  "results": [
    {
      "status": "Terminated",
      "city": "Raynham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "64937",
      "recalling_firm": "DePuy Spine, Inc.",
      "address_1": "325 Paramount Dr",
      "address_2": "N/A",
      "postal_code": "02767-5199",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide:  Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK",
      "recall_number": "Z-2133-2013",
      "product_description": "CONFIDENCE SPINAL CEMENT SYSTEM 11CC PLUS KIT, Product Code:  2839-10-000    The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures",
      "product_quantity": "3896 units (Overall quantity distributed appears greater than quantities manufactured due to the distribution pathways)",
      "reason_for_recall": "During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.",
      "recall_initiation_date": "20130417",
      "center_classification_date": "20130903",
      "termination_date": "20140813",
      "report_date": "20130911",
      "code_info": "HPBB0K, HPBB3C,HPCB7J, HPDBD6, HPBB0L,HPBB3D,HPCB7W,HPDBF0, HPBB0M,HPBB3F,HPDBBC,HPDBG4, HPBB0N,HPBBWH,HPDBBD,HPDBGB,  HPBB0P,HPBBWJ,HPDBBK,HPDBGC,HPBB0T,HPBBWL,HPDBBL,HPDBGJ,  HPBB19,HPBBZ8,HPDBBY,HPDBGM,HPBB27,HPBBZD,HPDBCD,HPDBGN,  HPBB29,HPBBZN,HPDBCF, HPDBGY,HPBB2B,HPCB7B,HPDBD0,HPBB2C,  HPCB7C,HPDBD3,HPBB2G,HPCB7H,HPDBD4",
      "more_code_info": ""
    }
  ]
}