  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Simpsonville",
      "address_1": "132 Citizens Blvd",
      "reason_for_recall": "During an FDA inspection, the firm was notified of Quality System deficiencies, which included lack of procedures for design control and customized modifications that were either not reviewed or received proper clearance from FDA.  There was no assurance that all systems in the field were working properly and operating at a minimum of 0.2ppm of ozone.",
      "address_2": "",
      "product_quantity": "44 units",
      "code_info": "Model WOZ -250-PE, Serial #s: 1031, 1032, 1033, 1035, 1036, 1038, 1039, 1042, 1043, 1044, 1045, 1046, 1047, 1048, 1049, 1050, 1051, 1053, 1054, 1055, 1056, 1057, 1058, 1059, 1060, 1061, 1062, 1063, 1065, 1066, 1067, 1068, 1069, 1070, 1071, 1072, 1073, 1074, 1075, 1076 & 1077.  Model WOZ-180-SS, Serial #s: 1034, 1037, 1040 & 1064.  Model WOZ-240-SS, Serial #s: 1041 & 1052.",
      "center_classification_date": "20120803",
      "distribution_pattern": "Product was shipped to the following states: CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MD, ME, NH, NJ, NM, OH, PA, RI, TX and VA.",
      "state": "KY",
      "product_description": "TANGO3 Water Storage Tank with Ozone Disinfection System, Tango3, LLC, 132 Citizens Blvd., Simpsonville, KY 40067. Models are WOZ-250-PE, WOZ-240-SS, WOZ-180-SS.  The TANGO3 Water Storage Tank with Ozone Disinfection System is intended to be used for disinfection of the water distribution system of a dialysis facility.",
      "report_date": "20120815",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Tango3 LLC",
      "recall_number": "Z-2138-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62514",
      "termination_date": "20130301",
      "more_code_info": "",
      "recall_initiation_date": "20120326",
      "postal_code": "40067-6547",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}