  "results": [
    {
      "country": "United States",
      "city": "Newark",
      "address_1": "500 Gbc Dr Ms 514",
      "reason_for_recall": "There is a misassembly issue with the  photometer installed on Dimension¿ Integrated Chemistry Systems instruments. Through internal investigation by  Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.",
      "address_2": "PO BOX 6101",
      "product_quantity": "15",
      "code_info": "DE271317, DE271320, DE271323, DE271324, DE271326, DE271327, DE271328, DE271332  DE271341, DE271343, DE271325, 12252589, 2004082898, 2004082899, 224719-AX",
      "center_classification_date": "20140730",
      "distribution_pattern": "FL  LA  AR  IA  OH  OH  WI  IL  LA  KS  MT",
      "state": "DE",
      "product_description": "Dimension¿ Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.",
      "report_date": "20140806",
      "classification": "Class III",
      "openfda": {},
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "recall_number": "Z-2133-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68400",
      "termination_date": "20141215",
      "more_code_info": "",
      "recall_initiation_date": "20140527",
      "postal_code": "19702-2466",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}