;  12930923 ;  12930924 ;  12930926 ;  12930927 ;  12930928 ;  12930929 ;  12930930 ;  12930931 ;  12930932 ;  12930933 ;  12930935 ;  12930936 ;  12930937 ;  12930938 ;  12930940 ;  12930941 ;  12930942 ;  12930943 ;  12930944 ;  12930945 ;  12930946 ;  12930947 ;  12930948 ;  12930949 ;  12930951 ;  12930952 ;  12930953 ;  12930955 ;  12930956 ;  12930958 ;  12930960 ;  12930961 ;  12930963 ;  12930965 ;  12930968 ;  12930969 ;  12930971 ;  12930972 ;  12930974 ;  12930975 ;  12930976 ;  12930977 ;  12930978 ;  12930979 ;  12930980 ;  12930981 ;  12930982 ;  12930983 ;  12930985 ;  12930986 ;  12930987 ;  12930989 ;  12930990 ;  12930991 ;  12930992 ;  12930994 ;  12930995 ;  12930996 ;  12930997 ;  12930998 ;  12931000 ;  12931001 ;  12931002 ;  12931003 ;  12931004 ;  12931005 ;  12931006 ;  12931007 ;  12931008 ;  12931009 ;  12931011 ;  12931012 ;  12931013 ;  12931014 ;  12931015 ;  12931016 ;  12931017 ;  12931018 ;  12931019 ;  12931020 ;  12931021 ;  12931022 ;  12931023 ;  12931024 ;  12931026 ;  12931027 ;  12931028 ;  12931030 ;  12931031 ;  12931032 ;  12931033 ;  12931034 ;  12931035 ;  12931036 ;  12931037 ;  12931038 ;  12931039 ;  12931040 ;  12931041 ;  12931042 ;  12931043 ;  12931044 ;  12931045 ;  12931046 ;  12931047 ;  12931048 ;  12931049 ;  12931050 ;  12931051 ;  12931052 ;  12931053 ;  12931054 ;  12931055 ;  12931056 ;  12931057 ;  12931058 ;  12931059 ;  12931060 ;  12931061 ;  12931062 ;  12931063 ;  12931064 ;  12931065 ;  12931066 ;  12931067 ;  12931068 ;  12931069 ;  12931070 ;  12931071 ;  12931072 ;  12931073 ;  12931074 ;  12931075 ;  12931076 ;  12931077 ;  12931078 ;  12931079 ;  12931080 ;  12931081 ;  12931082 ;  12931083 ;  12931084 ;  12931085 ;  12931086 ;  12931087 ;  12931089 ;  12931090 ;  12931091 ;  12931092 ;  12931093 ;  12931094 ;  12931095 ;  12931096 ;  12931098 ;  12931099 ;  12931100 ;  12931101 ;  12931103 ;  12931104 ;  12931105 ;  12931106 ;  12931107 ;  12931108 ;  12931109 ;  12931110 ;  12931111 ;  12931112 ;  12931113 ;  12931114 ;  12931115 ;  12931116 ;  12931117 ;  12931119 ;  12931120 ;  12931121 ;  12931122 ;  12931124 ;  12931125 ;  12931126 ;  12931127 ;  12931128 ;  12931129 ;  12931130 ;  12931131 ;  12931132 ;  12931133 ;  12931134 ;  12931135 ;  12931136 ;  12931137 ;  12931138 ;  12931139 ;  12931140 ;  12931141 ;  12931142 ;  12931143 ;  12931144 ;  12931145 ;  12931146 ;  12931147 ;  12931148 ;  12931149 ;  12931150 ;  12931151 ;  12931152 ;  12931153 ;  12931154 ;  12931155 ;  12931156 ;  12931157 ;  12931158 ;  12931159 ;  12931160 ;  12931161 ;  12931162 ;  12931163 ;  12931164 ;  12931165 ;  12931166 ;  12931167 ;  12931168 ;  12931169 ;  12931170 ;  12931171 ;  12931172 ;  12931173 ;  12931174 ;  12931175 ;  12931176 ;  12931177 ;  12931178 ;  12931179 ;  12931180 ;  12931181 ;  12931182 ;  12931183 ;  12931184 ;  12931185 ;  12931186 ;  12931187 ;  12931188 ;  12931189 ;  12931190 ;  12931191 ;  12931192 ;  12931193 ;  12931194 ;  12931195 ;  12931196 ;  12931197 ;  12931198 ;  12931199 ;  12931200 ;  12931201 ;  12931202 ;  12931203 ;  12931204 ;  12931205 ;  12931206 ;  12931207 ;  12931208 ;  12931210 ;  12931211 ;  12931212 ;  12931213 ;  12931214 ;  12931215 ;  12931216 ;  12931217 ;  12931218 ;  12931219 ;  12931220 ;  12931221 ;  12931222 ;  12931223 ;  12931224 ;  12931225 ;  12931226 ;  12931227 ;  12931228 ;  12931229 ;  12931230 ;  12931231 ;  12931232 ;  12931233 ;  12931234 ;  12931235 ;  12931236 ;  12931237 ;  12931238 ;  12931239 ;  12931240 ;  12931241 ;  12931242 ;  12931243 ;  12931244 ;  12931245 ;  12931246 ;  12931248 ;  12931249 ;  12931250 ;  12931251 ;  12931252 ;  12931253 ;  12931254 ;  12931255 ;  12931256 ;  12931257 ;  12931258 ;  12931259 ;  12931260 ;  12931261 ;  12931262 ;  12931263 ;  12931264 ;  12931265 ;  12931266 ;  12931267 ;  12931268 ;  12931269 ;  12931270 ;  12931271 ;  12931272 ;  12931273 ;  12931274 ;  12931275 ;  12931276 ;  12931277 ;  12931278 ;  12931279 ;  12931280 ;  12931281 ;  12931282 ;  12931283 ;  12931284 ;  12931285 ;  12931286 ;  12931287 ;  12931288 ;  12931289 ;  12931290 ;  12931291 ;  12931292 ;  12931293 ;  12931294 ;  12931295 ;  12931296 ;  12931297 ;  12931298 ;  12931299 ;  12931300 ;  12931301 ;  12931302 ;  12931303 ;  12931304 ;  12931305 ;  12931306 ;  12931307 ;  12931308 ;  12931309 ;  12931310 ;  12931311 ;  12931312 ;  12931313 ;  12931314 ;  12931315 ;  12931316 ;  12931317 ;  12931318 ;  12931319 ;  12931320 ;  12931322 ;  12931323 ;  12931324 ;  12931325 ;  12931326 ;  12931327 ;  12931328 ;  12931329 ;  12931330 ;  12931331 ;  12931332 ;  12931333 ;  12931334 ;  12931335 ;  12931336 ;  12931337 ;  12931338 ;  12931339 ;  12931340 ;  12931341 ;  12931342 ;  12931343 ;  12931344 ;  12931345 ;  12931346 ;  12931347 ;  12931348 ;  12931349 ;  12931350 ;  12931351 ;  12931352 ;  12931353 ;  12931354 ;  12931355 ;  12931356 ;  12931357 ;  12931358 ;  12931359 ;  12931360 ;  12931361 ;  12931362 ;  12931363 ;  12931364 ;  12931365 ;  12931366 ;  12931367 ;  12931368 ;  12931369 ;  12931370 ;  12931371 ;  12931372 ;  12931373 ;  12931374 ;  12931375 ;  12931376 ;  12931377 ;  12931379 ;  12931380 ;  12931381 ;  12931382 ;  12931383 ;  12931384 ;  12931385 ;  12931386 ;  12931387 ;  12931388 ;  12931389 ;  12931390 ;  12931391 ;  12931392 ;  12931393 ;  12931394 ;  12931395 ;  12931399 ;  12931400 ;  12931401 ;  12931402 ;  12931403 ;  12931404 ;  12931405 ;  12931406 ;  12931407 ;  12931408 ;  12931410 ;  12931411 ;  12931412 ;  12931413 ;  12931414 ;  12931415 ;  12931416 ;  12931418 ;  12931419 ;  12931420 ;  12931421 ;  12931423 ;  12931424 ;  12931425 ;  12931426 ;  12931427 ;  12931428 ;  12931429 ;  12931430 ;  12931431 ;  12931432 ;  12931433 ;  12931434 ;  12931435 ;  12931436 ;  12931437 ;  12931438 ;  12931439 ;  12931441 ;  12931442 ;  12931443 ;  12931445 ;  12931446 ;  12931447 ;  12931448 ;  12931449 ;  12931450 ;  12931451 ;  12931453 ;  12931455 ;  12931456 ;  12931457 ;  12931458 ;  12931459 ;  12931460 ;  12931461 ;  12931462 ;  12931463 ;  12931464 ;  12931465 ;  12931466 ;  12931467 ;  12931468 ;  12931470 ;  12931471 ;  12931472 ;  12931473 ;  12931474 ;  12931475 ;  12931476 ;  12931477 ;  12931478 ;  12931479 ;  12931480 ;  12931481 ;  12931482 ;  12931483 ;  12931484 ;  12931485 ;  12931486 ;  12931487 ;  12931488 ;  12931489 ;  12931490 ;  12931491 ;  12931492 ;  12931495 ;  12931496 ;  12931498 ;  12931499 ;  12931500 ;  12931501 ;  12931502 ;  12931503 ;  12931504 ;  12931505 ;  12931506 ;  12931507 ;  12931508 ;  12931509 ;  12931510 ;  12931511 ;  12931512 ;  12931513 ;  12931514 ;  12931515 ;  12931516 ;  12931517 ;  12931518 ;  12931520 ;  12931521 ;  12931522 ;  12931523 ;  12931524 ;  12931525 ;  12931526 ;  12931527 ;  12931528 ;  12931529 ;  12931530 ;  12931532 ;  12931533 ;  12931534 ;  12931535 ;  12931536 ;  12931537 ;  12931538 ;  12931539 ;  12931540 ;  12931541 ;  12931542 ;  12931543 ;  12931544 ;  12931545 ;  12931546 ;  12931547 ;  12931548 ;  12931549 ;  12931550 ;  12931551 ;  12931552 ;  12931553 ;  12931555 ;  12931558 ;  12931559 ;  12931560 ;  12931561 ;  12931562 ;  12931563 ;  12931564 ;  12931565 ;  12931566 ;  12931567 ;  12931568 ;  12931569 ;  12931570 ;  12931571 ;  12931572 ;  12931573 ;  12931574 ;  12931575 ;  12931576 ;  12931577 ;  12931578 ;  12931579 ;  12931580 ;  12931581 ;  12931582 ;  12931583 ;  12931584 ;  12931585 ;  12931586 ;  12931587 ;  12931588 ;  12931589 ;  12931590 ;  12931591 ;  12931592 ;  12931593 ;  12931594 ;  12931595 ;  12931596 ;  12931597 ;  12931598 ;  12931599 ;  12931600 ;  12931601 ;  12931602 ;  12931603 ;  12931604 ;  12931605 ;  12931606 ;  12931607 ;  12931608 ;  12931609 ;  12931610 ;  12931611 ;  12931612 ;  12931613 ;  12931614 ;  12931615 ;  12931616 ;  12931617 ;  12931618 ;  12931619 ;  12931620 ;  12931621 ;  12931622 ;  12931623 ;  12931624 ;  12931625 ;  12931626 ;  12931627 ;  12931628 ;  12931629 ;  12931630 ;  12931631 ;  12931632 ;  12931633 ;  12931634 ;  12931635 ;  12931636 ;  12931637 ;  12931638 ;  12931639 ;  12931640 ;  12931641 ;  12931642 ;  12931643 ;  12931644 ;  12931645 ;  12931646 ;  12931647 ;  12931648 ;  12931649 ;  12931650 ;  12931651 ;  12931652 ;  12931653 ;  12931654 ;  12931655 ;  12931656 ;  12931657 ;  12931658 ;  12931659 ;  12931660 ;  12931661 ;  12931662 ;  12931663 ;  12931664 ;  12931665 ;  12931666 ;  12931667 ;  12931668 ;  12931669 ;  12931670 ;  12931671 ;  12931672 ;  12931673 ;  12931676 ;  12931677 ;  12931678 ;  12931681 ;  12931682 ;  12931683 ;  12931684 ;  12931685 ;  12931686 ;  12931687 ;  12931688 ;  12931689 ;  12931690 ;  12931691 ;  12931692 ;  12931693 ;  12931694 ;  12931695 ;  12931696 ;  12931697 ;  12931699 ;  12931701 ;  12931703 ;  12931705 ;  12931707 ;  12931710 ;  12931712 ;  12931713 ;  12931714 ;  12931715 ;  12931717 ;  12931718 ;  12931719 ;  12931720 ;  12931721 ;  12931722 ;  12931723 ;  12931724 ;  12931725 ;  12931726 ;  12931727 ;  12931728 ;  12931729 ;  12931730 ;  12931731 ;  12931735 ;  12931736 ;  12931737 ;  12931738 ;  12931739 ;  12931740 ;  12931744 ;  12931745 ;  12931746 ;  12931747 ;  12931748 ;  12931750 ;  12931751 ;  12931752 ;  12931753 ;  12931754 ;  12931755 ;  12931757 ;  12931758 ;  12931760 ;  12931761 ;  12931763 ;  12931764 ;  12931768 ;  12931770 ;  12931771 ;  12931772 ;  12931773 ;  12931774 ;  12931775 ;  12931776 ;  12931777 ;  12931778 ;  12931779 ;  12931780 ;  12931781 ;  12931782 ;  12931785 ;  12931786 ;  12931787 ;  12931788 ;  12931789 ;  12931790 ;  12931791 ;  12931792 ;  12931793 ;  12931794 ;  12931795 ;  12931797 ;  12931798 ;  12931799 ;  12931800 ;  12931801 ;  12931802 ;  12931803 ;  12931804 ;  12931805 ;  12931806 ;  12931807 ;  12931808 ;  12931809 ;  12931811 ;  12931814 ;  12931815 ;  12931816 ;  12931817 ;  12931818 ;  12931819 ;  12931821 ;  12931822 ;  12931823 ;  12931824 ;  12931825 ;  12931826 ;  12931827 ;  12931828 ;  12931829 ;  12931830 ;  12931831 ;  12931832 ;  12931833 ;  12931834 ;  12931835 ;  12931836 ;  12931837 ;  12931838 ;  12931839 ;  12931840 ;  12931841 ;  12931842 ;  12931843 ;  12931844 ;  12931845 ;  12931846 ;  12931847 ;  12931848 ;  12931849 ;  12931850 ;  12931851 ;  12931853 ;  12931854 ;  12931855 ;  12931856 ;  12931857 ;  12931858 ;  12931859 ;  12931860 ;  12931861 ;  12931862 ;  12931863 ;  12931864 ;  12931865 ;  12931866 ;  12931867 ;  12931869 ;  12931870 ;  12931871 ;  12931872 ;  12931873 ;  12931875 ;  12931877 ;  12931878 ;  12931879 ;  12931880 ;  12931881 ;  12931882 ;  12931883 ;  12931884 ;  12931885 ;  12931886 ;  12931889 ;  12931891 ;  12931892 ;  12931894 ;  12931895 ;  12931896 ;  12931898 ;  12931901 ;  12931902 ;  12931903 ;  12931905 ;  12931906 ;  12931907 ;  12931909 ;  12931910 ;  12931911 ;  12931913 ;  12931914 ;  12931915 ;  12931916 ;  12931917 ;  12931918 ;  12931919 ;  12931920 ;  12931921 ;  12931922 ;  12931923 ;  12931924 ;  12931926 ;  12931927 ;  12931928 ;  12931930 ;  12931931 ;  12931932 ;  12931933 ;  12931934 ;  12931935 ;  12931937 ;  12931938 ;  12931940 ;  12931941 ;  12931942 ;  12931943 ;  12931944 ;  12931945 ;  12931946 ;  