  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74442",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US: AK, AZ, CA, COM, CT , DC, FL, GA, GU, HI, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MO, NV, NH, NJ, NY, NC, OH,OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WY.      OUS:  Algeria  Argentina  Australia  Austria  Bahrain  Belarus  Belgium  Bermuda  Brazil  Bulgaria  Canada  China  Egypt  France  Germany  Hungary  India  Indonesia  Ireland  Italy  Japan  Kazakhstan  Korea (Republic of)  Kuwait  Lebanon  Mexico  Netherlands  New Zealand  Poland  Qatar  Romania  Russia  Russian Federation  Saudi Arabia  Singapore  Spain  Sweden  Switzerland  Taiwan  Thailand  Turkey  United Arab Emirates  United Kingdom  Uruguay",
      "recall_number": "Z-2131-2016",
      "product_description": "GE Healthcare, Discovery MR750w",
      "product_quantity": "471 (184 US; 287 OUS)",
      "reason_for_recall": "GE Healthcare has recently become aware of a potential safety issue with the patient bore heating on the Discovery MR750w.    A small area on the bore surface, which is normally accessible to the patient, can get warm enough to cause a serious burn in the event the patient touches the bore and proper padding is not used.  No injury has been reported to GE as a result of this issue.",
      "recall_initiation_date": "20160608",
      "center_classification_date": "20160706",
      "termination_date": "20181004",
      "report_date": "20160713",
      "code_info": "Mfg Lot or Serial # System ID  00001250FMM03D 9075653T  00001308FMM00B 602863MR  Missing_SN_GON4205067 480342MR3  00001134FMM10E 928348MR  00001140FMM0FX RADNETMR597  N/A 530CHI750W  N/A 858824MR1  00000000UA0083 408356MR1  00001204FMM058 RADNETMR599  00000000UA0262 949760MR3  00000308185MR8 510449MR750W  00000000UA0075 619543HDMR4  00000000UA0118 858622HDMR1  00001204FMM052 415CB3TW  00001309FMM08B RADNETMR613  00001307FMM09Y 415UCBHMR3  1-4G33-86 805739MR3  310871MR9 310264MR2  309905MR8 RADNETMR621  00001204FMM04G 559734VIMR1  000005373011-3 RADNETMR598  N/A 720848MR4  N/A 720848MR3  311108MR5 303492MR  00001118FMM0WE 303604MR2  00001309FMM09E 720854MR  00000000UA0472 719365MR  Missing_SN_GON4077136 303837MR2  309815MR9 970624MR1  310950MR1 860446MR750W  00001251FMM00R 202966MR3TW  Missing_SN_GON4231312 561955MR1  00001309FMM097 386FNSBMR3T  00001308FMM042 954355MR750W  1309FMM09F 904483MR3T  00000000UA0465 305665MR750W  00001204FMM057 305461MR1  00001140FMM0FG 407SMSLMR1  00001147FMM0E5 386FHPCMR3T  00001223FMM08M 561844MR1  N/A 706787750DV  00001236FMM04Y 912819MR  00001301FMM043 671NAVYMR  00001234FMM0KL 671FHPMR750W  310330MR6 808HRAEMR750W  00001204FMM055 808545MR  00001250FMM035 808YMMR  00000306104MR1 808873MR750W  1-4GFG-928 319335MR2  310776MR0 773665SJMR2  PG75W1600017MR 312702MR750  00001250FMM02E 217876DMH3T  N/A 847316MR3T  00000305089MR5 708346MR3  00000307816MR9 708346MR4  310973MR3 309655IMRI  00001309FMM094 309655MR750W  N/A 815968MR  00001203FMM0VP 815395MR2  00000306062MR1 847872ZMR2  00000309665MR8 317887MR3  00001308FMM03T 317865MR4  00001308FMM006 913322MISHMRW  310253MR0 859334MR  00001221FMM06J 270798MR2  1406NMM0C2 318443MR2  00000307418MR4 225763MR2  00001236FMM051 337289MR1  1236FMM04H 337943MR1  00001251FMM01F BRIGHAM750WMR1  00001227FMM011 BOSTONMEDMR1  MISSING_SN_GON#4350388 508746MR3  