  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68406",
      "recalling_firm": "Carestream Health Inc.",
      "address_1": "150 Verona St",
      "address_2": "N/A",
      "postal_code": "14608-1733",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of   Canada, China, Italy, Germany, Greece, Kuwait, Spain, South Africa, and United Arab Emirates.",
      "recall_number": "Z-2137-2014",
      "product_description": "KODAK DirectView DR 9000 System,  Catalog Numbers 1966688 (US and Canada) and 8161937 (outside the US and Canada) ---  A permanently installed diagnostic x-ray system composed of 3 main components: an operator console, U-Arm assembly with x-ray tube including a collimator, and a stationary generator.",
      "product_quantity": "Domestic: 100 units; Foreign: 46 units",
      "reason_for_recall": "A control system failure may cause the U-Arm to move downward unexpectedly when the Radiology Technologist is positioning the equipment using the \"Up\" or \"Down\" control buttons. Downward movement occurs at three times the normal rate of speed and will continue until the device control button is released, the Emergency Stop button is activated or the equipment Bump Sensor contacts an object.",
      "recall_initiation_date": "20140530",
      "center_classification_date": "20140801",
      "termination_date": "20141217",
      "report_date": "20140813",
      "code_info": "Service Code 1534"
    }
  ]
}