12931947 ;  12931948 ;  12931949 ;  12931950 ;  12931953 ;  12931955 ;  12931956 ;  12931957 ;  12931958 ;  12931960 ;  12931961 ;  12931962 ;  12931963 ;  12931965 ;  12931966 ;  12931967 ;  12931968 ;  12931969 ;  12931970 ;  12931972 ;  12931973 ;  12931974 ;  12931975 ;  12931976 ;  12931977 ;  12931978 ;  12931979 ;  12931980 ;  12931981 ;  12931982 ;  12931983 ;  12931985 ;  12931986 ;  12931987 ;  12931990 ;  12931991 ;  12931992 ;  12931993 ;  12931994 ;  12931995 ;  12931996 ;  12931997 ;  12931998 ;  12931999 ;  12932000 ;  12932001 ;  12932002 ;  12932003 ;  12932004 ;  12932005 ;  12932006 ;  12932007 ;  12932009 ;  12932010 ;  12932011 ;  12932012 ;  12932015 ;  12932016 ;  12932018 ;  12932020 ;  12932021 ;  12932022 ;  12932023 ;  12932024 ;  12932025 ;  12932026 ;  12932028 ;  12932029 ;  12932030 ;  12932031 ;  12932032 ;  12932033 ;  12932034 ;  12932035 ;  12932036 ;  12932038 ;  12932039 ;  12932040 ;  12932041 ;  12932043 ;  12932044 ;  12932045 ;  12932046 ;  12932047 ;  12932049 ;  12932050 ;  12932051 ;  12932052 ;  12932054 ;  12932055 ;  12932056 ;  12932057 ;  12932058 ;  12932060 ;  12932062 ;  12932063 ;  12932064 ;  12932067 ;  12932068 ;  12932069 ;  12932070 ;  12932072 ;  12932073 ;  12932074 ;  12932075 ;  12932076 ;  12932077 ;  12932078 ;  12932079 ;  12932080 ;  12932081 ;  12932082 ;  12932083 ;  12932084 ;  12932085 ;  12932086 ;  12932088 ;  12932089 ;  12932090 ;  12932091 ;  12932092 ;  12932094 ;  12932095 ;  12932096 ;  12932097 ;  12932098 ;  12932099 ;  12932100 ;  12932102 ;  12932103 ;  12932105 ;  12932106 ;  12932107 ;  12932108 ;  12932111 ;  12932112 ;  12932113 ;  12932114 ;  12932115 ;  12932116 ;  12932118 ;  12932119 ;  12932121 ;  12932122 ;  12932123 ;  12932124 ;  12932125 ;  12932126 ;  12932127 ;  12932128 ;  12932129 ;  12932130 ;  12932131 ;  12932132 ;  12932133 ;  12932135 ;  12932137 ;  12932138 ;  12932139 ;  12932140 ;  12932141 ;  12932142 ;  12932143 ;  12932144 ;  12932145 ;  12932146 ;  12932147 ;  12932148 ;  12932149 ;  12932150 ;  12932151 ;  12932152 ;  12932153 ;  12932154 ;  12932155 ;  12932156 ;  12932157 ;  12932158 ;  12932159 ;  12932161 ;  12932162 ;  12932163 ;  12932164 ;  12932165 ;  12932166 ;  12932167 ;  12932168 ;  12932169 ;  12932170 ;  12932172 ;  12932173 ;  12932174 ;  12932175 ;  12932176 ;  12932177 ;  12932178 ;  12932179 ;  12932180 ;  12932181 ;  12932182 ;  12932183 ;  12932184 ;  12932185 ;  12932186 ;  12932189 ;  12932191 ;  12932192 ;  12932193 ;  12932194 ;  12932196 ;  12932197 ;  12932198 ;  12932200 ;  12932201 ;  12932203 ;  12932206 ;  12932207 ;  12932208 ;  12932209 ;  12932210 ;  12932211 ;  12932212 ;  12932213 ;  12932214 ;  12932215 ;  12932216 ;  12932217 ;  12932218 ;  12932219 ;  12932220 ;  12932221 ;  