  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83096",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US nationwide distribution in the states of AZ, CA, FL, GA, IL, MI, MN, NC, NJ, NY, PA, TX, and VT, and countries of Algeria, Australia, Belgium, Canada, Chile, Columbia, Czech Republic, Finland, France, Germany, Hungary, Iceland, Japan, Korea (Republic of), Kuwait, Lebanon, New Zealand, Norway, Poland, Saudi Arabia, Senegal, South Africa, Sri Lanka, Sweden, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.",
      "recall_number": "Z-2132-2019",
      "product_description": "GE Pristina Serena - Product Usage: Pristina Serena is an optional accessory of Senographe Pristina intended to provide accurate location of lesions in the breast in three dimensions.",
      "product_quantity": "16",
      "reason_for_recall": "Potential slippage of the biopsy positioner when the gantry is rotated during a biopsy procedure.",
      "recall_initiation_date": "20190516",
      "center_classification_date": "20190730",
      "termination_date": "20201216",
      "report_date": "20190807",
      "code_info": "Serial# 738192BU4, 738574BU3, 734192BU8, 733706BU6, 718590BU3, 735558BU9, 738023BU1, 737211BU3, 739178BU2, 733600BU1, 738696BU4,  738793BU9, 738695BU6, 734169BU6, 733707BU4 and 732113BU6"
    }
  ]
}