  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Innsbruck",
      "state": "N/A",
      "country": "Austria",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "81683",
      "recalling_firm": "MED-EL Elektromedizinische Gereate, Gmbh",
      "address_1": "Furstenweg 77a",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide in the states of  California and Illinois. Countries of ARGENTINA, AUSTRALIA, AUSTRIA, BRAZIL, BULGARIA, CHINA, FRANCE, GEORGIA, KOREA, REPUBLIC OF, POLAND, RUSSIAN FEDERATION, SPAIN, SWEDEN.",
      "recall_number": "Z-2137-2019",
      "product_description": "MED EL Cochlear Implant System, SONATATIi100, sold under the following implant types:      a. SONATAti100 FLEXsoft;      b. SONATAti100 H Standard;      c. SONATAti100 Standard    Product Usage: The SONATATIi100 Implant is an implantable part of the MED EL Cochlear Implant System and can only be used together with compatible MED EL external components. The MED EL Cochlear Implant System is intended to evoke auditory sensations via electrical stimulation of the auditory pathways for severely to profoundly hearing impaired individuals who obtain little or no benefit from acoustic amplification in the best aided condition.",
      "product_quantity": "10 devices",
      "reason_for_recall": "Devices were distributed despite the Helium-fine leak test results not being within the cochlear implant standards specification.",
      "recall_initiation_date": "20181016",
      "center_classification_date": "20190802",
      "termination_date": "20200522",
      "report_date": "20190814",
      "code_info": "a. SONATAti100 FLEXsoft, Serial Number 150489;      b. SONATAti100 H Standard, Serial Numbers 146714, 164324;      c. SONATAti100 Standard, Serial Numbers 145179, 217623, 144443, 145228, 158919, 147748, 223013."
    }
  ]
}