12932222 ;  12932223 ;  12932225 ;  12932227 ;  12932228 ;  12932229 ;  12932230 ;  12932231 ;  12932232 ;  12932234 ;  12932235 ;  12932236 ;  12932237 ;  12932238 ;  12932239 ;  12932241 ;  12932243 ;  12932244 ;  12932245 ;  12932246 ;  12932248 ;  12932249 ;  12932250 ;  12932251 ;  12932252 ;  12932253 ;  12932255 ;  12932257 ;  12932258 ;  12932259 ;  12932260 ;  12932261 ;  12932262 ;  12932263 ;  12932264 ;  12932265 ;  12932268 ;  12932271 ;  12932272 ;  12932273 ;  12932274 ;  12932275 ;  12932276 ;  12932278 ;  12932280 ;  12932281 ;  12932282 ;  12932283 ;  12932284 ;  12932285 ;  12932286 ;  12932287 ;  12932288 ;  12932289 ;  12932290 ;  12932291 ;  12932292 ;  12932293 ;  12932294 ;  12932295 ;  12932296 ;  12932297 ;  12932298 ;  12932299 ;  12932300 ;  12932301 ;  12932302 ;  12932303 ;  12932304 ;  12932305 ;  12932307 ;  12932308 ;  12932309 ;  12932311 ;  12932312 ;  12932313 ;  12932314 ;  12932315 ;  12932316 ;  12932317 ;  12932318 ;  12932319 ;  12932320 ;  12932321 ;  12932323 ;  12932325 ;  12932326 ;  12932327 ;  12932328 ;  12932329 ;  12932330 ;  12932331 ;  12932332 ;  12932333 ;  12932334 ;  12932335 ;  12932336 ;  12932337 ;  12932338 ;  12932339 ;  12932340 ;  12932341 ;  12932342 ;  12932343 ;  12932344 ;  12932345 ;  12932346 ;  12932347 ;  12932348 ;  12932349 ;  12932350 ;  12932351 ;  12932352 ;  12932353 ;  12932354 ;  12932355 ;  12932356 ;  12932357 ;  12932358 ;  12932359 ;  12932360 ;  12932361 ;  12932362 ;  12932363 ;  12932364 ;  12932365 ;  12932366 ;  12932367 ;  12932368 ;  12932369 ;  12932370 ;  12932371 ;  12932372 ;  12932373 ;  12932374 ;  12932376 ;  12932377 ;  12932378 ;  12932379 ;  12932380 ;  12932381 ;  12932382 ;  12932383 ;  12932384 ;  12932385 ;  12932386 ;  12932387 ;  12932388 ;  12932389 ;  12932390 ;  12932391 ;  12932392 ;  12932393 ;  12932394 ;  12932395 ;  12932396 ;  12932397 ;  12932398 ;  12932399 ;  12932400 ;  12932401 ;  12932402 ;  12932403 ;  12932404 ;  12932405 ;  12932406 ;  12932407 ;  12932408 ;  12932409 ;  12932410 ;  12932411 ;  12932412 ;  12932413 ;  12932414 ;  12932415 ;  12932416 ;  12932417 ;  12932418 ;  12932420 ;  12932421 ;  12932422 ;  12932423 ;  12932424 ;  12932425 ;  12932426 ;  12932427 ;  12932428 ;  12932429 ;  12932430 ;  12932431 ;  12932432 ;  12932433 ;  12932434 ;  12932435 ;  12932436 ;  12932437 ;  12932438 ;  12932439 ;  12932440 ;  12932441 ;  12932442 ;  12932443 ;  12932444 ;  12932448 ;  12932449 ;  12932450 ;  12932451 ;  12932452 ;  12932454 ;  12932459 ;  12932460 ;  12932461 ;  12932462 ;  12932463 ;  12932464 ;  12932465 ;  12932466 ;  12932467 ;  12932468 ;  12932469 ;  12932470 ;  12932471 ;  12932472 ;  12932473 ;  12932474 ;  12932475 ;  12932476 ;  12932477 ;  12932478 ;  12932479 ;  12932480 ;  