  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Wauwatosa",
      "state": "WI",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68596",
      "recalling_firm": "Enhancement Medical, LLC",
      "address_1": "10201 Innovation Drive",
      "address_2": "Suite 450",
      "postal_code": "53226",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Visit",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide including Puerto Rico) and Internationally to Canada.",
      "recall_number": "Z-2135-2014",
      "product_description": "Enhancement Medical expression.  model # EXP-150.  Injectable Gel 1.5cc Syringe with hyaluronic acid concentration of 20 mg/ml.     Product Usage:  Expression is labelled as an Intranasal Splint, intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. However, Expression is currently being used off-label as a subcutaneously injected dermal filler.",
      "product_quantity": "17,875",
      "reason_for_recall": "Manufacturer made a change in the production process that resulted in a change in final gel weight.   RECALL EXPANDED 7/8/2014   Firm expanded their recall to include all lots of product.",
      "recall_initiation_date": "20130807",
      "center_classification_date": "20140808",
      "termination_date": "20170929",
      "report_date": "20140820",
      "code_info": "Lot numbers to follow:  Z36-074, Q36-136, Q36-137, Q36-138, Q36-139, Q36-140, R36-168, R36-169, R36-170, R36-171, R36-172, S36-189, S36-190, S36-191, S36-192, S36-194, S36-202, S36-203.    RECALL EXPANDED 7-8-2014 to include ALL LOTS."
    }
  ]
}