  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Allentown",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88198",
      "recalling_firm": "B. Braun Medical, Inc.",
      "address_1": "901 Marcon Blvd",
      "address_2": "N/A",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Affected products were distributed to the following US states:  CA, MI, OR, and UT.",
      "recall_number": "Z-2137-2021",
      "product_description": "Infusomat Space Volumetric Pump Administration Set, Product Code 490100",
      "product_quantity": "36,144 units",
      "reason_for_recall": "There is a potential for the Anti-free flow clip of the administration set to be inverted which may cause unintended free flow if the user fails to follow on screen prompt to close the roller clamp prior to set removal from the pump.",
      "recall_initiation_date": "20210625",
      "center_classification_date": "20210723",
      "termination_date": "20231016",
      "report_date": "20210804",
      "code_info": "Lot Number: 0061761278"
    }
  ]
}