12932481 ;  12932482 ;  12932483 ;  12932484 ;  12932485 ;  12932486 ;  12932487 ;  12932488 ;  12932489 ;  12932490 ;  12932491 ;  12932492 ;  12932493 ;  12932494 ;  12932495 ;  12932496 ;  12932498 ;  12932499 ;  12932500 ;  12932501 ;  12932502 ;  12932505 ;  12932506 ;  12932508 ;  12932509 ;  12932510 ;  12932511 ;  12932512 ;  12932513 ;  12932514 ;  12932515 ;  12932516 ;  12932517 ;  12932518 ;  12932519 ;  12932520 ;  12932521 ;  12932522 ;  12932523 ;  12932524 ;  12932525 ;  12932526 ;  12932527 ;  12932528 ;  12932529 ;  12932530 ;  12932531 ;  12932532 ;  12932533 ;  12932534 ;  12932535 ;  12932536 ;  12932537 ;  12932538 ;  12932539 ;  12932540 ;  12932541 ;  12932542 ;  12932543 ;  12932544 ;  12932546 ;  12932547 ;  12932548 ;  12932549 ;  12932550 ;  12932551 ;  12932552 ;  12932553 ;  12932555 ;  12932556 ;  12932557 ;  12932558 ;  12932559 ;  12932560 ;  12932562 ;  12932563 ;  12932564 ;  12932567 ;  12932568 ;  12932569 ;  12932570 ;  12932571 ;  12932572 ;  12932573 ;  12932574 ;  12932575 ;  12932576 ;  12932577 ;  12932579 ;  12932580 ;  12932581 ;  12932583 ;  12932584 ;  12932585 ;  12932586 ;  12932587 ;  12932588 ;  12932589 ;  12932590 ;  12932591 ;  12932592 ;  12932594 ;  12932596 ;  12932597 ;  12932598 ;  12932599 ;  12932600 ;  12932601 ;  12932602 ;  12932603 ;  12932604 ;  12932605 ;  12932606 ;  12932607 ;  12932610 ;  12932612 ;  12932613 ;  12932614 ;  12932615 ;  12932616 ;  12932617 ;  12932618 ;  12932619 ;  12932620 ;  12932621 ;  12932622 ;  12932623 ;  12932624 ;  12932625 ;  12932626 ;  12932627 ;  12932628 ;  12932629 ;  12932630 ;  12932631 ;  12932632 ;  12932633 ;  12932634 ;  12932635 ;  12932636 ;  12932637 ;  12932638 ;  12932639 ;  12932640 ;  12932641 ;  12932642 ;  12932643 ;  12932644 ;  12932645 ;  12932646 ;  12932647 ;  1293 2648 ;  12932649 ;  12932650 ;  12932651 ;  12932652 ;  12932653 ;  12932654 ;  12932655 ;  12932656 ;  12932657 ;  12932658 ;  12932659 ;  12932660 ;  12932661 ;  12932662 ;  12932663 ;  12932664 ;  12932665 ;  12932666 ;  12932667 ;  12932669 ;  12932670 ;  12932671 ;  12932672 ;  12932673 ;  12932674 ;  12932675 ;  12932676 ;  12932677 ;  12932678 ;  12932679 ;  12932680 ;  12932681 ;  12932682 ;  12932683 ;  12932684 ;  12932685 ;  12932686 ;  12932687 ;  12932688 ;  12932689 ;  12932690 ;  12932691 ;  12932692 ;  12932693 ;  12932694 ;  12932695 ;  12932696 ;  12932697 ;  12932698 ;  12932699 ;  12932700 ;  12932701 ;  12932702 ;  12932703 ;  12932704 ;  12932706 ;  12932707 ;  12932708 ;  12932709 ;  12932710 ;  12932711 ;  12932712 ;  12932713 ;  12932714 ;  12932715 ;  12932716 ;  12932717 ;  12932718",
      "recall_initiation_date": "20130308",
      "postal_code": "60045-2579",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}