  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Feldkirchen",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68737",
      "recalling_firm": "Brainlab AG",
      "address_1": "Kapellenstrasse 12",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) including the states of AL, AK, AZ, CA, FL, GA, IA, IL, KS, MD, MO, MS, NC, NY, OH, PA, SD, TX, VA, VT, WA, and WI, and the countries of Australia, Austria, Belgium, Brazil, Chile, China, Ecuador, France, Germany, Japan, Netherlands, New Zealand, Norway, Qatar, Russia, Spain, Switzerland, Taiwan, United Arab Emirates, and United Kingdom.",
      "recall_number": "Z-2134-2014",
      "product_description": "Patient Data Manager 2.0  (Content manager 2.0, Patient Browser 4.0, DICOM Viewer 2.0)   Brainlab Digital Lightbox.  A system for the display of medical images.   Model numbers 25100-05 and 25100-06.  The software can transfer images to and from picture archiving and communication systems (PACS), file servers, or removable storage media.",
      "product_quantity": "154 systems",
      "reason_for_recall": "The following defects have been identified for data loaded with Patient Data Manager version 2.0 (including subversions 2.0.0, 2.0.1 and 2.0.2):   1) When loading non-square pixel image data with PDM 2.0, the data set is not correctly transferred and displayed in the corresponding Brainlab navigation or planning software. When a data set is loaded, the pixels of the data set are sorted in order to",
      "recall_initiation_date": "20140627",
      "center_classification_date": "20140730",
      "termination_date": "20170424",
      "report_date": "20140806",
      "code_info": "Model/Catalog Numbers: 25100-05 PATIENT DATA MANAGER 2.0.11  25100-06 PATIENT DATA MANAGER 2.0.2.     Software Versions: 2.0.0, 2.0.1, and 2.0.2",
      "more_code_info": ""
    }
  ]
}