00001250FMM03C VAWESTROX750W  00001309FMM085 410328MR3T  00001204FMM04U RADNETMR1652  00000306342MR7 RADNETMR1651  00001203FMM0W2 RADNETMR1635  00000303329MR7 RADNETMR1629  223355 3014643TMR  Missing_SN_GON4216878 240707MR3T  00001301FMM047 301570MR  00000000UA0301 RADNETMR1662  N/A RADNETMR6071  N/A RADNETMR1686  00001308FMM046 207MMC3TMR  130FMM07E 734845VA750MR  1442RG01YE 248533750W  00001227FMM00W 616267IMRI  00001134FMM10U 616774MR750W  00001308FMM01H 763795MR1  00000306613MR1 MR229047  00001250FMM01U MR229046  00001308FMM007 MR229048  UA0368 MR229050  00000000UA0091 314577CGMR3T  00000007T5MV12 336832MR750  00001233FMM06K 6033343TW  00001204FMM04E RADNETMR1631  00001203FMM0UK RADNETMR1632  311106MR9 856537MR  N/A 732MSKMR  00001251FMM01C 856581MR3T  00000304496MR3 856247MR3  00001233FMM06X 856247MR1  00001227FMM00T 201445750W  911001 775882GBIMR  00001308FMM02E 702382MR750W  00001309FMM065 702731MR750W  310333MR0 RADNETMR1687  PG75W1600002MR 718854MR  00001134FMM10R 71872MR750W  00000122FMM015 RADNETMR1648  310949MR3 631623MR750  N/A 516NS7113T  00001309FMM07F 914750W  00001203FMM0VB EAST750W  PG75W1600021MR MSK68MR03  00001308FMM008 MSK68MR7  00000308347MR4 MSK68MR1  00000000UA0363 MSK53MR02  00001140FMM0FT 21277MR1  N/A 212288MR4  00001236FMM04X COR526967  1406NMM058 716740MR1  00000305184MR4 RADNETMR1639  00001308FMM02T 718226MR3T  Missing_SN_GON4292927 914631MR  00000305090MR3 330344MR750W  00001308FMM02C 513636MR4  00001111FMM0B2 61456MR5  309752MR4 614293MR9  00001309FMM09G 6142343T  00001250FMM02H 937641MR2  N/A 937208MR1  00001142FMM08Y 405271OUCMR  N/A 405636SMR2  1307FMM088 503251PMR3  1406NMM0C5 412914MR  310902MR2 215829MR750W  N/A 215728MR3T  00001308FMM032 PRESBYMR1  00001308FMM028 843815750W  00000306781MR6 843750WGEM  00000306801MR2 864529MR  309780MR5 423493MR750W  00001309FMM047 865984MR750  309948MR8 817433HSWMR2  310952MR7 512544SD3T  00001250FMM02W 214645UT3TMR1  00001308FMM02A 2148203TMR  N/A 9157473TMR  N/A 409772MR6  00000000UA0396 713792AMR14  00001308FMM05P 713798MR1  PG75W1600019MR 713792AMR16  N/A 903236MR750W  PG75W1600011MR 713363MR750  00001251FMM01E 210617HUBMR  N/A 210WHASCMR1  00000000UA0304 210AMVAMR  00001134FMM10C 281274MR2  N/A 903525MR750W  00001204FMM04Y 435251MR1  PG75W1500005MR 434MJMR2  PG75W1500001MR 703698MR5  00000000UA0124 703698MR4  00001307FMM089 703573MR1DV  00000304862MR6 757827MR2  00001309FMM0AF 703478MR2DV  00000000UA0141 757983MR  N/A 206223VM750W  00001221FMM06M 206292SR750W  N/A 206320CH750W  310775MR2 206781SB750W  1307FMM086 920738MR750  00001250FMM02N 608775MR3T  00001251FMM012 608263MR5W  00001309FMM07N 608CHILDMR1  310777MR8 608VAMR750W  00000310697MR8 608265MR750W  N/A 414649WB2  00001250FMM03A 414647MR1  00001309FMM05F 715847MR3T  Missing_SN_GON4216470 307BHBBJ3T  00000000UA0437 DZ1060MR04  00000310539MR2 DZ1638MR01  00001234FMM0KB MRUA0136  00001309FMM08P MRUA0343  00000309907MR4 83112923995  00000309950MR4 83112923777  00000000UA0226 83096535911  301100984 83092403928  00000308593MR3 83096536248  00000308883MR8 83104203907  00000309051MR1 83096322711  00000309127MR9 83105089955  00000309327MR5 83104653897  00000309453MR9 83108956002  00001134FMM10P 910274040  00001203FMM0VW 910273043  00000309423MR2 AP7645MR01  00001308FMM02N BH1017MR01  00001139FMM0D7 BY4160MR01  00001219FMM0GM 140024MR01  00001233FMM06W 130083MR05  00001308FMM02R 220078MR01  0000310163MR01 220826MR02  00001406NMM0C0 220826MR01  00001309FMM084 441239MR750W  00000000UA0467 MRUA0467  00000309092MR5 MRUA0272  00001309FMM072 MRUA0191  00000310904MR8 MRUA0394A  00001301FMM03E MRUA0122  00001406NMM00K BG4680MR01  663 506648MR4  00000000UA0364 E416249MR2  00001134FMM104 613737MR1  00001437RG01DD 82427040139  00001441RG00L1 82427200086  00001442RG01XE 82427160158  00001437RG01DK 82427160160  00001442RG01XC 82427180035  00000309522MR1 82427160149  00001442RG01Y0 82427170020  00001442RG01XK 82427070157  N/A 82427010087  00001441RG00KT 82427010080  00000305814MR6 82427210118  00001442RG01YT 82427210129  00001438RG00L1 82427210149  00000306638MR8 82427120116  00001406NMM082 82427120122  00000307589MR2 82427120124  00001438RG00L6 82427120159  00001437RG01DM 82427120149  00000307210MR5 82427120110  00001406NMM0BP 82427120136  00000306700MR6 82427040146  00000306350MR0 82427040135  00001442RG01Y9 82427040167  00001308FMM035 82427160127  00001406NMM09V 82427110051  00001406NMM0BE 82427120131  00001309FMM0A8 82427100225  00001406NMM0BJ 82427180028  00001406NMM0B7 82427130076  00001442RG01YR 82427120135  00000309521MR3 82427060052  00000307568MR6 82427100219  00001442RG01Y4 82427100267  00001442RG01YK 82427100254  00001438RG00L3 82427100265  00001442RG01YJ 82427100246  00001438RG00LG 82427140124  00000000UA0482 EG1526MR05  00000310953MR5 A1765918  00001309FMM0A6 M4018549  00001134FMM10K M4166994  00000310538MR4 M4172341  00000311037MR6 M4035023  00000303954MR2 M9481853  00001406NMM0CN M365865301  00001204FMM05C N4192951  00001441RG00JR X41944010  00000309695MR5 M54606125  00000000UA0016 X59007321  00001204FMM048 M2322246  00001204FMM05F M45052106  00001204FMM05J PC9953MR01  00000305145MR5 UC2547MR01  00001236FMM05D HC0220MR03  00001307FMM08H TP4447MR02  00001309FMM04U PC3003MR01  00001117FMMOCN PNUE08MR01  00001117FMM0CP PC9595MR01  0000000UA0332A 409740MR02  00000302668MR9 HC3068MR01  00001140FMM0FF HU1098MR02  00001221FMM06K 8.30278E+13  00000301768MR8 8.30278E+13  00001406NMM05A 8.30272E+13  00000310060MR9 8.30272E+13  00000309164MR2 8.30279E+13  00000309951MR2 8.30279E+13  00001236FMM05Y 8.30272E+13  00000306902MR8 8.30272E+13  00001118FMM0WJ 8.3027E+13  00001250FMM01T 8.3027E+13  00001204FMM054 8.30271E+13  00000311040MR0 8.3027E+13  00001203FMM0V5 835270026  00001308FMM065 IE1010MR01  00000310162MR3 A5547422  00001309FMM05G A5360437  00001301FMM03X A5111746  00001309FMM07R B5174327  00001309FMM0AJ A5719869  00001203FMM0VY B5129925  00001236FMM04P EM0271  00001308FMM05M EM0272  00001308FMM04T EM0229  00001308FMM063 EM0232  00001236FMM00M EM0215  00001234FMM0KA EM0221  00001236FMM01R EM0214  00001309FMM06A EM0245  00001112FMM081 EM0202  00001227FMM00J EM0208  00001250FMM01W EM0227  00001308FMM012 EM0254  00000310839MR6 EM0276  00001236FMM01G EM0213  00001301FMM04H EM0235  00001308FMM05B EM0234  00001406NMM0C7 EM0262  00000310905MR5 EM0269  00000310906MR3 EM0273  00001250FMM02U EM0228  00001236FMM016 EM0238  00000310702MR6 EM0277  00000310474MR2 EM0267  00001308FMM01Y EM0249  00001140FMM0FU EM0205  00001236FMM05E EM0239  00001147FMM0EE EM0222  00001227FMM00C EM0225  00000304861MR8 EM0206  00001221FMM0BB EM0210  00001308FMM03L EM0248  PG75W1500006MR EM0286  00000000UA0488 EM0285  PG75W1500007MR EM0290  00000310677MR0 EM0268  00000310475MR9 EM0264  00000000UA0507 EM0284  PG75W1600006MR EM0289  0000000UA0427A EM0282  00001406NMM080 EM0263  N/A EM0288  00000310588MR9 EM0287  00000310903MR0 EM0274  00000310420MR5 EM0266  00000310720MR8 EM0280  00000310421MR3 EM0281  00001406NMM05U EM0270  00001308FMM034 EM0253  0000000UA0397B EM0283  00001236FMM01H EM0241  00001236FMM020 EM0240  00001308FMM022 EM0244  00001308FMM055 EM0231  00001309FMM09Y EM0257  00001220FMM01J EM0207  00001309FMM051 EM0261  00001308FMM01U EM0265  00001309FMM095 EM0275  00001203FMM0VX EM0219  00001308FMM01P EM0255  00001308FMM020 EM0256  00001203FMMOUW EM0212  00001236FMM05U EM0246  00001234FMM0KH EM0223  00001309FMM08Y EM0243  00001203FMM0V2 EM0230  00001221FMM06R EM0220  00001301FMM04V EM0233  00001250FMM02G EM0226  00001309FMM05Y EM0247  00001250FMM02Y EM0236  00001117FMM0CE EM0204  00000310976MR6 EM0278  N/A EM0279  00001134FMM10L EM0201  00000308133MR8 EM0237  00001147FMM0E8 EM0217  00001308FMM027 EM0242  00001309FMM08H EM0258  00001227FMM00V EM0211  00001140FMM0FL EM0218  00000305986MR2 EM0224  00000302811MR5 EM0203  00001233FMM076 EM0250  00001301FMM03C EM0260  00001220FMM01H EM0209  00000000UA0358 EM0259  00001236FMM00X EM0216  00001406NMM02J KA1023MR02  0000000JWB0DZ1 850270459  00001204FMM04V 850270386  00000306791MR5 850270430  00001250FMM03E 850270426  00000000UA0170 850270427  00000000UA0483 850270478  00001233FMM06M 850270414  00000308159MR3 850270436  00001221FMM06X 850270409  00001309FMM04W 850270446  00001309FMM055 850270455  00000311338MR8 850270481  00001227FMM00X 850270411  00000310542MR6 KW1100MR02  00001236FMM05A KW1005MR04  00001309FMM096 KW1019MR03  00000301969MR2 LB1331MR01  00001250FMM028 LB1160MR02  00001301FMM04M LB1006MR02  00001308FMM02P MR308183MR3  00000304860MR0 RTD0191  00001147FMM0E6 UDN0121  00000310773MR7 85110163303  00001140FMM0G4 920279031  00000307192MR5 85092252519  00001308FMM05H PL3116MR01  00000000UA0480 PL3554MR01  00001250FMM02P 600042MR02  00001309FMM0AE PL3277MR01  00001308FMM05C QA1042MR01  00001204FMM04A RO1098MR02  00000310304MR1 RU1111MR02  00001406NMM01E RU1717MR01  00000000UA0227 RU8438MR01  00000000UA0229 RU8438MR03  00001234FMM0KG RU5599MR01  00001307FMM08M RU8792MR01  00001308FMM03B RU8615MR01  00001406NMM02X RU8937MR01  00001308FMM064 SA1223MR01  00001135FMM019 SA1033MR01  00001309FMM05E SA1255MR03  00001309FMM087 SA1058MR03  00000000UA0333 6073MR0003A  00000000UA0023 847270026  00001307FMM07X 6073MR0003  00001142FMM096 MPX94401  00001309FMM08R UE929526  00000000UA0425 DPN36500  PG75W1600001MR A202MR02  00001251FMM014 O009MR03  00000306101MR7 M001MR02  00001251FMM00J M004MR01  00001236FMM04T T002MR01  00001134FMM10D A010MR08  00001309FMM076 C001MR01  00001309FMM05J N002MR03  00001250FMM039 387040MR10  00000310540MR0 259130MR12  00001309FMM05N 150250MR01  00001309FMM06W 725100MR02  00001142FMM092 120810MR02  0000000UA0430A 568987MR01  00000000UA0049 850020MR02  00000310703MR4 123953MR01  00001308FMM048 853270141  00000310604MR4 856270071  00001309FMM07G 06162MRS01  00001227FMM00N 26001MRS01  00001204FMM04B 34348MRS01  00000309908MR2 53008MRS01  00000000UA0097 AE1016MR01  00001142FMM090 10639MRS01  00001309FMM04T 10692MRS01  00001309FMM07M 77266MRS03  00001142FMM094 10678MRS01  00001250FMM038 00010MRS02  00001111FMM0B7 70431MRS04  00000000UA0265 00230MRS01  N/A 00440MRS08  00000310424MR7 10722MRS01  00000310701MR8 20312MRS01  00001406NMM09X MRUA0353"
    }
  ]
}